An in Vitro Evaluation of Pressure Generated by Programmed Intermittent Epidural Bolus (PIEB) or Continuous Epidural Infusion (CEI)

HTML  Download Download as PDF (Size: 791KB)  PP. 214-217  
DOI: 10.4236/ojanes.2013.34049    5,866 Downloads   8,893 Views  Citations

ABSTRACT

PIEB has been reported as being superior to CEI for labor analgesia. The aim of this study was to measure the pressures generated by two commercially available pumps (CADD PIB, Smiths and Gemstar, Hospira) when delivering PIEB or CEI at 2 commonly used rates of infusion (10 or 5 mL). The two pumps were set to deliver fluid at four rates (CEI: 5 or 10 mL/h; PIEB: 5 or 10 mL every h) and connected to a pressure transducer (PCB Piezotronics 1500 connected to a NI USB-6251 Screw Terminal) to determine the pressures applied during each infusion. The peak pressure generated during the PIEB mode was consistently higher when compared to the CEI mode in both pumps. When comparing the two pumps in the PIEB 10 mL every hour mode, the peak pressures were approximately the same, while the pressure pattern differed. For each cycle the pressure generated by Gemstar oscillated from 25 mmHg to 0 mmHg; whereas with the CADD pump the baseline pressure was above 0. The Gemstar pump piston frequency (1.2 Hz) was twice as high as the CADD (0.5 Hz), so the volume delivered per cycle was lower. In the PIEB 5 mL every hour mode the peak pressures followed the same wave patterns. However, unlike the 10 mL mode, the pump piston frequency was approximately the same in both pumps. In both the CEI 5 or 10 mL/h modes, the CADD peak pressure was four times greater than the Gemstar. Both cycled between their respective peak pressures and 0 mmHg. Since the peak pressure of the CADD was approximately the same with PIEB and CEI, we speculated that the most important factor affecting the distribution of the solution in the epidural space was not the peak pressure per se, but the mode of delivery of the bolus.

Share and Cite:

S. Stirparo, S. Fortini, S. Espa, A. Cenedese and G. Capogna, "An in Vitro Evaluation of Pressure Generated by Programmed Intermittent Epidural Bolus (PIEB) or Continuous Epidural Infusion (CEI)," Open Journal of Anesthesiology, Vol. 3 No. 4, 2013, pp. 214-217. doi: 10.4236/ojanes.2013.34049.

Cited by

[1] Estudio comparativo de diferentes sets de infusión epidural a velocidades máximas para analgesia del parto
2020
[2] Comparative study of different epidural infusion sets at maximum speeds for labor analgesia
2020
[3] Distribution of a Solution in the Epidural Space
2020
[4] Estudio comparativo de la combinación de diferentes modos de administración de anestésicos locales en la analgesia del parto
2020
[5] Comparaison du taux d'intervention médicale chez des parturientes au cours du travail obstétrical lorsque l'entretien analgésique péridural est réalisé par …
2019
[6] Programmed intermittent epidural boluses (PIEB) in labor analgesia: a narrative review of the present
Acta …, 2019
[7] Epidural contrast spread in the porcine model: bolus versus infusion
2019
[8] Comparaison du taux d'intervention médicale chez des parturientes au cours du travail obstétrical lorsque l'entretien analgésique péridural est réalisé par boli …
2019
[9] Continuous epidural infusion versus programmed intermittent epidural bolus for labor analgesia: optimal configuration of parameters to reduce physician-administered …
Ships and Offshore Structures, 2017
[10] An in vitro evaluation of the pressure generated during programmed intermittent epidural bolus injection at varying infusion delivery speeds
Journal of Clinical Anesthesia, 2016
[11] Obstetric analgesia–update 2016
Journal of Perinatal Medicine, 2016
[12] Maintenance of Labor Analgesia
Epidural Labor Analgesia, 2015
[13] Estudo comparativo entre três volumes de Ropivacaína 1mg-mL e Sufentanil 0, 25mcg-mL por via epidural em bolus intermitentes e sua relação com a duração da …
2013
[14] Techniques for the maintenance of epidural labor analgesia
Current Opinion in Anesthesiology, 2013
[15] Estudo comparativo entre três volumes de Ropivacaína 1mg-mL e Sufentanil 0, 25mcg-mL por via epidural em bolus intermitentes e sua relação com a duração da analgesia durante o trabalho de parto
2013

Copyright © 2025 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.