International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2013, 2, 133-138
Published Online November 2013 (http://www.scirp.org/journal/ijmpcero)
http://dx.doi.org/10.4236/ijmpcero.2013.24018
Open Access IJMPCERO
Dosimetric Study of Coplanar and Non-Coplanar
Intensity-Modulated Radiation Therapy Planning for
Esophageal Carcinoma*
Ying Li, Bing Liu, Fushan Zhai#, Yongfeng Yang, Ming Liu, Chaoen Bao, Qingxiang Zhou
Department of Radiation Oncology, Third Hospital of Hebei Medical University, Shijiazhuang, China
Email: #zhaifushan@126.com
Received April 25, 2013; revised May 15, 2013; accepted June 5, 2013
Copyright © 2013 Ying Li et al. This is an open access article distributed under the Creative Commons Attribution License, which
permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
ABSTRACT
Purpose: To compare the dosimetric impact of coplanar intensity modulated radiation therapy (IMRT) and non-copla-
nar IMRT for the esophageal carcinoma. Methods: There are forty-five esophageal carcinoma patients, fifteen of whom
were cervical and upper thoracic (Group 1) and thirty were middle and lower thoracic (Group 2). Gross tumor volume
(GTV), clinical target volume (CTV), and organs at risk (OAR) were contoured by the chief physician in the CMS-XiO
treatment planning system. For each patient, one coplanar plan and two non-coplanar plans have been created using the
same physical objective function. A detailed dose-volume histogram (DVH) comparison among three plans was then
carried out in a tabulated format. Results: 1) In Group 1 patients with PTV volume less than 100cc, the mean dose and
dose gradient of non-coplanar plan were much better than those in coplanar plan. 2) In Group 2 patients, the conformity
index (CI) for coplanar and two non-coplanar plans were 0.69 ± 0.13, 0.41 ± 0.13, and 0.68 ± 0.15, respectively. The
V5, V10, V20, and the mean dose to the lung were lower in the non-coplanar plans compared to ones in coplanar plan.
However, the non-coplanar plans resulted in an increase in a dose to the heart, but the dose was still within heart toxicity
tolerance. Conclusion: For Group 1 patients, the non-coplanar IMRT plan had less dose gradient and better mean dose
than the coplanar IMRT plan. For Group 2 patients, the non-coplanar IMRT could the decrease dose to the lung tissue,
thus lowering the probability of radiation pneumonia to esophageal cancer patients. The drawback of non-coplanar IMRT
is that, even within toxicity tolerance, it could deliver a higher dose to the heart and spinal cord compared to the coplanar
plan. Therefore, for patients with cardiology and neurology concern, non-coplanar IMRT should be used with caution.
Keywords: Esophageal Carcinoma; Coplanar IMRT; Non-Coplanar IMRT
1. Introduction
Radiotherapy is the primary treatment modality for the
inoperable or unresectable esophageal carcinoma. The
goal of radiotherapy for esophageal cancer is to kill the
cancer cell inside the target volume while sparing the
normal tissues. In the recent years, intensity-modulated
radiation therapy (IMRT) has been broadly used in treat-
ing esophageal carcinoma [1]. IMRT has been proven to
be superior to Three Dimensional Conformal Radiother-
apy (3DCRT) with respect to dose conformity in Plan-
ning Treatment Volume (PTV) and the normal tissue
preservation [2,3]. However, depending on the volume
and location of the esophageal cancer tumor, the normal
lung tissue may be exposed to high doses of radiation.
Several studies have shown that the incidence and sever-
ity of radiation pneumonia were related to the irradiation
to normal lung tissue in esophageal carcinoma radio-
therapy. For instance, Wang et al. [4] found that, among
100 patients, 49% of them have developed various sever-
ity pneumonia, in whom 27% have grade 1, 16% have
grade 2, 6% have grade 3 pneumonia. From a clinical
aspect, decreasing the side effect of radiotherapy is an-
other way to improve the survival rate.
Since non-coplanar IMRT plan usually takes longer
treatment planning time and it requires complex patient
setup and treatment, the majority of the esophageal car-
cinoma cases are treated by coplanar IMRT, and clini-
cians have yet to explore the feasibility of non-coplanar
IMRT in treating esophageal cancer. The main purposes
*Thanks to Dr. Junfang (Jeff) Gao who is medical physicist at Procure
p
roton therapy center at Oklahoma city in USA for language assistance!
#Corresponding author.
Y. LI ET AL.
Open Access IJMPCERO
134
of this study were to optimize the treatment approach for
the esophageal cancer and improve the treatment quality
by investigating the dosimetric impact of non-coplanar
IMRT in our hospital. Additionally, this study aims to
establish radiotherapy treatment strategies for future cli-
nical trials, especially for esophageal carcinoma.
2. Methods and Materials
2.1. Patient Selection
Forty-five patients with esophageal carcinoma were en-
rolled in this study. Fifteen of them had cervical and up-
per thoracic tumors (Group 1), and the remaining thirty
patients had middle and lower thoracic section tumors
(Group 2). Median age of patients was 61 years (range,
48 - 72 years), and the median PTV was 114.98cc (range,
30.7cc - 235.12cc). All patients in this study had tumors
located away from the distal esophagus and gastro-eso-
phageal junction.
2.2. Simulation
Both the Groups 1 and 2 patients were simulated in a
supine position under the thermoplastic mask immobili-
zation of the head, neck and shoulders. Patients were
simulated on a computed tomography (CT) scanner (So-
motom-sensation Plus-16) using slice thickness of 5-mm
from mandible to the costophrenic angle. This extended
CT scan volume was used to fully visualize the non-co-
planar beams in treatment planning system (TPS). The
acquired CT images covered the entire thorax and upper
abdomen. The CT images were then transferred to the
TPS via a local area network.
2.3. Contouring
The gross tumor volume (GTV) was contoured by a ra-
diation oncologist. The clinical target volume (CTV) was
expanded with a 0.5 cm in radial direction and a 3 to 5
cm superior-inferior direction, which followed our clini-
cal guidelines. The PTV was defined as an additional 0.5
cm expansion around the CTV. The organ at risk (OAR)
included the spinal cord, lung, and heart.
2.4. Treatment Planning
For each patient, one coplanar IMRT plan (referred as
Plan A) and two non-coplanar IMRT plans (referred as
Plan B and C) were created in the XiO TPS, version 4.40
(ELEKTA, CMS St Louis, USA). Specifically, the co-
planar plan (Plan A) was considered as the reference plan
with three to five beams arrangement. For Group 1 pa-
tients, if the plan A has three beams, we used the beam
setup of one posterior-anterior (PA) and two anterior-
oblique (AO) beams (gantry angle at 180˚ ± 10˚, 50˚ ±
10˚, and 310˚ ± 10˚). If it is a four-beam Plan A, we used
one PA, one anterior-posterior (AP), and two AO (gantry
angle at 0˚, 50˚ ± 10˚, 230˚ ± 10˚, and 310˚ ± 10˚). Simi-
larly, five-beam Plan A consisted of one AP, two AO,
and two PO beams (gantry angle at 0˚, 50˚ ± 10˚, 130˚ ±
10˚, 230˚ ± 10˚, and 310˚ ± 10˚).
For Group 2 patients, if it is a three-beam Plan A, one
AP and two PO (gantry angle at 0˚, 130˚ ± 10˚, and 230˚ ±
10˚) were used; if it is four-beam Plan A, one PA, one AP,
and two AO or one parallel-opposed oblique beams (gan-
try angle at 0˚ and/or 180˚, 50˚ ± 10˚ and/or 310˚ ± 10˚,
130˚ ± 10˚ and/or 230˚ ± 10˚) were used; if it is five-beam
Plan A, one AP, one PA, two AO, and one PO beams (gan-
try angle at 0˚, 50˚ ± 10˚, 130˚ ± 10˚, 180˚ ± 10˚, and 310˚ ±
10˚) were used. The couch angle was always set to 0˚.
For each patient case, the non-coplanar Plan B was
morphed from the reference Plan A by converting one or
two of the coplanar beams in Plan A into the non-copla-
nar beam/s. For Plan B, the total beams of 3 to 5 were
used, and the gantry angle of the non-coplanar beam was
set to 330˚ or 30˚ or 150˚, and couch angle was set to 90˚.
The non-coplanar Plan C was using the identical beam
parameters as in the Plan B, with addition of two more
non-coplanar beams with gantry angles of 330˚ and 30˚,
thus, making the total of 5 to 7 beams in Plan C.
All the IMRT plans were generated using 6 mega-
voltage (MV) X-ray beam. The intermediate dose pre-
scribed to the PTV was 60 Gy in 30 fractions. Dose to
the OARs, such as the lung, spinal cord, and heart, were
minimized to the acceptable tolerances. The treatment
planning was done with an objective of meeting the
planning criteria: Dose to the 95% (D95%) of the PTV
volume receives the prescribed dose (60 Gy); D100% of
the PTV is greater than 57 Gy; D5% of PTV is less than
63 Gy. The spinal cord dose was limited to 45 Gy for
0.1cc. For the total lung, the V5, V20, and mean dose were
expected to be lower than the 50%, 25%, and 13 Gy,
respectively (in absolute percentage of the lung volume
at 5 and 20 Gy). For the heart, the planning goals were to
keep V40 less than 40% and mean dose less than 30 Gy
(in absolute percentage of the heart volume). All the
plans were calculated using superposition algorithm with
a dose calculation grid size of 0.2 cm.
2.5. Plan Evaluation
Each plan was evaluated with respect to the dose distri-
bution, dose-volume histograms (DVHs), and additional
dosimetric parameters described below. Comparisons of
treatment plans were based on doses delivered to the
PTV and OARs. The dose distribution of PTV was as-
sessed by evaluating the maximum dose, mean dose, and
minimum dose.
To evaluate the plan quality with respect to the dose
delivered to the tumor, the conformity index (CI) and
heterogeneity index (HI) were computed.
Y. LI ET AL.
Open Access IJMPCERO
135
ref ref
ref
VT VT
CI= VT V
× (1)
where, VT is the volume of PTV; VTref is the volume of
PTV enclosed by the 95% prescription isodose cloud;
Vref is enclosed by the 95% prescription isodose cloud.
The CI is usually 0 - 1, with a larger value indicating
better conformity.
5%
95%
D
HI= D (2)
where, D5% and D95% correspond to the dose delivered to
5% and 95% volume of the PTV, respectively. Greater
HI values indicate doses exceeding the prescription dose
and, thus, a greater degree of dose heterogeneity in the
PTV.
To evaluate the effect of IMRT on normal lung tissue,
heart, and spinal cord irradiation, we computed several
different dosimetric indices, including V5, V10, V20, and
V30 for the normal lung and mean dose delivered to the
normal lung (MLD). The rationale behind using V5-V30
for the normal lung evaluation in comparing the different
plans was based on observations that lung tissue tends to
have a low dose tolerance. We also calculated V30, V40,
V45, V50, and V55 for the heart, as well as mean dose and
D1cc volume of spinal cord dose.
2.6. Statistical Analysis
The different plans were compared using mean statistics.
Quantile-quantile plots showed the data to be approxi-
mately normally distributed, so the differences between
means were tested for significance using a two-tailed
paired Student’s t-test. The null hypothesis was that there
was no difference between the coplanar IMRT technique
and the non-coplanar IMRT treatment techniques. Statis-
tical significance was set at P < 0.05.
3. Results
3.1. Group 1 Patients
Table 1 shows the doses to the PTV, CI, and HI for three
different IMRT plans (A, B and C). The results showed
that all three different plans are very similar. We further
classify this group patient into two sections based on the
PTV volume: First section includes the patients with
PTV volume less than 100cc, and the second section con-
sisted of patients with PTV volume greater than 100cc.
In the section with PTV volume less than 100cc, the PTV
mean doses (cGy) were 6528.33 ± 286.93, 6354.00 ±
270.87, 6354.00 ± 270.87 for Plans A, B and C, respec-
tively. The PTV mean dose of Plan B was closest to the
prescription dose and showed statistical difference com-
pared with Plan A (P = 0.037). The HI of Plan B was
closest to 1, and showed statistical difference compared
with Plan A (P = 0.020). More detailed comparisons of
the Plan A, B, and C are presented in Table 2.
3.2. Group 2 Patients
Table 3 shows that the CI for Plans A, B, and C were
0.69 ± 0.13, 0.41 ± 0.13 and 0.68 ± 0.15, respectively.
The CI of plan B was the lowest and had statistical dif-
ference compared to the reference Plan A (P = 0.000).
Plan C and Plan A has no statistical difference (P =
0.807). More detailed comparisons of the three plans are
presented in Table 3.
The OAR analysis is presented in Table 4, Figure 1,
and Figure 2. The results showed that the lung V5(%),
V10(%), V20(%), V30(%), and mean lung dose (MLD)
(cGy) of reference Plan A were 42.33 ± 8.24, 29.65 ±
8.57, 15.71 ± 4.42, 7.66 ± 4.74, and 882.70 ± 191.83,
respectively. The results were consistent with the finding
from the isodose distribution as lung V5(%), V10(%),
V20(%), V30(%), and MLD (cGy) were reduced to 16.27,
10.59, 5.81, 2.35 and 248 using the Plan B (P < 0.000).
Table 5 shows the dosimetric results of the heart. The
V30(%),V40(%), V45(%), V50(%), V55(%), and MLD (cGy)
of Plan A were 19.00 ± 10.23, 13.00 ± 11.98, 9.09 ± 9.35,
5.88 ± 6.46, 3.30 ± 5.02, and 1729.71 ± 1025.13, respec-
tively. Compared to these values in Plan A, the Plans B
had an increase of 11.05 for V30(%), 9.79 for V40(%),
7.84 for V45(%), 5.86 for V50(%), 4 for V55(%), and
626.94 for MLD with statistical significance (P < 0.05).
Additionally, when compared to the Plan A, the Plan C
also showed an increment in V30(%), V40(%), V45(%), but
not in V50(%), V55(%), and MLD when a non coplanar
field was used. The spinal cord D1cccGyof reference
Plan A was 3127.36 ± 740.8. Compared to Plan A, both
the Plan B and C had slightly higher dose; however, all
sets of plans had D1cc 4500 cGy for the spinal cord.
4. Discussion
In external beam radiation therapy, the choice of beam
parameters plays an important role when we optimize the
IMRT treatment plan. The selection of beam angle is par-
ticularly important as the beam angle directly affects the
patient setup and treatment plan quality, especially in the
cases where the tumor target is wrapped around multiple
OARs [5].
Beam angle selection in IMRT plan optimization has
been investigated by several institutions [5-8]. Allen [9] and
Tucker [10] reported that non-coplanar IMRT significantly
can improve the dose distribution when tumor is close to
the spinal cord. Bedford et al. [11] reported that the use of
inverse planning algorithm to generate 3 to 6 beam
non-coplanar plans without intensity-modulation could
provide better rectal sparing in conformal prostate plan
compared to a three-field coplanar plan. Olivier et al. [12]
Y. LI ET AL.
Open Access IJMPCERO
136
Table 1. Averaged dosimetric results of target volume in Group 1 patients.
Dmax (cGy D
min (cGy D
mean (cGy CI HI
Plan A 6810.51 ± 777.00 5096.70 ± 930.80 6239.52 ± 623.31 0.76 ± 0.17 1.11 ± 0.08
Plan B 6616.70 ± 470.80 5144.51 ± 1021.33 6191.50 ± 569.60 0.75 ± 0.19 1.09 ± 0.06
Plan C 7012.22 ± 821.00 5222.73 ± 917.21 6339.74 ± 645.13 0.80 ± 0.10 1.14 ± 0.08
T value
P value
1.13* - 0.76#
0.340* 0.498#
0.67* - 4.11#
0.546* 0.054#
1.46* - 0.87#
0.240* 0.448#
0.65* - 0.79#
0.561* 0.484#
1.14* - 0.80#
0.334* 0.479#
Note:* = P l an B compared with Plan A; # = Plan C compared with Plan A.
Table 2. Average dosimetric results of PTV less than 100cc in Group 1 patients.
Dmax (cGy D
min (cGy D
mean (cGy CI HI
Plan A 7119.66 ± 576.36 5391.00 ± 883.36 6528.33 ± 286.93 0.73 ± 0.19 1.11 ± 0.06
Plan B 6797.33 ± 370.04 5424.00 ± 1046.88 6354.00 ± 270.87 0.72 ± 0.21 1.08 ± 0.04
Plan C 7385.66 ± 418.06 5535.00 ± 822.72 6661.66 ± 49.16 0.79 ± 0.12 1.13 ± 0.07
T value
P value
2.01* - 0.73#
0.182* 0.537#
0.33* - 4.11#
0.767* 0.054#
5.05* - 0.85#
0.037* 0.483#
0.38* - 0.98#
0.736* 0.429#
3.34* - 1.44#
0.020* 0.209#
Note: * = Pl an B compared with Plan A; # = P l an C compared with Plan A.
Table 3. Averaged dosimetric results of target volume in Group 2 patients.
Dmax (cGy D
min (cGy D
mean (cGy CI HI
Plan A 6198.00 ± 1155.47 4413.53 ± 965.83 6230.32 ± 1171.13 0.69 ± 0.13 1.08 ± 0.04
Plan B 6156.21 ± 1021.97 4484.68 ± 1020.97 5382.15 ± 980.93 0.41 ± 0.13 1.09 ± 0.03
Plan C 6230.32 ± 1171.13 4482.15 ± 980.93 5828.52 ± 1053.96 0.68 ± 0.15 1.09 ± 0.05
T value
P value
0.80* - 1.40#
0.432* 0.178#
0.79* - 1.50#
0.438* 0.149#
2.00* 1.01#
0.064* 0.326#
7.98* 0.24#
0.000* 0.807#
0.80* - 0.80#
0.434* 0.434#
Note: * = Plan B compared with Plan A; # = Plan C compared with Plan A.
Table 4. Averaged dosimetric results of normal lung tissue in Group 2 patients.
V
5 (%) V10 (%) V20 (%) V30 (%) MLD (cGy)
Plan A 42.33 ± 8.24 29.65 ± 8.57 15.71 ± 4.42 7.66 ± 4.74 882.71 ± 191.83
Plan B 26.06 ± 10.14 19.06 ± 8.50 9.90 ± 5.03 5.31 ± 3.32 634.00 ± 244.85
Plan C 40.04 ± 7.83 25.51 ± 7.85 12.88 ± 5.18 5.05 ± 4.02 775.94 ± 186.12
T value
P value
7.53* 3.66#
0.000* 0.002#
5.82* 5.07#
0.000* 0.000#
6.97* 4.09#
0.000* 0.001#
2.37* 4.39#
0.020* 0.000#
6.86* 6.64#
0.000* 0.000#
Note: * = Pl an B compared with Plan A; # = Pl an C compared with Plan A.
Table 5. Averaged dosimetric results hear t in Group 2 patients.
V
30 (%) V40 (%) V45 (%) V50 (%) V55 (%)
Plan A 19.00 ± 10.23 13.00 ± 11.98 9.09 ± 9.35 5.88 ± 6.46 3.30 ± 5.02
Plan B 30.05 ± 15.35 22.79 ± 17.34 16.93 ± 13.15 11.74 ± 10.12 7.30 ± 8.95
Plan C 21.14 ± 13.83 15.85 ± 14.52 10.80 ± 10.27 6.39 ± 7.14 3.73 ± 5.56
T value
P value
5.32* - 2.25#
0.000* 0.020#
5.36* - 2.23#
0.000* 0.043#
5.99* - 2.87#
0.000* 0.010#
4.88* - 2.05#
0.000* 0.061#
3.43* - 1.25#
0.004* 0.225#
Note: * = Pl an B compared with Plan A; # = Pl an C compared with Plan A.
found that using non-coplanar fields in three-dimensional
conformal radiotherapy (3DCRT) and IMRT can dra-
matically reduces the dose to the heart in irradiation of
middle and lower lung tumors. Liu et al. [13] compared
and analyzed the 3DCRT plans with coplanar beam ante-
rior field and non-coplanar beam anterior field, as well as
Y. LI ET AL.
Open Access IJMPCERO
137
Figure 1. Discrepancy be tween Plan B and Plan A in Group
2 patients (Note: Dashed line = Plan B; Solid line = Plan A.
Brown = PTV, Green = Heart, Red = Spinal cord, Blue =
Lungs).
Figure 2. Discrepancy be tween Plan C and Plan A in Group
2 patients (Note: Dashed line = Plan B. Solid line = Plan A,
Brown = PTV, Green = Heart, Red = Spinal cord, Blue =
Lungs).
assess the dose distribution in the planning target volume
PTV and OARs in treating thoracic esophagectomy un-
der the conditions of the PTV length 19 cm. That study
[13] found that the use of non-coplanar beam can get
better dose distribution on target area and reduce the dose
to spinal cord.
To date, this is the first study, which investigates the
dose distribution around esophageal target and OARs in
the coplanar IMRT and non-coplanar IMRT in detail for
a large group of patients (n = 45). In our study, for Group
1 patients, we found that Plan B produced the mean PTV
dose close to the prescribed dose and HI was also closer
to 1. It showed some potential benefit of using non-co-
planar IMRT plan as non-coplanar beam can reduce the
effect of the dose distribution to the body surface to some
extent, especially when the target volume is relatively
small (<100cc).
Radiation pneumonitis is the main factor to limit cli-
nical therapeutic effect and our goal was to reduce the
irradiation to the lung. Graham [14] found that the inci-
dence and degree of radiation pneumonitis are closely
correlated with radiation volume and dose. There was a
significant association between the degree of the radia-
tion pneumonitis and the V20, V30, and mean dose of lung.
Meanwhile, Allen et al. [15] and Tucker et al. [16] found
that, with the increase of V5, the mortalition of radiation
pneumonitis will increase. In our study, non-coplanar
IMRT plans could be an effective tool to reduce irradia-
tion volume of lung. However, as shown for Plan B, re-
duction in the irradiation of lung may also reduce the
conformality of plan. The reason we found that, compare
to Plan A, the volume of PTV enclosed by the 95% pre-
scription isodose cloud was similar, but the 95% pre-
scription isodose cloud significantly increase. Using Plan
C, it was possible to limit the irradiation of lung without
reducing the conformality of plan.
We also found that non-coplanar IMRT plans in-
creased the dose to the heart while reducing dose to the
lung. In comparing our data with those from other studies
on prediction of radiation injury of the heart, we found
that two types of non-coplanar IMRT plans investigated
in this study can meet the heart clinical constraints,
which traditionally have been considered acceptable [17,
18]. Based on these results, it clearly appears that non-
coplanar IMRT Plan C provided better benefit. It will
decrease the severe acute toxicity of lung without creat-
ing the radioactivity injury of heart.
In our treatment planning study, the delivery technique
was limited to the IMRT. Currently, the volumetric in-
tensity modulated arc therapy (VMAT) technique is
available for the treatment of cancer. The VMAT is gain-
ing more popularity since VMAT requires less number of
monitor units and treatment time when compared to
IMRT. Recently, Rana et al. utilized the VMAT planning
technique on esophageal cancer treatment plans to inves-
tigate the impact of dose calculation algorithms in terms
of dosimetric [19] and radiobiological study [20]. The
findings from those studies [19,20] of Rana et al. showed
that the VMAT has a great potential of reducing dose to
the lung tissue and heart while providing adequate target
coverage for the esophageal cancer patients. Since Rana
et al. [19,20] used the coplanar arcs in their studies, it
would be interesting to further investigate the impact of
non-coplanar arcs on dosimetric results of esophageal
cancer treatment plans generated by the VMAT tech-
nique.
5. Conclusion
For cervical and upper thoracic (Group 1) patients, the
non-coplanar IMRT plan had less dose gradient and bet-
ter mean dose than the coplanar IMRT plan. For mid- dle
and lower thoracic (Group 2) patients, the non-coplanar
IMRT can decrease the irradiation to the lung, thus low-
ering the probability of radiation pneumonia to the eso-
phageal cancer patients. The drawback of non-coplanar
IMRT is that, even within toxicity tolerance, it will de-
liver the higher dose to the heart and spinal cord com-
Y. LI ET AL.
Open Access IJMPCERO
138
pared to the coplanar IMRT plan. Therefore, for patients
with cardiology and neurology concern, non-coplanar
IMRT should be used with caution.
REFERENCES
[1] Q. Wu, M. Manning, R. Schmidt-Ullrich, et al., “The
Potential for Sparing of Parotids and Escalation of Bio-
logically Effective Dose with Intensity-Modulated Radia-
tion Treatments of Head and Neck Cancers: A Treatment
Design Study,” International Journal of Radiation On-
colo gy*Biol ogy*Physi cs, Vol. 46, No. 1, 2000, pp. 195-
205. http://dx.doi.org/10.1016/S0360-3016(99)00304-1
[2] V. W. Wu, D. L. Kwong and J. S. Sham, “Target Dose
Conformity in 3-Dimensional Conformal Radiotherapy
and Intensity Modulated Radiotherapy,” Radiotherapy &
Oncology, Vol. 71, No. 2, 2004, pp. 201-206.
http://dx.doi.org/10.1016/j.radonc.2004.03.004
[3] C. Y. Hsiung, E. D. Yorke, C. S. Chui, et al., “Intensity-
Modulated Radiotherapy versus Conventional Three-Di-
mensional Conformal Radiotherapy for Boost or Salvage
Treatment of Nasopharyngeal Carcinoma,” International
Journal of Radiation Oncology*Biology*Physics, Vol. 53,
No. 3, 2002, pp. 638-647.
http://dx.doi.org/10.1016/S0360-3016(02)02760-8
[4] L. Wang, C. Han, X. Zhang, et al., “Three-Dimensional
Conformal Radiotherapy for Esophageal Cancer,” Chi-
nese Journal of Clinical Oncology, Vol. 35, No. 8, 2008,
pp. 424-427.
[5] A. Pugachev, J. G. Li, A. L. Boyer, S. L. Hancock, et al.,
“Role of Beam Orientation Optimization in Intensity-
Modulated Radiation Therapy,” International Journal of
Radiation Oncology*Biology*Physics, Vol. 50, No. 2,
2001, pp. 551-560.
http://dx.doi.org/10.1016/S0360-3016(01)01502-4
[6] E. Schreibmann, et al., “Feasibility Study of Beam Ori-
entation Class Solutions for Prostate IMRT,” Medical
Physics, Vol. 31, No. 10, 2004, pp. 2863-2870.
http://dx.doi.org/10.1118/1.1797571
[7] Y. Li, J. Yao and D. Yao, “Automatic Beam Angle Selec-
tion in IMRT Planning Using Geneticalgorithm,” Physics
in Medicine and Biology, Vol. 49, No. 10, 2004, pp.
1915-1932.
http://dx.doi.org/10.1088/0031-9155/49/10/007
[8] E. K. Lee, T. Fox, I. Crocker, et al., “Simultaneous Beam
Geometry and Intensity Map Optimization in Intensity-
Modulated Radiation Therapy,” International Journal of
Radiation Oncology*Biology*Physics, Vol. 64, No. 1,
2006, pp. 301-320.
[9] J. Meyer, J. A. Mills and O. C. Haas, “Accommodation of
Couch Constraits for Coplanar Intensity Modulated Ra-
diation Therapy,” Radiotherapy & Oncology, Vol. 61, No.
1, 2001, pp. 23-32.
http://dx.doi.org/10.1016/S0167-8140(01)00393-0
[10] P. Andrei, G. L. Jonathan, L. Arthur, et al., “Role of
Non-Coplanar Beams in IMRT,” Proceedings of the 22nd
Annual EMBS International Conference, 23-28 July 2000,
pp. 456-459.
[11] J. L. Bedford, A. J. Henrys, D. P. Dearnaley, et al.,
“Treatment Planning Evaluation of Non-Coplanar Tech-
niques for Conformal Radiotherapy of the Prodtate,” Ra-
diotherapy and Oncology, Vol. 75, No. 3, 2005, pp. 287-
292. http://dx.doi.org/10.1016/j.radonc.2005.03.023
[12] C. T. Olivier, K. Mustapha, J. Patrice, et al., “Potential
Benefits of Using Non-Coplanar Field and Intensity
Moduhad Radiation Therapy to Preserve the Heart in Ir-
radiation of Lung Tumors in the Middle and Lower
Lobes,” Radiotherapy & Oncology, Vol. 80, No. 3, 2006,
pp. 333-340.
http://dx.doi.org/10.1016/j.radonc.2006.07.009
[13] H. Liu, J.-K. Li, X.-P. Wang, et al., “Dosimetry Study on
Non-Coplanar Beam in 3D-CRT for Thoracic Esophagec-
tomy,” Chinese Journal of Cancer Prevention and Treat-
ment, Vol. 18, No. 13, 2011, pp. 1036-1038, 1053.
[14] M. V. Graham, J. A. Purdy, B. Emami, et al., “Clinical
Dose-Volume Histogram Ananlysis for Pnumonifis after
3D Treatment for Non-Small Cell Lung Cancer (NSCLc),”
International Journal of Radiation Oncology*Biology*
Physics, Vol. 45, No. 2, 1999, pp. 323-329.
http://dx.doi.org/10.1016/S0360-3016(99)00183-2
[15] A. M. Allen, M. Czerminska, P. A. Janne, et al., “Fatal
Pneumonitis Associated with Intensity-Modulated Radia-
tion Therapy for Mesothelioma,” International Journal of
Radiation Oncology*Biology*Physics, Vol. 65, No. 3,
2006, pp. 640-645.
http://dx.doi.org/10.1016/j.ijrobp.2006.03.012
[16] L. Tucker, H. Liu and S. Wang, “Dose-Volume Modeling
of the Risk of Postoperative Pulmonary Complications
among Esophageal Cancer Patients Treated with Concur-
rent Chemoradiotherapy Followed by Surgery,” Interna-
tional Journal of Radiation Oncology*Biology*Physics,
Vol. 66, No. 3, 2006, pp. 754-761.
http://dx.doi.org/10.1016/j.ijrobp.2006.06.002
[17] S. Darbys, P. McGale, R. Peto, et al., “Mortality from
cardiovascular Disease More than 10 Years after Radio-
therapy for Breast Cancer Nation Wide Cohort Study of
90000 Swedish Women,” BMJ, Vol. 326, No. 7383, 2003,
pp. 256-257.
[18] X. Wei, H. H. Liu, S. L. Tucker, et al., “Risk Factors for
Perianrdial Effusion in Inoperable Esophageal Cancer Pa-
tients Treated with Definitive Chemoradiation Therapy,”
International Journal of Radiation Oncology*Biology*
Physics, Vol. 70, No. 3, 2008, pp. 707-714.
http://dx.doi.org/10.1016/j.ijrobp.2007.10.056
[19] S. Rana, K. Rogers, S. Pokharel, et al., “Acuros XB Al-
gorithm vs. Anisotropic Analytical Algorithm: A Do-
simetric Study Using Heterogeneous Phantom and Com-
puted Tomography (CT) Data Sets of esophageal Cancer
Patients,” Journal of Cancer Therapy, Vol. 4, No. 1, 2013,
pp. 138-144. http://dx.doi.org/10.4236/jct.2013.41019
[20] S. Rana and K. Rogers, “Radiobiological Evaluation of
Dose Calculation Algorithms in RapidArc Planning of
Esophageal Cancer Treatment Plans,” Journal of Solid
Tumors, Vol. 3, No. 3, 2013, pp. 44-52.
http://dx.doi.org/10.5430/jst.v3n3p44