Differentiation between Benign and Malignant Breast Lesions Using ADC on Diffusion-Weighted Imaging at 3.0 T ()
Received 25 November 2015; accepted 5 March 2016; published 9 March 2016

1. Introduction
One of the most common types of tumor affecting female nowadays is breast cancer. Annually, over 1,000,000 new cases are diagnosed worldwide [1] . So to reduce mortality rates caused due to breast cancer we need early diagnosis and treatment to increase survival and improve the quality of life, hence diagnostic imaging is necessary with accuracy. Over the past few years, MRI has been used to study breast lesions [2] -[4] . Magnetic resonance imaging is a widely accepted modality as an adjunct to mammography and ultrasonography in the detection of primary or recurrent breast cancer [5] . Hence, several studies have been investigating the role of advanced MRI techniques, such as diffusion-weighted imaging (DWI), to improve the specificity of diffusion weighted MRI for the evaluation of breast carcinoma [6] -[10] . DWI measures the random movement of water molecules in tissues, which is detected as the attenuation of the measured signal intensity [11] on Diffusion weighted MRI (DWI). Diffusion can be quantified by using the apparent diffusion coefficient (ADC) value to increase the specificity of DWI [12] . The movement of water molecules is much more restricted in tissues with a higher cellular content and associated with numerous intact cell membranes [13] [14] as seen in malignant breast lesion. The cell density is increased in malignant breast cancer, which corresponds to lower ADC values [13] [15] -[17] on DWI.
While the sensitivity of dynamic contrast-enhanced (DCE) MRI in the diagnosis of breast cancer is approximately 85% - 99%, its specificity ranges between 40% and 80% as stated by Kuhl, C.K. et al. [18] . Diffusion- weighted imaging (DWI) is the techniques that are performed to increase MRI specificity [7] [8] . The sensitivity of MRI for breast cancer detection ranges from 85% - 100% [19] [20] , and specificity ranges from only 37% - 88% [21] [22] . DWI-MRI is used to detect cancers of the body easily as it has shown that diffusion of the water molecules in malignant tumors is slower compared to the normal body parenchyma.
The purpose of this study is to investigate the diagnostic efficiency of the apparent diffusion coefficient value (ADC) in differentiating benign from malignant breast lesions at 3.0 T Diffusion weighted MRI with different pairs of b value.
2. Materials and Methods
2.1. Subjects
A retrospective clinical breast DWI MRI examination was reviewed from March 2011 to April 2015 in our hospital. The indication for MR imaging were mainly women suspected of breast cancer, high risk women, preoperative local staging, follow up after conservative breast surgery, detection of unknown primary cancer and women suspected of breast lump in conventional mammography or ultrasonography. This study was approved by the institutional review board, and informed written consent was obtained from all patients after the procedure had been explained. Patients enrolled into this study had diagnostic procedure, including the following: a) 3.0T DW-MRI done prior to surgery b) histopathological confirmed diagnosis after surgery c) Good‑quality DW MR scans images. Exclusion criteria were 1) Patient with inflammation and cysts 2) Patient with neo-adjuvant chemotherapy 3) those with poor DWI or DCE-MRI image quality 4) those with breast implants 5) those who fail to fat suppression were excluded from the study as it could give us false readings.
On the basis of these criteria, we excluded eight lesions in eight patients (two were breast cyst, three were due to chemotherapy, two due to poor image quality and one due to breast implants). As a result, the study included a total of 110 women with 107 lesions (mean age 46.09 years; range 25 - 68 years) diagnosed with benign and malignant breast lesion constituted our study group. The mean size of benign lesions was 1.9 cm (range 0.9 cm - 4.2 cm), while that of malignant lesions was 4.5 cm (range 1.0 cm - 10.2 cm). Of the 107 lesion 44 were benign and 63 were malignant lesion. Among the 44 benign included 22 fibroadenoma, 18 adenosis, 2 papilloma, 1 lobulated tumor and 1 scleorsing adenosis and among 63 malignant tumor included 37 IDC (Invasive ductal Carcinoma), 12 DCIS (Ductal Carcinoma in situ), 9 Invasive lobular carcinoma(ILC), 5 mucinous carcinoma, 2 medullary carcinoma, 1 squamous cell carcinoma and 1 papillary carcinoma.
2.2. MRI Protocol
All DWI-MRI was performed on 3.0-T MR System (Magnetom Verio, Siemens, Germany) by using a dedicated bilateral breast coil with patients in prone position. Before DWI, standard sequences were acquired, including an axial fat-suppressed T2-weighted images turbo inversion recovery magnitude (TIRM; TR/TE,4300/61; inversion time, 150 milliseconds; bandwidth, 319 Hz/Px; matrix size, 320 × 320; field of view, 340 mm; number of signals averaged, 2; slice thickness, 4 mm;), axial T1-weighted spin-echo sequence (TR/TE, 6.04/2.45; bandwidth, 400 Hz/Px; matrix size, 512 × 512; field of view, 340 mm; number of signals averaged, 1; slice thickness, 1.3 mm;), axial T2-weighted spin-echo sequence (TR/TE, 1000/87; bandwidth, 244 Hz/Px; matrix size, 512 × 512; field of view, 340 mm; number of signals averaged, 1; slice thickness, 2.0 mm), and axial T1-weighted 3D fat-suppressed fast spoiled gradient-echo sequence (TR/TE, 4.67/1.66; bandwidth, 320 Hz/Px; matrix size, 384 × 384; field of view, 360 mm; slice thickness, 1.20 mm ) before and eight times after the rapid injection of a bolus of intravenous contrast medium (gadopentetate dimeglumine) at a dose of 0.2 mmol/kg body weight infused at the rate of 2 ml/sec, followed by 20 ml of saline solution. After the examination, the unenhanced images were subtracted from the first and last contrast-enhanced images.
DCE-MRI images were obtained with a T1W 3D FLASH sequence using spectrally adiabatic inversion recovery (SPAIR) for fat suppression. DWI was performed before the DCE-MRI acquisition in an axial plane using a 2D spin echo-echo planar imaging (SE-EPI) sequence centered on the lesion acquired in 7 minutes 21 seconds for the DWI scan (b = 0, 400, 800, 1200, and 1600 s/mm2; repetition time (TR)/time to echo (TE): 10,500/87 ms; bandwidth, 1184 Hz/Px; matrix size, 192 × 192; field of view, 350 mm; number of slices, 26; slice thickness, 4 mm; number of signal averaged, 3; PAT mode GRAPPA; acceleration factor PE, 2; total scan time 7:21sec.
2.3. Analysis of MR Images
All images were transferred to a workstation (Magnetom Verio, Siemens) and the DWI sequence was post processed with the software (Magnetom Verio, Siemens) to obtain ADC maps (black-and-white). Two researchers retrospectively analyzed the diffusion-weighted (DW) Images and at the time of DWI analysis, readers were unaware of histological results. The ADC maps of each lesion were calculated using two b values (0 and 400, 0 and 800, 0 and 1200, 0 and 1600 s/mm2). The dynamic contrast enhancement MR (DCE-MR) was used as the reference to locate the lesion. To achieve standardized conditions for analyses and to avoid contamination of the data by adjacent structures, three circular regions of interest (ROIs) were drawn manually in the central region of the homogeneous breast lesions in the same location as the five DW imaging and the four ADC maps cited above, and the average ADC value was recorded for each b value combination. Apparent necrotic, cystic or hemorrhagic components were avoided by referring to conventional MR images.
The ROI sizes were set to a mean diameter of 50 mm2 for breast lesions and the mean value from 3 ROIs were recorded for ADC.
2.4. Statistical Analysis
Data were analyzed using the IBM SPSS 21.0 software package (SPSS Inc, Chicago, IL, USA). All Quantitative data were expressed by mean ± standard deviation. Mean ADC value for benign and malignant breast lesions were calculated. Differences in mean ADC between benign and malignant breast lesion were calculated using independent sample t-test. Receiver operating characteristic (ROC) analysis was performed to determine the best cut off score and area under curve (AUC) for each pair of b value using ADC, so that benign and malignant lesions could be differentiated. ROC curve analysis was also used to determine the diagnostic efficiency of the ADC to calculate the sensitivity, specificity, PPV, NPV and accuracy rate. Statistical significance was set at p < 0.05.
3. Results
3.1. Comparison of the ADC Values of Benign and Malignant Lesions for Different Pairs of b-Values
Table 1 illustrates the mean ADC values for benign and malignant breast lesions for all the pairs of b value (0 - 400, 0 - 800, 0 - 1200, 0 - 1600 s/mm2) and the results shows that there is significant differences between the benign and malignant breast lesions (p < 0.05). The mean ADC value for benign were found to be 1.73 × 10−3 mm2/s for b0 and 400, 1.57 × 10−3 mm2/s for b0 and 800, 1.43 × 10−3 mm2/s for b 0 and 1200 and 1.30 × 10−3 mm2/s for b0 and 1600 s/mm2. The mean ADC value for the malignant breast lesion were found to be 1.21 × 10−3 mm2/s for b0 and 400, 1.06 × 10−3 mm2/s for b0 and 800, 0.94 × 10−3 mm2/s for b0 and 1200 and 0.86 × 10−3 mm2/s for b0 and 1600 s/mm2. The results shows that the mean ADC values for the malignant breast lesion had lower ADC value compared to benign breast lesion for all b values in pair as shown in Figure 1, and the mean ADC value decreases with the increase in b value pair for both the benign and malignant breast lesion.
3.2. Diagnostic Efficiency of ADC Value Using ROC Curve for Different Pairs of b-Values
ROC (Receiver Operating Characteristics) curves for all the pair of b values are represented in Figure 2. The
![]()
Table 1. Comparison of the ADC values for benign and malignant breast lesions.
Note: *Quantitative data represented as Mean ± S.D. **p ?value shows statistically significant (p < 0.05). ADC: apparent diffusion coefficient; SD: standard deviation.
![]()
Figure 1. Mean ADC value at 3.0 Tesla using different pairs of b value in differentiating benign and malignant breast lesion. The Mean ADC value is indicated along the y-axis. The mean ADC for malignant breast lesions are significantly lower than benign breast lesion for all pair of b-value (0 - 400, 0 - 800, 0 - 1200, 0 - 1600 s/mm2).
![]()
Figure 2. ROC curves. Receiver operating characteristic (ROC) curves derived from Apparent Diffusion Co-efficient (ADC) in differentiating malignant from benign lesions at 3.0 Tesla DWI. The areas under the curves (AUC) at b values of 0 - 400, 0 - 800, 0 - 1200 and 0 - 1600 s/mm2 were 0.87, 0.88, 0.89, and 0.89 respectively.
ROC diagnostic efficiency of ADC for different b value pair were statistically significant in differentiation of benign and malignant breast lesion (p = 0.00) (Figure 3, Figure 4). The cut-off ADC value was found to be 1.35 × 10−3 mm2/s for b 0 and 400, 1.22 × 10−3 mm2/s for b 0 and 800, 1.08 × 10−3 mm2/s for b 0 and 1200 and 0.99 × 10−3 mm2/s for b0 and 1600 s/mm2 as demonstrated in Table 2. The mean ADC value, which was equal or less than the cut-off value, suggested malignant breast cancer. The Sensitivity, Specificity, PPV, NPV and accuracy were 80.95%, 90.9%, 92.72%, 76.92%, 85.04% for b 0 and 400; 84.12%, 90.9%, 92.98%, 80%, 86.91% for b0 and 800; 84.12%, 90.9%, 92.98%, 80%, 86.91% for b0 and 1200; 84.12%, 90.9%, 92.98%, 80%, 86.91% for b0 and 1600 s/mm2 respectively. Significant result were obtained for all the b value combination (p < 0.05) as shown in Table 2 obtained from ROC curves analysis using ADC values in differentiating benign and malignant breast lesion at 3.0 Tesla DW-MRI.
![]()
Figure 3. 44 year female, fibroadenoma left breast. (a)-(e), DWI with b-value of 0 s/mm2, 400 s/mm2, 800 s/mm2, 1200 s/mm2, 1600 s/mm2 reveals homogeneous high signal intensity in the left breast (arrow) (f)-(i), Apparent diffusion coefficient (ADC) map with b-value of 400 s/mm2, 800 s/mm2, 1200 s/mm2, 1600 s/mm2 shows hypo intensity lesion (arrow). The signal intensity of the benign lesion was significantly higher than the normal glandular tissue, meanwhile with the increasing of b-value the signal intensity decreased and the declining degree of signal intensity of benign breast lesion is higher than that of malignant lesions shown in Figure 4 for each of the b-values.
![]()
Figure 4. A 41-year-old woman with invasive ductal carcinoma (IDC) left breast. (a)-(e), DWI with b-value of 0 s/mm2, 400 s/mm2, 800 s/mm2, 1200 s/mm2, 1600 s/mm2 reveals homogeneous high signal intensity in the left breast (arrow) (f)-(i), Apparent diffusion coefficient (ADC) map with b-value of 400 s/mm2, 800 s/mm2, 1200 s/mm2, 1600 s/mm2 shows low signal intensity lesion as restricted diffusion (arrow).
![]()
Table 2. Diagnostic efficiency of ADC in differentiation of breast lesion using ROC curves.
Note: *AUC: Area under Curve from ROC curves. **p value shows statistically significant (p < 0.05). PPV = Positive predictive value; NPV = Negative predictive value; ACC = Accuracy Rate.
4. Discussion
Water diffusion in body tissues is the process that reflects the exchange of different components between the extra cellular (ECF) and intracellular (ICF) fluid compartment and the interstitial fluid in human body. In normal human tissues water movement is affected by the cell membranes and macromolecules within the cell. Diffusion Weighted Imaging (DWI) shows the random motion of water molecules and is the only method of measuring molecule diffusion [23] . Many research shows that the breast lesions may be differentiated by calculating the mean ADC values, with malignant cancer yielding lower ADC values compared to benign or normal tissues [24] -[26] .
In our study we tested the diagnostic efficiency of ADC in differentiating benign and malignant breast lesion. The mean ADC value for the malignant breast lesion was significantly lower than the benign lesion for all the combination of b values used in our study. This finding was as same as that of the literature stated by [16] [24] [27] -[34] . We also noted that with the higher b values combination the ADC value decreases. However, statistical differences (p < 0.05) was seen for all the mean ADC values calculated from different b values combination in our study. Many studies related to DWI have shown results in differentiating benign and malignant breast lesions with sensitivity ranging from 81% to 93% and specificity from 80% to 88.5% [13] [15] [24] [26] [35] -[37] . Our results corroborate with those previous studies and the calculated ADC values from different b values combinations and all of them were statically significant.
We found that diffusion restriction in a lesion as a bright signal (high intensity) on DWI and as dark signal (low intensity) on corresponding ADC map, having a Sensitivity, Specificity, PPV, NPV and accuracy rate were 80.95%, 90.9%, 92.72%, 76.92%, 85.04% for b0 and 400; 84.12%, 90.9%, 92.98%, 80%, 86.91% for b0 and 800; 84.12%, 90.9%, 92.98%, 80%, 86.91% for b0 and 1200; 84.12%, 90.9%, 92.98%, 80%, 86.91% for b0 and 1600 s/mm2 respectively for diagnosing malignant breast cancer. This study demonstrates ADC value were same for b0 - 800, b0 - 1200, and b0 - 1600 s/mm2 (Table 2). So we conclude that the use of higher b value in combination can be used in differentiating benign and malignant breast lesion at 3.0 Tesla DWI as per recent meta-analysis by Dorrius et al. [38] , also stated result similar to our finding that those of b = 0 and 1000 s/mm2 are recommended for the best improvement of differentiating between benign and malignant lesions, hence instead of using multiple DWI with different b values it is wise to perform this sequence as per our studied b value combination, either with an intermediate or high b value. This way, the acquisition scan time will be shortened, which is comfortable for the patients and less artifacts due to movement of patient. Our study has some limitations firstly, the patients group included in this study was comparatively small, and secondly it was a retrospective study performed in our university so our study could be selection bias. Thirdly, use of only 5 b values in our study, so we need to include more b value range in our study. Nevertheless, further study of larger population may be needed to confirm the diagnostic efficiency of ADC in differentiating benign and malignant breast lesions.
5. Conclusion
The determined mean and cut-off ADC values in different studies should not be accepted as absolute values. We authors believe that each institution should define its own mean and cut-off ADC values for differentiating benign and malignant breast lesions.