Sequential Injection Analysis of Ampicillin in Pharmaceuticals by Using Potentiometric Detectors Based on PVC and Sol-Gel Membranes

Abstract

This work describes the construction and the evaluation of the general performance of new ampicilli-nate-selective electrodes based on manganese (III) tetraphenylporphyrin [Mn(III) TPP-Cl] as ionophore, incorporated in both PVC and sol-gel membranes and directly applied onto a conductive graphite/epoxy resin support. The units were constructed without inner reference solution adopting conventional configuration and in the case of PVC membrane the tubular configuration was also adopted. The good working characteristics of these electrodes, made possible its application to the determination of ampicillin in pharmaceuticals formulations, both in conventional batch analysis and in flow conditions, when the electrodes were coupled to a SIA system. In the last case the potentiometric sensors presented linear response towards ampicillin concentration between 5.0 × 10–4 and 5.0 × 10–2 mol?l–1 with slopes of –57.4 and –63.5 mV?dec–1 for the PVC and sol-gel membranes, respectively. The developed procedures enable mean relative standard deviations better than 3% for all the samples analysed. The obtained results do not statistically differ from those furnished by applying the HPLC reference method.

Share and Cite:

N. Zárate, A. Araujo, M. Montenegro and R. Pérez-Olmos, "Sequential Injection Analysis of Ampicillin in Pharmaceuticals by Using Potentiometric Detectors Based on PVC and Sol-Gel Membranes," American Journal of Analytical Chemistry, Vol. 2 No. 4, 2011, pp. 491-499. doi: 10.4236/ajac.2011.24059.

Conflicts of Interest

The authors declare no conflicts of interest.

References

[1] E. Malinowska, J. Niedziólka, E. Rozniecka and M. E. Meyerhoff, “Salicylate-Selective Membrane Electrodes Based on Sn(IV) and O=Mo(V)-Porphyrin: Differences in Response Mechanism and Analytical Performance,” Journal of Electroanalytical Chemistry, Vol. 514, No. 1-2, 2001, pp. 109-117. doi:10.1016/S0022-0728(01)00615-5
[2] E. M. G. Santos, A. N. Araújo, C. M. C. M. Couto, M. C. B. S. M. Montenegro, A. Kejzlarová and P. Solich, “Ion Selective Electrodes for Penicillin-G Based on Mn (III) TPP-Cl and Their Application in Pharmaceutical Formu-lations by Sequential Injection Analysis,” Journal of Pha- rmaceutical and Biomedical Analysis, Vol. 36, No. 4, 2004, pp. 701-709. doi:10.1016/j.jpba.2004.08.006
[3] T. Katsu, K. Ido, A. Moriya, Y. Nakae and I. Sakata, “Valproate-Selective Membrane Electrode Based on Gal-lium(III)Tetraphenylphorfyrin,” Electroanalysis, Vol. 12, No. 16, 2000, pp. 1282-1285. doi:10.1002/1521-4109(200011)12:16<1282::AID-ELAN1282>3.0.CO;2-D
[4] E. M. G. Santos, A. N. Araújo, C. M. C. M. Couto and M. C. B. S. M. Montenegro, “Construction and Evaluation of PVC and Sol-Gel Sensor Membranes Based on Mn(III) TPP-Cl. Aplication to Valproate Determination in Phar-maceutical Preparation,” Analytical and Bioanalytical Chemistry, Vol. 384, No. 4, 2006, pp. 867-875. doi:10.1007/s00216-005-0170-y
[5] S. S. M. Hassan, W. H. Mahmoud, M. A. F. Elmosallamy and M. H. Almarzooqi, “Nobel Ibuprofen Potentiometric Membrane Sensors Based on Tetraphenylphorfyrinato Indium(III),” Analytical Sciences, Vol. 19, No. 5, 2003, pp. 675-680. doi:10.2116/analsci.19.675
[6] J. H. A. Badr, M. E. Meyerhoff and S. S. M. Hassan, “Metallophorfyrin-Based Polymer Membrane Electrode with High Selectivity for 2-Hydroxybenzhydroxamate,” Analytica Chimica Acta, Vol. 321, No. 1, 1996, pp. 11-19. doi:10.1016/0003-2670(95)00580-3
[7] E. M. G. Santos, A. N. Araújo, C. M. C. M. Couto and M. C. B. S. M. Montenegro, “Potentiometric Behaviour of Ion Selective Electrodes Based on Iron Porphyrins: The Influence of Porphyrins Substituents on the Response Properties and Analytical Determination of Diclofenac in Pharmaceutical Formulations,” Journal of Pharmaceutical and Biomedical Analysis, Vol. 42, No. 5, 2006, pp. 535-542. doi:10.1016/j.jpba.2006.05.019
[8] O. Dinten, U. E. Spichiger, N. Chaniotakis, P. Gehring, B. Rusterholtz, W. E. Morf and W. Simon, “Lifetime of Neutral-Carrier-Based Liquid Membranes in Aqueous Samples and Blood and the Lipophilicity of Membrane Components,” Analytical Chemistry, Vol. 63, No. 6, 1991, pp. 596-603. doi:10.1021/ac00006a009
[9] J. Lin and C. W. Brown, “Sol-Gel Glass as a Matrix for Chemical and Biochemical Sensing,” Trends in Analytical Chemistry, Vol. 16, No. 4, 1997, pp. 200-211. doi:10.1016/S0165-9936(97)00021-6
[10] W. Kim, S. Chung, S. B. Park, S. Ch. Lee, C. Kim and D. D. Sung, “Sol-Gel Method for the Matrix of Chlo-ride-Selective Membranes Doped with FridodecylMethy-lammonium Chloride,” Analytical Chemistry, Vol. 69, 1997, pp. 95-98. doi:10.1021/ac9604088
[11] W. Kim, D. D. Sung, G. S. Cha and S. B. Park, “Chlo-ride-Selective Membranes Prepared with Different Ma-trices Including Polymers Obtained by the Sol-Gel Me-thod,” Analyst, Vol. 123, 1998, pp. 379-382. doi:10.1039/a705788a
[12] M. Mazloum, M. Salavati-Niasari, M. Khayat and S. M. Ghoreishi, “A Copper Ion-Selective Electrode with High Selectivity Prepared by Sol-Gel Coated Wire Techniques,” Analytical and Bioanalytical Chemistry, Vol. 378, No. 6, 2004, pp. 1659-1665. doi:10.1007/s00216-003-2462-4
[13] S. Z. Yao, J. Shiao and L. H. Nie, “Potentiometric De-termination of Penicillins with Ion-Selective Electrodes,” Talanta, Vol. 36, No. 12, 1989, pp. 1249-1252. doi:10.1016/0039-9140(89)80058-X
[14] E. G. Kulapina, V. V. Baraguzina and O. I. Kulapina, “Ion-Selective Electrodes for Determining Penicillin An-tibiotics in Biological Fluids and Pharmaceutical Forms,” Journal of Analytical Chemistry, Vol. 59, No. 9, 2004, pp. 971-975. doi:10.1023/B:JANC.0000040704.24966.46
[15] M. S. Rizk, Y. M. Issa and A. F. Shoukry, “New Ampi-cillin Selective Plastic Membrane and Coated Metal Electrodes Based on Ampicillinium Phosphotungstate Ion Pair,” Analytical Letters, Vol. 27, No. 6, 1994, pp. 1055- 1065.
[16] N. V. Shvedene and S. V. Borovskaya, “Determination of β-Lactam Antibiotics by Potentiometry with Ion-Sele- ctive Electrodes,” Journal of Analytical Chemistry, Vol. 58, No. 11, 2003, pp. 1208-1213. doi:10.1023/A:1027393608302
[17] G. J. Papariello, A. K. Muhkerji and C. M. Shearer, “Pe-nicillin Selective Enzyme Electrode,” Analytical Chemistry, Vol. 45, No. 4, 1973, pp. 790-792. doi:10.1021/ac60326a032
[18] I. S. Park, D. K. Kim and N. Kim, “ Characterization and Food Application of a Potentiometric Biosensor Measur-ing β-Lactam Antibiotics,” Journal of Microbiology and Biotechnology, Vol. 14, No. 4, 2004, pp. 698-706.
[19] E. S. Gil and G. R. Melo, “Electrochemical Biosensors in Pharmaceutical Analysis,” Brazilian Journal of Pharma-ceutical Sciences, Vol. 46, No. 3, 2010, pp. 375-391.
[20] K. Gupta, et al., “Recent Advances on Potentiometric Membrane Sensors for Pharamaceutical Analysis,” Com-binatorial Chemistry & High Throughput Screening, Vol. 14, No. 4, 2011, pp. 284-302.
[21] S. Alegret, J. Alonso, J. Bartrolí, A. A. S. C. Machado, J. L. F. C. Lima and J. M. Paulis, “Construction of Equip-ment for Potentiometric Determinations in Flow Analy- sis,” Química Analítica, Vol. 6, No. 3, 1987, pp. 278-292.
[22] J. Alonso, J. Bartrolí, J. L. F. C. Lima and A. A. S. C. Machado, “Sequential Flow-Injection Determinations of Calcium and Magnesium in Waters,” Analytica Chimica Acta, Vol. 179, 1986, pp. 503-508. doi:10.1016/S0003-2670(00)84500-6
[23] J. L. F. C. Lima, M. C. B. S. M. Montenegro and A. M. Roque da Silva, “A Phenobarbital-Selective Electrode without an Inner Reference Solution, an Its Application to Pharmaceutical Analysis,” Journal of Pharmaceutical and Biomedical Analysis, Vol. 8, 1990, pp. 701-704. doi:10.1016/0731-7085(90)80106-Y
[24] J. L. F. C. Lima, M. C. B. S. M. Montenegro and A. M. Roque da Silva, “FIA Potentiometric Determination of Salicylate in Pharmaceutical Preparations with a Tubular Detector,” Journal of Flow Injection Analysis, Vol. 7, 1990, pp. 19-33.
[25] European Directorate for the Quality of Medicines, “Di-rectorate for the Quality of Medicines of the Council of Europe,” European Pharmacopoeia, 4th Edition, 2002.
[26] IUPAC, Pure and Applied Chemistry, Vol. 53, No. 10, 1981, pp. 1907-1912. doi:10.1351/pac198153101907
[27] E. M. G. Santos, C. M. C. M. Couto, M. C. B. S. Monte-negro, M. G. P. M. S. Neves, S. L. H. Rebelo, J. A. S. Cavaleiro and B. F. Reis, “Ion-Selective Electrodes Based on Metalloporphyrins for Gibberellic Acid Determination in Agricultural Products,” Analytical and Bioanalytical Chemistry, Vol. 375, No. 4, 2003, pp. 511-516.
[28] A. Caneiro, P. Fabry, H. Khireddine and E. Siebert, “Performance Characteristics of Sodium Superionic Con- ductor Prepared by the Sol-Gel Route for Sodium Sen-sors,” Analytical Chemistry, Vol. 63, No. 22, 1991, pp. 2550-2557. doi:10.1021/ac00022a004
[29] U. Oesch, D. Amman, H. Pham, U. Wuthier, R. Zund and W. J. Simon, “Design of Anions-Selective Membranes for Clinically Relevant Sensors,” Chemical Society Faraday Transactions I, Vol. 82, No. 4, 1986, pp. 1179- 1186. doi:10.1039/f19868201179
[30] A. M. Pimenta, A. N. Araújo and M. C. B. S. M. Monte-negro, “A Sequential Injection Analysis System for Po-tassium Clavulanate Determination Using Two Potenti-ometric Detectors,” Journal of Pharmaceutical and Bio-medical Analysis, Vol. 30, No. 4, 2002, pp. 931-937. doi:10.1016/S0731-7085(02)00474-0
[31] J. C. Miller and J. N. Miller, “Statistics for Analytical Chemistry,” 2nd Edition, Ellis Horwood Ltd., Chichester, 1988.
[32] Analytical Methods Committee, “Recommendations for the Definition, Estimation, and Use of Detection Limit,” Analyst, Vol. 112, 1987, pp. 199-204. doi:10.1039/an9871200199

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.