ABSTRACT

Performance evaluation of computed tomography is not significant in the Republic of Cameroon in general and more particularly in the city of Yaoundé. Therefore, this work aimed to analyze image performances of 05 CT scanners in 05 medical facilities in the city of Yaoundé, by using the Catphan 700 phantom and test procedures of the Atomic Energy Agency. Five quality control tests were performed on each CT-Scanner. All the CT scanners evaluated showed good results and were not disapproved in three of the five tests performed. The difference between the measured spatial resolution for each of the five CT scanners and the spatial resolution specified by the manufacturer was very high. The low contrast resolution of four of the five CT scanners was found low when compared to the low contrast resolution specified by the manufacturers.

Share and Cite:

Njiki, C. , Ndjaka Manyol, J. , Ebele Yigbedeck, Y. , Abou’ou, D. , Yimele, B. and Sabouang, J. (2018) Assessment of Image Quality Parameters for Computed Tomography in the City of Yaoundé. Open Journal of Radiology, 8, 37-44. doi: 10.4236/ojrad.2018.81005.

Cited by

[1] Automated slice thickness measurement on the Nessoft CT QA Phantom
Int. J. Sci. Res. Sci …, 2023
[2] AUTOMATIC MEASUREMENT OF CT NUMBER LINEARITY IN THREE TYPES OF CATPHAN PHANTOMS
Jurnal Teknologi, 2023
[3] Implementación de métricas de calidad de imagen de TC en estudios clínicos SPECT/CT de tórax
… Brasileira de Física …, 2023
[4] CT image quality metrics implementation during thorax SPECT/CT clinic studies
… Brasileira de Física …, 2023
[5] Evaluación “a distancia” de calidad de imagen en Tomografía Computada: validación de una herramienta
Ingeniería Electrónica …, 2023
[6] Evaluación" a distancia" de calidad de imagen en Tomografía Computada: validación de una herramienta/" At distance" computer tomography image quality check …
… Electrónica, Automática y …, 2023
[7] Calidad de la imagen de tomografía computarizada versus dosis en estudios híbridos: resultados preliminares en maniquí.
, 2022
[8] Assessment of Image Quality Parameters for Computed Tomography in Tripoli, Libya
2022 IEEE 2nd International …, 2022
[9] Calidad de la imagen de tomografía computarizada versus dosis en estudios híbridos: resultados preliminares en maniquí
MediSur, 2022
[10] Evaluation of the computed tomography number for water, field uniformity, image noise and contrast resolutions in Kano Metropolis, Nigeria
2020
[11] Efectos adversos por el uso de la biomodulina T y la corticotropina en pacientes con esclerosis múltiple
Medisur, 2007

Copyright © 2025 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.