Validating HeartSmart® against the Cardiopulmonary Bypass Machine

HTML  XML Download Download as PDF (Size: 438KB)  PP. 21-27  
DOI: 10.4236/ss.2012.31004    7,111 Downloads   9,991 Views  Citations

ABSTRACT

Purpose: To assess the utility of HeartSmart®, a new method that uses empirical physiological formulae to estimate hemodynamic variables, in estimating cardiac index during open heart surgery when compared with the cardiopulmonary bypass machine. Methods: This was a prospective, blinded study of patients undergoing elective cardiac bypass surgery. We monitored and compared the flow rates provided by the cardiopulmonary bypass machine (roller pump) with estimates derived from the empirical physiological formulae embedded in the HeartSmart® monitor in 32 patients. Cardiac index values were compared at the start of cardiopulmonary bypass, during re-warming, and at the end of cardiopulmonary bypass. Results: A total of 256 paired sets of measurements were suitable for comparison. The mean of the differences or bias (95% limits of agreement) was 0.09 l/min/m2 and the limits of agreement –0.86 to 1.05 l/min/m2. The mean difference of the sets of 256 measurements was 585.5 l/min/m for the pump and 575.0 l/min/m2 for the empirical physiological formulae—a difference of 5% l/min/m2. The range of flow rates for the pump was 1.2 to 2.85 l/min/m2; for the empirical physiological formulae, the range was 1.2 to 3.0 l/min/m2. Conclusion: The cardiac index estimates derived from the empirical physiological formulae in the HeartSmart® software are in good agreement with pump cardiac output rates. These results suggest that HeartSmart® measurements are sufficiently accurate for assessing hemodynamic variables in many groups of patients.

Share and Cite:

K. Warring-Davies and M. Bland, "Validating HeartSmart® against the Cardiopulmonary Bypass Machine," Surgical Science, Vol. 3 No. 1, 2012, pp. 21-27. doi: 10.4236/ss.2012.31004.

Cited by

[1] Bedside Hemodynamic Monitoring Providing Equity of Critical and Maternity Care for the Critically Ill Pregnant or Recently Pregn
JOURNAL OF CARDIOVASCULAR DISEASE, 2014
[2] A New Simple Empirical Formula for Calculating Body Surface Area in Children & Adults
Frontiers in Clinical Medicine (FCM), 2014
[3] Bedside Hemodynamic Monitoring Providing Equity of Critical and Maternity Care for the Critically Ill Pregnant or Recently Pregnant Woman
Journal of Cardiovascular Disease, 2014
[4] Cardiac Dynamic Monitoring in Patients with Systemic Inflammatory Response Syndrome: a Double-Blind Study
Journal of Cardiovascular Disease, 2013
[5] HeartSmart (TM): A New Method of Assessing Hydration in Neurosurgical Patients
Surgical Science, 2012
[6] HeartSmart™: A new method of assessing hydration in neurosurgical patients
2012
[7] HeartSmartTM: A New Method of Assessing Hydration in Neurosurgical Patients
Surgical Science, 2012

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.