Detection of Stiff Nodules Embedded in Soft Tissue Phantoms, Mimicking Cancer Tumours, Using a Tactile Resonance Sensor

HTML  Download Download as PDF (Size: 1596KB)  PP. 181-193  
DOI: 10.4236/jbise.2014.74022    4,375 Downloads   6,291 Views  Citations

ABSTRACT

Background: Prostate cancer (PCa) is the most common form of cancer among males in Europe and in the USA and the most common curative treatment is removal of the prostate, i.e. prostatectomy. After the removal, the prostate is histopathologically analysed. One area of interest is to examine the perifery of the prostate, as tumours on and near the surface can indicate that the PCa has spread to other parts of the body. There are no current methods to examine the surface of the prostate at the time of surgery. Tactile resonance sensors can be used for detecting areas of different stiffness in soft tissue. Human prostate tissue affected by cancer is usually stiffer than healthy tissue, and for this purpose, a tactile resonance sensor was developed. The aim of this study was to investigate the depth at which embedded stiffer volumes could be detected, using soft tissue phantoms. Methods: With the tactile resonance sensor used in this study, the shift of the resonance frequency and the force at contact with tissue can be measured, and combined into a tissue stiffness parameter. The detection sensitivity of the sensor at impression depths, 0.4 and 0.8 mm, was measured for detection of inserted nodules of stiff silicone in softer silicone and in chicken muscle tissue, mimicking prostate tissue with cancer tumours. Results: Measurements on the silicone samples detected the hidden stiffer object at a depth of 1 - 4 mm with a difference in the stiffness parameter of 80 - 900 mN/kHz (p < 0.028, n = 48). At the depth 5 - 6 mm the difference was smaller but still significant < 30 mN/kHz (p < 0.05, n = 24). For the measurements on chicken muscle, the detectable depth was 4 mm (p < 0.05, n = 24). Conclusion: This model study suggests that, with only a small impression depth of ≤1 mm, the resonance sensor system described here can detect stiffness variations located at least 4 mm in silicone and chicken muscle, mimicking tumours in prostate tissue.

Share and Cite:

Åstrand, A. , Jalkanen, V. , Andersson, B. and Lindahl, O. (2014) Detection of Stiff Nodules Embedded in Soft Tissue Phantoms, Mimicking Cancer Tumours, Using a Tactile Resonance Sensor. Journal of Biomedical Science and Engineering, 7, 181-193. doi: 10.4236/jbise.2014.74022.

Cited by

[1] Liquid Metal Enabled Biodevices
Advanced Intelligent …, 2021
[2] Mechanical Imaging of Soft Tissues with a Highly Compliant Tactile Sensing Array
2017
[3] Prostate Cancer Detection with a Tactile Resonance Sensor—Measurement Considerations and Clinical Setup
Sensors, 2017
[4] Piezoelectric tactile sensor for submucosal tumor detection in endoscopy
Sensors and Actuators A: Physical, 2016
[5] Towards Bio-Inspired Tactile Sensing Capsule Endoscopy for Detection of Submucosal Tumours
2016
[6] Tumor Detection and Morphology Assessment in the Liver Tissue Using an Automatic Tactile Robot
2016
[7] Optimization of hyperelastic model parameters of soft tissue based on genetic algorithm utilizing experimental mechanical datase
2015
[8] Dual-modality probe intended for prostate cancer detection combining Raman spectroscopy and tactile resonance technology—discrimination of normal human …
Analytical Chemistry Letters, 2015
[9] Initial Measurements on Whole Human Prostate ex vivo with a Tactile Resonance Sensor in Order to Detect Prostate Cancer
16th Nordic-Baltic Conference on Biomedical Engineering. Springer International Publishing, 2015
[10] Dynamic modeling of breast tissue with application of model reference adaptive system identification technique based on clinical robot-assisted palpation
Journal of the mechanical behavior of biomedical materials, 2015
[11] بهینه‌سازی ضرایب مدل هایپرالاستیک بافت نرم با استفاده از داده های تجربی و الگوریتم ژنتیک
Modares Mechanical Engineering, 2015
[12] Investigating reservoir's initial oil saturation effect on in-situ combustion recovery method performance using numerical simulation
2015
[13] Optimization of hyperelastic model parameters of soft tissue based on genetic algorithm utilizing experimental mechanical dataset
Modares Mechanical Engineering, 2015

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.