Rapid Determination of Residual Quinolones in Honey Samples by Fast HPLC with an On-Line Sample Pretreatment System

HTML  Download Download as PDF (Size: 611KB)  PP. 200-205  
DOI: 10.4236/ajac.2011.22023    5,726 Downloads   10,206 Views  Citations

ABSTRACT

A method of on-line pretreatment coupled to HPLC with fluorescence detection was developed and validated for the determination of nine quinolones in honey samples. This method simplified the complicated process of sample pretreatment and reduced sample treatment time. Recovery of the quinolones was between 92% - 101% for spiked honey samples. The limit of detection was 0.22 - 3.78 ng/mL (signal/noise ratio = 3). There was good linear correlation between HPLC peak area and concentration of the quinolones, with a linear range of 1.0 - 100.0 ng/mL and correlation coefficients >0.9997. The method proposed was validated for detecting quinolones in honey samples.

Share and Cite:

W. Du, J. Yao, Y. Li and Y. Hashi, "Rapid Determination of Residual Quinolones in Honey Samples by Fast HPLC with an On-Line Sample Pretreatment System," American Journal of Analytical Chemistry, Vol. 2 No. 2, 2011, pp. 200-205. doi: 10.4236/ajac.2011.22023.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.