Background and Purpose: To perform a retrospective in vivo dosimetry study of 129 total body irradiation (TBI) on leukemia and bone marrow transplant patients treated in our clinic from 2008 to 2011 and to find out if there is any indication of the necessity of developing a new efficient TBI approach. Materials and Methods: The in vivo dosimetry data of 129 patients treated with TBI between 2008 and 2011 were retrieved from the database and analyzed. These patients were mostly treated with the regime of a single fraction or 6 fractions with some exceptions of 8-fraction or 2-fraction treatments depending on the protocols that were applied. For every fraction of treatment, 10 pairs of diode dosimeters were used to monitor the doses to the midline of head, neck, arms, mediastinum, left lung, right lung, umbilicus, thigh, knee, and ankle for both AP and PA fields. The doses to the midline of the above body parts were considered to be the average of the AP and PA readings of each diode pair. Dose deviation from the prescribed value for each body part was studied by plotting the histogram of the frequency versus deviation and comparing this with the dose delivered to the midline of the umbilicus to where the dose was prescribed. The correlation of dose deviation to body part thickness was also studied. By studying the dose deviations, we can find the uniformity of general dose distributions for conventional TBI treatments. Results: The retrospective dosimetry study of the 129 TBI patient treatments indicates that for most of the patients treated in our clinic, the doses received by different body parts monitored with in vivo dosimetry were within the window of 10% difference from the prescribed dose. The inhomogeneity of dose on different body parts could be manually improved by using compensators, but the method is cumbersome and time consuming. The dose deviation in many histograms ranging from about ?10% to 10% indicates some incongruity of dose distribution. This could be due to the method of using lead compensators for a manual dose adjustment which could not ideally compensate for different body thicknesses everywhere. Conclusions: The conventional TBI could give uniform dose to the major body parts under the online in vivo dosimetry monitoring at the level of 10%, but the treatment procedure is cumbersome and time consuming. This implies the importance of developing a new and efficient TBI method by adopting modern radiation therapy technique.

Share and Cite:

Lu, L. , Filippi, J. , Patel, A. , Gupta, N. , Weldon, M. , Woollard, J. , Ayan, A. , Rong, Y. , Martin, D. , Pelloski, C. and Welliver, M. (2014) A Clinical Dosimetry Analysis of Total Body Irradiation for Leukemia Patients. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 3, 31-42. doi: 10.4236/ijmpcero.2014.31006.

Cited by

[1] Total body irradiation delivered using a dedicated Co‐60 TBI unit: Evaluation of dosimetric uniformity and dose verification
Journal of Applied …, 2024
[2] Treatment robustness of total body irradiation with volumetric modulated arc therapy
Bosman, A Zoetelief… - Physics and Imaging in …, 2024
[3] Archive of SID. ir
2023
[4] Practice patterns of pediatric total body irradiation techniques: a children's oncology group survey
International Journal of …, 2021
[5] Dosimetric comparison of AP/PA and bilateral geometries for total body irradiation treatment
Radiation and …, 2021
[6] Evaluation of the surface dose for total body irradiation (TBI) technique with parallel-opposed anterior posterior geometry
2021
[7] Comparison of different TBI techniques in terms of dose homogeneity–review study
2021
[8] ارزیابی یکنواختی توزیع دوز در تکنیک پرتودهی فوتونی تمام بدن در هندسه قدامی خلفیِ مقابل-موازی با استفاده از شتاب‌دهنده خطی واریان مدل Clinac 2100C/D.‎
2020
[9] Comparison of calculated and measured basic dosimetric parameters for total body irradiation with 6-and 18-MV photon beams
2020
[10] بررسی مروری تکنیک‌های بهبود یکنواختی توزیع دز در تابش فوتونی تمام بدن (TBI) با هندسه درمان مخالف و موازی قدامی و خلفی‎
مجله سنجش و ایمنی …, 2020
[11] Review of techniques for improving the uniformity of dose distribution in total body irradiation (TBI) with parallel â opposed anterior and posterior geometry
Radiation Safety and …, 2020
[12] Caracterização de dosímetros semicondutores para dosimetria in Vivo na técnica de TBI
2018

Copyright © 2025 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.