In-Vivo Dosimetry Method for Measuring Peak Surface Dose Using Radiochromic Films during Computed Tomography Scanning of the Sinus

HTML  XML Download Download as PDF (Size: 566KB)  PP. 151-159  
DOI: 10.4236/ijmpcero.2018.72013    578 Downloads   1,475 Views  Citations

ABSTRACT

Purpose: During computed tomography (CT) helical scanning mode the patient surface dose distribution is assumed to be non-uniform, therefore point dose measurement methods may lead to imprecise estimation of the radiation dose received by the patient skin in particular. We have used XRQA2 films as in-vivo dosimeters to measure the entrance skin dose during sinus exams. Methods: The films were placed under the patient head rest in order to sample the entrance surface dose in-vivo. We have performed in-vivo film irradiation on 23 patients in this study to verify the clinical suitability of the method and were found adequate. Results: The measured average ESD in the sinus exam was 11.7 ± 1.0 mGy, the PSD was 15.7 ± 1.7 mGy and the CTDI(vol) was 13.3 ± 0.1 mGy. The ratio of ESD/CTDI(vol) and PSD/CTDI(vol) was 0.88 and 1.18 respectively. The results indicate that the scanner registered CTDI(vol) underestimates the PSD and in the same time it overestimates the ESD by 18% and 13.6% respectively. Conclusion: The observed differences between the ESD, PSD and CTDI(vol) although seem small for the radiation dose range measured during CT of the sinus [13.2 - 13.4] mGy, but important for the medical physicist to know, since monitoring of patients’ doses from CT examinations is becoming more mandatory. The use of radiochromic film as in-vivo dosimeter does not interfere with the clinical radiological exam and does not produce any image artifacts. The method can be used to study other CT examinations specially the ones with large beam width, high pitch factor and high dose exams. The method allows measurement of the peak skin dose, examination of the CT dose profile and the 2D dose distribution in the XZ plan.

Share and Cite:

Soliman, K. , Altimyat, S. , Alrushoud, A. , Alenezi, A. and Alkhorayef, M. (2018) In-Vivo Dosimetry Method for Measuring Peak Surface Dose Using Radiochromic Films during Computed Tomography Scanning of the Sinus. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 7, 151-159. doi: 10.4236/ijmpcero.2018.72013.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.