Study on Backward Scattering Characteristics of Submicron Particles ()
1. Introduction
Submicron particles are widely present in production and daily life [1] [2] [3] [4] [5]. For example, the line width and pitch of large-scale integrated circuits are very small, which puts forward higher requirements on the production technology of the circuit and its manufacturing environment. Not only should submicron or even nano lines be etched, but circuit defects must be controlled to a certain range in order to ensure the function and yield of the chip. However, the particulate matter generated in the production process and the particulate matter in the production environment will make the integrated circuit lose its proper function and become a waste product. Therefore, we need real-time detection of submicron particles in the environment. In this paper, based on the generalized Mie scattering theory [6] [7], the relationship between backscattered light energy and particle size is obtained through numerical simulation calculations. The relationship between the backlight scattering energy and particle size and particle concentration was studied based on the backlight scattering particle testing technology.
2. Backscatter Calculation Principle
The experimental system for measuring the backscattering characteristics of particles is shown in Figure 1. The laser light emitted by the laser is irradiated by the Gaussian beam condensed by the transflective mirror and lens combination into the flowing sample cell. The backscattered light of the particles passes through the same lens combination and transflective mirror to be accepted and transformed into electrical signals by the detector. Finally, the electrical signals are sent to the computer for processing.
Figure 2 is the calculated coordinates of the spatial distribution of scattered light of particles based on the generalized Mie theory [8] [9] [10] (abbreviated as GLMT), where
is the beam coordinate system, the beam propagates along the w direction, and the gaussian beam waist center is OG. The coordinate
Figure 1. Schematic diagram of backward light scattering particles.
Figure 2. Model for calculating the spatial distribution of scattered light from particles.
system
is the particle coordinate system, the origin of the coordinate system is at the particle center, the z-axis is parallel to the w axis.
According to the calculation model shown in Figure 2, in the actual calculation, it is necessary to set a position of the particles in the beam coordinate system; set the integration range
and
of the detection point. Calculate the scattered light energy within this integration range, namely:
(1)
among them
(2)
In the above formula,
and
are scattering intensity functions, which can be calculated by GLMT.
3. Relationship between Particle Size and Backscattered Light Energy
The backscattering characteristics of the particles are closely related to the particle size. With different particle sizes, the backscattered light energy is significantly different. Based on the principle of backscattering of particles, a program written by Monte Carlo algorithm is used to calculate the backscattered light energy of the particles, and the corresponding relationship between the scattered light energy of the particles and the particle size is obtained by changing the parameters. Set the beam waist radius of the Gaussian beam to be
, the wavelength to be
, and the refractive index is 1.57, change the particle size. From the simulation calculations, it can be known that the backscattered light energy has a linear relationship of linear micro-oscillation on the logarithmic coordinate as the particle size changes. The line relationship is shown in Figure 3.
Figure 3. Relationship between particle backscattered light energy and particle size.
4. Experimental Research
The experimental device is built according to the experimental principle of Figure 1, in which the amplifier circuit adopts the two-stage amplification mode of the LT1793 op-amp in series with LT1012. By adjusting the resistance ratio of the chip input terminal to control the amplification factor, the input optical signal can be converted into a current signal and then into a voltage signal. The space where the liquid flows inside the sample cell is a rectangular parallelepiped, in which the height of the rectangular parallelepiped is 150 mm, and the cross-section is a square with a side length of 10 mm. The circulating pump system adopts BT800 type peristaltic pump, and the circulating speed is adjustable from 500 rev/min to 2000 rev/min. The signal acquisition software is the top view 2000 oscilloscope simulation system of Sichuan Tuopu Measurement & Control Technology Co., Ltd. It adopts the internal trigger mode and sets the trigger voltage to 0 V - 10 V.
The standard polystyrene particles with radii of 0.1 μm, 0.2 μm, 0.3 μm, 0.5 μm were measured by the above experimental device, and the pulse waveform signal was obtained. Before putting the particles to be tested, start the peristaltic pump, adjust the photoelectric detector, and reduce the noise.
Figure 4 shows the pulse waves of these four particles measured in the laboratory and Gaussian fitting the waveform with origin. Figure 5 shows the relationship between the peak values of these four-particle pulse waveforms and particle size, and it is found that the linear relationship is basically consistent with Figure 3.
5. Relationship between Backscatter Signal and Particle Concentration
Under the same experimental conditions, particle samples were gradually added to the cyclic sampling system. Figure 6 shows the experimental results of dropping 2 drops, 3 drops and 4 drops of particles with a particle size of 0.3. From the figure, we can see that the number of particle backscattering response signals detected increases with increasing concentration during the same period. The higher the concentration in the same period of time, the more particles pass through the measurement area.
6. Statistics of Particle Light Scattering Pulse Signal Count Distribution
In Figure 4, the scattered light energy of different particles is the maximum value selected from a large amount of data obtained in the experiment, but this does not guarantee that the particles pass through the center of the beam waist, so a large amount of data is required to count the relationship between the light scattering pulse signal and the particles. During the experimental measurement, the particles in the liquid passing through the Gaussian beam waist area are random. Therefore, the amplitude of the backscattered pulse signal generated by
Figure 5. Relationship between particle size and its peak.
Figure 6. Scattered signals at different concentrations. (a) 2 drops of particles; (b) 3 drops of particles; (c) 3 drops of particles.
the particles is also random, and the amplitude of these pulse signals contains the inherent characteristics of the interaction between the particles and the beam (such as particle size, particle refractive index, etc.).
In this paper, we try to fit the logarithmic normal distribution to the amplitude of the experimental data, equation [11] is as follows:
,
(3)
where
represents the classification interval of voltage, V represents the voltage value, and
represents the probability value between V and
.
Under the same experimental device conditions, random measurement is performed on the standard particles with particle sizes of 0.2 μm, 0.3 μm, and 0.5 μm to obtain 1000 pieces of data respectively, and the peak value of these 1000 pieces of data was counted according to formula (3) using Python language. The data results are shown in Figure 7, where the x-axis represents voltage and the y-axis represents probability. It can be seen from the figure that the distribution of the pulse signal does not satisfy the center symmetry, and the voltage corresponding to the peak of the amplitude distribution increases as the particle size increases.
7. Conclusions
This article introduces a submicron particle test method for detecting backscattered light signals. In the detection device of this method, the incident Gaussian
Figure 7. Experimental (point) and fitting (solid line) results of the amplitude distribution of the laser scattering pulse signal.
beam and the scattered beam pass through the same lens, which makes the measurement area independent of the optical device, thereby enhancing the shockproof performance of the measurement device, which is convenient for practical applications.
Through theoretical analysis and experimental research on the backscattering method, the article obtains the variation law between the particle size and particle concentration and the backscattered light energy. When the number of samples is collected enough, the peak distribution of the pulse waveform is analyzed and it is found that the statistical amplitude distribution of the particles agrees well with the lognormal distribution. The experimental measurement shows that the backscattered light energy distribution of the particles has a linear relationship with the particle size and is consistent with the theoretical analysis results, and the number of particle backscattering response signals in the same time period increases with increasing concentration.