Acetabular Component Positioning and Risk of Dislocation in Hip Arthroplasty: Is Lewinnek’s Safe Zone Truly Safe?

Abstract

Objective: The present research aims to determine if adherence to the Lewinnek safe zone, when exclusively considered, constitutes a pivotal element for ensuring stability in the context of total hip arthroplasty. This is done by examining the acetabular placement in instances of hip dislocation after total hip arthroplasty (THA). Methodology: The authors searched 2653 patient records from 2015 to 2022 looking for patients who had total hip arthroplasty at our facility. For the analysis, 23 patients were culled from 64 individuals who exhibited post-THA dislocations, employing a stringent exclusion criterion, and the resultant acetabular angulation and anteversion were quantified utilizing PEEKMED software (Peek Health S.A., Portugal) upon radiographic evidence. Results: Within the operational timeframe, from the cohort of 2653 subjects, 64 presented with at least a singular incident of displacement. Post-exclusion criterion enforcement, 23 patients were eligible for inclusion. Of these, 10 patients conformed to the safe zone demarcated by Lewinnek for both inclination and anteversion angles, while 13 exhibited deviations from the prescribed anteversion and/or inclination benchmarks. Conclusion: Analysis of the 23 patients reveals that 13 did not confirm to be in the safe zone parameters for anteversion and/or inclination, whereas 10 were within the safe zone as per Lewinnek’s guidelines. This investigative review, corroborated by extant literature, suggests that the isolated consideration of the Lewinnek safe zone does not suffice as a solitary protective factor. It further posits that additional variables are equally critical as acetabular positioning and mandate individual assessment.

Share and Cite:

Meyer, O. , Wink, L. , Iserhard, W. , Schwartsmann, C. , Yépez, A. and Spinelli, L. (2025) Acetabular Component Positioning and Risk of Dislocation in Hip Arthroplasty: Is Lewinnek’s Safe Zone Truly Safe?. Open Journal of Orthopedics, 15, 47-53. doi: 10.4236/ojo.2025.152006.

1. Introduction

Currently, beyond one million total hip arthroplasties are performed worldwide each year [1]. The most common surgical indications are severe osteoarthritis, fractures of the femoral neck or head, and femoral head osteonecrosis. Among the most feared postoperative complications, dislocation stands out, both for its severity and the added cost to the procedure. According to literature data, dislocation occurs at 1% - 3% and ranks as the leading reason for revision surgery [2] [3]. Different factors have been studied as potential contributors to postoperative instability, with acetabular positioning being of particular interest [4]-[11].

In 1978, Lewinnek et al. studied a series of postoperative radiographs and found that patients predisposed to dislocation had their acetabulum positioned outside a predefined range of angles. The author concluded that the safest zone for acetabular positioning would be within 40˚ ± 10˚ of inclination and 15˚ ± 10˚ of anteversion. This concept, known as Lewinnek’s SAFE ZONE, has been widely disseminated and remained an absolute truth for decades [11].

Recent studies have questioned the concept of the safe zone described by Lewinnek, finding no absolute correlation between acetabular positioning and the occurrence of dislocation [4]-[10]. Seagrave et al. conducted a systematic review, analyzing 11 articles related to acetabular positioning. In only 2 of the 11 articles, it was statistically significant that the majority of patients experiencing dislocations were outside the safe zone. In 7 of the evaluated articles, patients with prosthesis dislocations were primarily located within the safe zone [10].

The present study aims to assess the relationship between acetabular positioning and the occurrence of dislocation in patients treated at our institution by measuring acetabular anteversion and inclination. It also aims to determine whether the safe zone genuinely reduces the risk of postoperative dislocation.

2. Materials and Methods

This study was submitted and approved by the Ethics Committee Board of our institution and performed in accordance with the Declaration of Helsinki. We reviewed medical records and radiographs of patients who underwent hip arthroplasty between 2015 and 2022 at our institution and experienced at least one episode of postoperative dislocation.

Patients who experienced dislocation in revision or partial prostheses were excluded, as were those with unavailable or poorly executed imaging examinations that prevented anteversion and inclination measurements, and patients with dislocations associated with periprosthetic fractures.

The measurements of acetabular component anteversion and inclination angles were performed using Peekmed software (Peek Health S.A., Portugal) on plain radiographs. This software is FDA-certified in the United States. Statistical analysis was conducted using SPSS 26, considering mean values, standard deviation, and percentages.

3. Results

Between 2015 and 2022, 2653 hip arthroplasties were performed at our institution, and 64 patients experienced at least one episode of prosthesis dislocation. After applying the exclusion criteria, 23 patients remained to be evaluated.

Of these patients, 9 were male, and 14 were female. The mean age at the time of dislocation was 66 years, ranging from 51 to 82 years. The time between surgery and the first episode of dislocation ranged from 1 day to 7 years.

Anteversion measurements of the acetabular component ranged from 1.8˚ to 39.5˚, with a mean of 14.9˚. Inclination measurements of the acetabular cup ranged from 23.3˚ to 67.1˚, with a mean of 44.7˚, as shown in Table 1.

Out of the 23 patients evaluated, only 5 had anteversion measurements outside the safe zone (15˚ ± 10˚), and 10 had inclination measurements outside the safe zone (40˚ ± 10˚), as indicated in Table 2. Among the 23 patients, 10 had both anteversion and inclination measurements within Lewinnek’s established limits (Figure 1).

4. Discussion

Lewinnek et al.’s original article published in 1978 continues to be widely cited and debated in orthopedic literature, despite its evident limitations. The authors relied on a small sample, analyzing only 9 cases of dislocation, with 6 of these

Table 1. Angle values found in the 23 evaluated patients.

Anteversion

Inclination

Mean Value

14.9

44.7

Standard Deviation

10.0

12.5

Minimum Value

1.8

23.3

Maximum Value

39.5

67.1

Table 2. Anteversion and inclination values found versus number of samples.

Angle

N

%

Anteversion

5 - 25 degrees

17

73.9

<5 or >25 degrees

6

26.1

Inclination

30 - 50 degrees

13

56.5

<30 or >50 degrees

10

43.5

Figure 1. Graph demonstrating the values of inclination and anteversion found in the 23 evaluated cases with a highlight on the safety values according to Lewinnek et al.

involving revision prostheses, which are considered an independent risk factor for dislocation. In this study, we evaluated the relationship between acetabular positioning and dislocation occurrence in patients treated at our institution by measuring acetabular anteversion and inclination to determine whether the safe zone genuinely reduces the risk of postoperative dislocation [1]-[3] [11].

Several recent articles have questioned the efficiency of Lewinnek’s safe zone in preventing acetabular dislocation. These authors are unable to pinpoint a single factor to explain cases of dislocation following hip arthroplasty, suggesting a possible multifactorial cause. Other factors may, therefore, explain postoperative instability in patients with the acetabulum positioned within the safe zone, such as femoral head size, femoral anteversion, femoral offset, and others [4]-[10].

The ideal positioning remains a topic of discussion. Incorrect acetabular positioning could result in femoral neck impingement with the acetabular rim or the polyethylene or ceramic insert. Such impingement can act as a fulcrum, causing femoral head displacement or even causing microtrauma in the polyethylene, ceramic, or metal, releasing abrasive debris that accelerates wear and aseptic loosening [12]-[14].

In a study involving prostheses removed from patients during revision, Yamaguchi et al. observed that patients with excessive anteversion of the acetabular component in radiographs exhibited signs of anterior acetabular rim impingement. These impingement episodes would occur when the patient flexed and internally rotated the hip while sitting and were related to episodes of posterior dislocation [13]. Greater acetabular inclinations were associated with increased polyethylene wear and a higher risk of dislocation. However, smaller inclinations are associated with a loss of range of motion [14] [15].

The ideal positioning is likely one that allows for maximum range of motion without the consequence of femoroacetabular impingement and instability. Kummer et al. conducted a study on synthetic models and suggested that this positioning would be between 35 and 45 degrees of inclination and 0 to 10 degrees of anteversion [12]. Subsequent studies on three-dimensional models demonstrated that this depends not only on acetabular anteversion but also on the combined anteversion of the acetabulum and femoral neck [14] [16]-[18]. According to some authors, the combined anteversion value with the lowest dislocation risk is between 40 and 60 degrees [16] [18].

Another concept currently under debate is the so-called functional acetabular positioning. It is based on the fact that the pelvis is not static throughout the day. When assuming a seated position, for example, the patient tilts the pelvis approximately 20 degrees posteriorly, reducing lumbar lordosis and tilting the acetabulum. For every degree of pelvic tilt, it is estimated that the acetabulum undergoes 0.7 to 0.8 degrees of anteversion, allowing the femur to flex freely over the acetabulum without impacting its anterior rim. Therefore, in patients with reduced spine-pelvic movement due to previous arthrodesis or degenerative disease, the acetabulum should theoretically be positioned with more anteversion to compensate for this loss of mobility and avoid anterior femoroacetabular impingement and potential dislocations when sitting [19]-[21]. Similarly, patients with reduced lumbar lordosis require less acetabular anteversion [14] [20]. Some authors advocate the use of sagittal radiographs of the spine and pelvis for preoperative planning as a standardized approach.

In this study, it was demonstrated that out of the 23 patients who experienced dislocation after total hip arthroplasty, only 13 had acetabular angles outside the recommended safe zone. Analyzing articles in the literature and the results obtained in our study, it is evident that acetabular positioning does not constitute an isolated risk factor. It should be considered alongside other patient characteristics, such as age, muscular condition, ligament laxity, spine-pelvic balance, surgical approach, and femoral head size, among others. When determining acetabular positioning, the orthopedic surgeon must carefully evaluate each patient individually to arrive at an angle that allows for better mobility and lower dislocation risk for each case.

Further studies encompassing a larger case volume and incorporating instances where dislocation did not occur post-THA, while individually addressing other potential causes for the occurrence of dislocation, may elucidate the emergence of this undesirable outcome more clearly.

5. Conclusion

Of the 23 patients analyzed who experienced dislocation after total hip arthroplasty, 13 had anteversion and inclination angles outside Lewinnek’s safe zone. The current understanding found in the literature and corroborated by the results of this study is that acetabular positioning cannot be seen in isolation as a determining factor for the occurrence of dislocation. The ideal angle is likely different for each patient, and the surgeon should personalize the planning based on the numerous variables that can influence the risk of postoperative dislocation.

Conflicts of Interest

The authors declare no conflicts of interest regarding the publication of this paper.

References

[1] Ferguson, R.J., Palmer, A.J., Taylor, A., Porter, M.L., Malchau, H. and Glyn-Jones, S. (2018) Hip Replacement. The Lancet, 392, 1662-1671.
https://doi.org/10.1016/S0140-6736(18)31777-X
[2] Campbell, W.C., Canale, S.T., Beaty, J.H., Azar, F.M. and Anderson, R.B. (2021) Campbell’s Operative Orthopaedics. 14th Edition, Elsevier, Amsterdam, 178-333.
[3] Hinton, Z.W., Wu, C.J., Ryan, S.P., Cunningham, D.J., Green, C.L. and Lachiewicz, P.F. (2022) Current Trends in Revision Hip Arthroplasty: Indications and Types of Components Revised. The Journal of Arthroplasty, 37, S611-S615.E7.
https://doi.org/10.1016/j.arth.2022.03.008
[4] Guo, L., Yang, Y., An, B., et al. (2017) Risk Factors for Dislocation after Revision Total Hip Arthroplasty: A Systematic Review and Meta-Analysis. International Journal of Surgery, 38, 123-129.
https://doi.org/10.1016/j.ijsu.2016.12.122
[5] Sadhu, A., Nam, D., Coobs, B.R., Barrack, T.N., Nunley, R.M. and Barrack, R.L. (2017) Acetabular Component Position and the Risk of Dislocation following Primary and Revision Total Hip Arthroplasty: A Matched Cohort Analysis. The Journal of Arthroplasty, 32, 987-991.
https://doi.org/10.1016/j.arth.2016.08.008
[6] Esposito, C.I., Gladnick, B.P., Lee, Y.Y., et al. (2015) Cup Position Alone Does Not Predict Risk of Dislocation after Hip Arthroplasty. The Journal of Arthroplasty, 30, 109-113.
https://doi.org/10.1016/j.arth.2014.07.009
[7] Abdel, M.P., von Roth, P., Jennings, M.T., Hanssen, A.D. and Pagnano, M.W. (2016) What Safe Zone? The Vast Majority of Dislocated THAs Are Within the Lewinnek Safe Zone for Acetabular Component Position. Clinical Orthopaedics and Related Research, 474, 386-391.
https://doi.org/10.1007/s11999-015-4432-5
[8] Biedermann, R., Tonin, A., Krismer, M., Rachbauer, F., Eibl, G. and Stöckl, B. (2005) Reducing the Risk of Dislocation after Total Hip Arthroplasty: The Effect of Orientation of the Acetabular Component. The Bone & Joint Journal, 87, 762-769.
https://doi.org/10.1302/0301-620X.87B6.14745
[9] Bozic, K.J., Kurtz, S.M., Lau, E., Ong, K., Vail, T.P. and Berry, D.J. (2009) The Epidemiology of Revision Total Hip Arthroplasty in the United States. The Journal of Bone & Joint Surgery, 91, 128-133.
https://doi.org/10.2106/JBJS.H.00155
[10] Seagrave, K.G., Troelsen, A., Malchau, H., Husted, H. and Gromov, K. (2017) Acetabular Cup Position and Risk of Dislocation in Primary Total Hip Arthroplasty. Acta Orthopaedica, 88, 10-17.
https://doi.org/10.1080/17453674.2016.1251255
[11] Lewinnek, G.E., Lewis, J.L., Tarr, R., Compere, C.L. and Zimmerman, J.R. (1978) Dislocations after Total Hip-Replacement Arthroplasties. The Journal of Bone & Joint Surgery, 60, 217-220.
https://doi.org/10.2106/00004623-197860020-00014
[12] Kummer, F.J., Shah, S., Iyer, S. and DiCesare, P.E. (1999) The Effect of Acetabular Cup Orientations on Limiting Hip Rotation. The Journal of Arthroplasty, 14, 509-513.
https://doi.org/10.1016/S0883-5403(99)90110-9
[13] Yamaguchi, M., Akisue, T., Bauer, T.W. and Hashimoto, Y. (2000) The Spatial Location of Impingement in Total Hip Arthroplasty. The Journal of Arthroplasty, 15, 305-313.
https://doi.org/10.1016/S0883-5403(00)90601-6
[14] Widmer, K.H. and Zurfluh, B. (2004) Compliant Positioning of Total Hip Components for Optimal Range of Motion. Journal of Orthopaedic Research, 22, 815-821.
https://doi.org/10.1016/j.orthres.2003.11.001
[15] Shon, W.Y., Baldini, T., Peterson, M.G., Wright, T.M. and Salvati, E.A. (2005) Impingement in Total Hip Arthroplasty a Study of Retrieved Acetabular Components. The Journal of Arthroplasty, 20, 427-435.
https://doi.org/10.1016/j.arth.2004.09.058
[16] Nakashima, Y., Hirata, M., Akiyama, M., et al. (2014) Combined Anteversion Technique Reduced the Dislocation in Cementless Total Hip Arthroplasty. International Orthopaedics, 38, 27-32.
https://doi.org/10.1007/s00264-013-2091-2
[17] Yoshimine, F. (2006) The Safe-Zones for Combined Cup and Neck Anteversions That Fulfill the Essential Range of Motion and Their Optimum Combination in Total Hip Replacements. Journal of Biomechanics, 39, 1315-1323.
https://doi.org/10.1016/j.jbiomech.2005.03.008
[18] Hidaka, R., Matsuda, K., Nakamura, M., Nakamura, S. and Kawano, H. (2022) Optimal Combined Anteversion Range for Obtaining a Wider Range of Motion without Prosthetic Impingement after Total Hip Arthroplasty: A Three-Dimensional Analysis Study. Journal of Orthopaedic Surgery and Research, 17, Article No. 226.
https://doi.org/10.1186/s13018-022-03112-6
[19] Tezuka, T., Heckmann, N.D., Bodner, R.J. and Dorr, L.D. (2019) Functional Safe Zone Is Superior to the Lewinnek Safe Zone for Total Hip Arthroplasty: Why the Lewinnek Safe Zone Is Not Always Predictive of Stability. The Journal of Arthroplasty, 34, 3-8.
https://doi.org/10.1016/j.arth.2018.10.034
[20] Heckmann, N.D. and Lieberman, J.R. (2021) Spinopelvic Biomechanics and Total Hip Arthroplasty: A Primer for Clinical Practice. Journal of the American Academy of Orthopaedic Surgeons, 29, e888-e903.
https://doi.org/10.5435/JAAOS-D-20-00953
[21] Ike, H., Dorr, L.D., Trasolini, N., Stefl, M., McKnight, B. and Heckmann, N. (2018) Spine-Pelvis-Hip Relationship in the Functioning of a Total Hip Replacement. The Journal of Bone and Joint Surgery, 100, 1606-1615.
https://doi.org/10.2106/JBJS.17.00403