ABSTRACT

Purpose: The isocenter of a medical linac system is a frequently used concept in clinical practice. However, so far not all the isocenters are rigorously defined. This work is intended as an attempt of deriving consistent and operable isocenter definitions. Methods: The isocenter definition is based on a fundamental concept, the axis of rotation of a rigid body. The axis of rotation is determined using the trajectory of any point on a plane that intersects the rigid body. A point on the axis of rotation is found through the minimal bounding sphere of the trajectory when the rigid body makes a full rotation. The essential mathematical tool of the isocenter definition system is three-dimensional coordinate transformation. Results: The axes of rotation of the linac collimator, gantry, and couch are established first. The linac mechanical isocenter (linac isocenter) is defined as the center of a circle that best fits the trajectory of a select linac X-ray source position. The axis of rotation and the minimal bounding sphere are cornerstones for the rotation isocenters of the collimator, gantry and couch. The definition of radiation isocenter incorporates a surrogate of the useful beam axis. Conclusions: A framework of isocenter definitions for medical linacs is presented in this manuscript. Consistent meanings of the mechanical and radiation isocenters can be achieved using this approach.

Share and Cite:

Zhang, M. , Zhou, S. and Qu, T. (2015) What Do We Mean When We Talk about the Linac Isocenter?. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 233-242. doi: 10.4236/ijmpcero.2015.43028.

Cited by

[1] Fast Isocenter Determination Using 3D Polymer Gel Dosimetry with Kilovoltage Cone-Beam CT Reading and the PolyGeVero-CT Software Package for Linac Quality …
Materials, 2022
[2] Biological Effects of Intense Terahertz Pulses
2021
[3] CADTH Health Technology Review Implementation of Stereotactic Ablative Radiotherapy for the Treatment of Oligometastatic Cancer in Canada
2021
[4] Implementation of Stereotactic Ablative Radiotherapy for the Treatment of Oligometastatic Cancer in Canada
2021
[5] A novel method to determine linac mechanical isocenter position and size and examples of specific QA applications
2021
[6] Characterization of motion of the LINAC flat-panel detector
2020
[7] Approximating a Classic Cone-Based Winston-Lutz Test Using a Continuously Rotating Collimator Concurrent with Beam Delivery and Its Applicability for …
2020
[8] Assessment of an Elekta Versa HD linear accelerator for stereotactic radiosurgery with circular cone collimators
2020
[9] Teaching and learning threshold concepts in radiation physics for professional practice
2020
[10] Approximating a Classic Cone-Based Winston-Lutz Test Using a Continuously Rotating Collimator Concurrent with Beam Delivery and Its Applicability for Single …
2020
[11] Setup error en radioterapia de próstata con corrección online
2019
[12] MapPHAN for conformal arc SABR: an analysis of normal tissue dosimetry
2019
[13] STARU TERAPIJAS AR LĒZENIEM DOZAS GRADIENTIEM VERIFICĒŠANA
2019
[14] A Quality Assurance Approach for Linear Accelerator Mechanical Isocenters with Portal Images
2018
[15] Threshold Concepts in Radiation Physics Underpinning Professional Practice in Radiation Therapy
2018
[16] Comparison of Commercial Quality Assurance Programs For Radiation Isocenter Checks Of Linacs
2016
[17] The Measurement Accuracy of Ball Bearing Center in Portal Images Using an Intensity-Weighted Centroid Method
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2015
[18] Image based mechanical isocenter runout measurement

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.