Open Journal of Anesthesiology

Volume 5, Issue 11 (November 2015)

ISSN Print: 2164-5531   ISSN Online: 2164-5558

Google-based Impact Factor: 0.23  Citations  

Forearm Loss Caused by Automated Non-Invasive Blood Pressure Cuff Malfunction: A Hearsay Report

HTML  XML Download Download as PDF (Size: 249KB)  PP. 227-232  
DOI: 10.4236/ojanes.2015.511041    4,446 Downloads   5,438 Views  

ABSTRACT

Failure of an automated blood pressure cuff to deflate when a patient is under general anesthesia can lead to catastrophic consequences if unnoticed for more than three hours [1]. We present this as a hearsay case in which an automated blood pressure cuff of the Spacelabs Ultraview Clinical Workstation monitor (model No. 90385) applied pressure for about five hours resulting in limb thrombosis. In order to analyze this catastrophe, simulation scenarios were tested to elucidate the possible errors and malfunctions that may have led to this injury. We present the analysis of the advantages and validity of the hearsay case report. We also include our proposed criteria that should be required when a hearsay case is considered for publication.

Share and Cite:

Shulman, S. , Namn, Y. , Lando, S. and Discepola, P. (2015) Forearm Loss Caused by Automated Non-Invasive Blood Pressure Cuff Malfunction: A Hearsay Report. Open Journal of Anesthesiology, 5, 227-232. doi: 10.4236/ojanes.2015.511041.

Cited by

No relevant information.