Microwave Detection, Disruption, and Inactivation of Microorganisms ()
1. Introduction
The scientific origins of vaccination and thermal inactivation of bacteria, fungi, and viruses go back to the pioneering work of Edward Jenner (1749-1823) and Louis Pasteur (1822-1895). By 1900, Herbert George Wells brought the idea of beneficial and pathogenic microorganisms to the forefront in his novels “The Time Machine” [1] and the “War of the Worlds” [2]. The latter manifested itself in the 1918-1920 influenza pandemic which was caused by the H1N1 virus [3]. Since 2009, the world has undergone two further major influenza pandemics (caused by the subtype of the H1N1 [4] ), and the SARS-CoV-2 virus [5]. In the latter pandemic, wealth and political pressure have generated a global disparity in vaccine uptake between nations, leading to travel restrictions between world regions [6]. In addition, single-use personal protective equipment (such as surgical masks and N95-like respirators) has become an environmental waste problem [7]. Due to the rapid spread of the Omicron variant some two years into the COVID-19 pandemic, some nations (Greece, Christmas—New Year, 2021) have introduced the compulsory wearing of N95-like respirators by the general public. Arguably, the N95 particle filtering standards have brought to the limelight within the public perception of the transmissibility of influenza viruses that have a spheroid diameter of 80 to 300 nm.
This paper reviews a range of potential interactions between microwave energy and microorganisms. For example, there are a number of microwave-based technologies in the development of in-vitro rapid detection of coronavirus (COVID-19) within human blood [8] [9] [10] [11]. These use mono-mode 50-Ohm transmission-line applicators, plus in-vivo inactivation of COVID-19 of an infected individual [12]. For in-vitro, single batch multi-mode microwave oven inactivation of bacteria, fungi, and influenza virus [13] [14] [15] [16] [17], cell membrane disruption [16] [18], sterilization of towels [17], inactivation of Hepatitis C Virus (HCV), Human Immunodeficiency Virus (HIV) inoculated syringe, cigarette filters [19], continuous flow inactivation of airborne and liquid-borne microorganisms within modified microwave ovens have been reported [20] [21] [22]. In the latter [22], the calorimetric calculation method was used to estimate the energy required to inactivate bacteria in both airborne and water-borne phases. Microwave Generation Steam (MGS) decontamination of virus inoculated N95-like respirators within domestic and industrial multi-mode microwave ovens has also been reported [23] - [28], along with thermodynamic analysis and dielectric considerations of MGS have been explored [29] [30]. Mono-mode microwave feed-horn inactivation of influenza virus has been examined [31] [32] [33] [34] [35].
Microwave applicators may be classified into two types: mono-mode (or single-mode) and multi-mode. The microwave oven is a well-known example of a multi-mode applicator in which a complex electromagnetic field pattern is formed by the superimposition of different electromagnetic waves and multiple cavity wall reflections [36] [37] [38]. Consequently, with sufficient energy, dielectric volume heating of organic material (food stuff, or culture fluid containing microorganisms) may undergo non-uniform heating due to size, geometry, and location within the oven’s multi-mode cavity [29] [30]. Mono-mode applicators, such as a 50-Ohm transmission line produce coherent standing waves on the line, where the Standing Wave Ratio (SWR) is determined by the line impedance load termination. Indeed, open-ended transmission lines are used to measure the dielectric properties of polar molecules such as water [39].
This paper is constructed as follows. Section 2 examines in-vitro rapid detection of influenza virus within human blood. Section 3 looks at the inactivation of influenza viruses by mono-mode microwave feed-horn applicators. Section 4 examines multi-mode microwave ovens technology reported for the inactivation and disruption of bacteria, fungi, influenza virus, HCV and HIV. In this work, the term gram-negative and gram-positive (originally used by Hans Christian Gram (1853 to 1938) who developed specific dyes to differentiate microorganisms when viewed under a microscope) is used to differentiate microorganisms that have different membranes or wall structures. Gram-negative microorganisms have an outer membrane (Escherichia coli (E. coli) Pseudomonas aeruginosa (P. aeruginosa) and Pseudomonas fluorescens (P. fluorescens)), while gram-positive microorganisms (Bacillus subtilis var niger (B. subtilis var niger), Candida albicans (C. albicans) and Staphylococcus aureus (S. aureus)) have no outer membrane. Section 5 looks at MGS decontamination of virus inoculated N95-like respirators. Section 6 provides a discussion on the reviewed papers, Section 6.1: virus rapid detection methods, Section 6.2: mono-mode feed-horn inactivation of microorganism and mechanisms of destruction, Section 6.3: microorganism sample size, geometry, and location within microwave oven, Section 6.4: respirator metal nose clips, Section 6.5: bubble formation and Section 6.6: non-thermal and thermal microwave oven inactivation mechanism. Section 7 provides a summary and outlook of this work. In-Vivo inactivation of COVID-19 of an infected individual [12] is not considered in this work.
2. In-Vitro Rapid Detection of Influenza Virus within Human Blood
It is well known that normal (healthy) human blood is an amalgamation of approximately 55% plasma and 45% blood cells (erythrocytes and leukocytes and thrombocytes). These blood cells are typically 10 to 50 time’s size of influenza viruses. Since the early 1950s [40] [41] microwave (1.7 to 24 GHz) measurements have shown that the dielectric constant (εr¢) of human blood exhibits an inverse frequency dependency. Within this dependency there are at least two flat (relaxation) regions: β-dispersion arising from the polarization of the cell membranes in the 10 kHz to 200 MHz region and the γ-dispersion region (near 18 GHz) due to the reorientation of water molecules [42]. Glucose [43] [44] [45] and anti-coagulant agents [46] are also known to alter the dispersion which has a significant importance in biosensor diagnostics. In addition, the dispersion is known to be gender specific [46]. Unlike most bacterial and virus infections that cause an increase in leukocyte and lymphocyte count, recent studies of influenza virus infections found a increase in leukocyte lymphocytes count [47] [48]. Given this knowledge in vitro studies into human blood (with and without coronavirus) within a 50-Ohm transmission-line that incorporates a blood sample test chamber have been undertaken using a Vector Network Analyzer (VNA). In this configuration, the test chamber plus blood acts as a variable non 50-Ohm discontinuity [49] [50]. Examples of the use of both contact and non contact antenna measurements [8] [9] are summarized below.
Figure 1(a) shows a schematic of a bench-top Agilent 2-port PNA-L (VNA) scattering parameter (S-parameter, where S12 is transmission and S11 is reflection) measurement of human blood [8]. Figure 1(b) shows a typical S11 swept frequency response, with no blood added (control; solid line) to a planer circular ring, healthy blood (dotted line) and COVID-19 blood (dashed line). A feature of note is that when healthy blood is in contact, the primary S11 zero frequency registration reduces from 4.7 to 4.38 GHz with respect to the control. When a COVID-19 blood sample is in contact, the primary zero shifts to 5.1 GHz which equals to a positive frequency shift of 720 MHz with respect to healthy. In addition, the apex of the shifted S11 zero falls approximately from −12 to −15 dB, with respect to the healthy blood zero. The secondly S11 zero (not shown here) is in the region of 6.5 GHz forms the beginning of a ripple that is related to the transmission-line length, see Equation (1) [49] [50].
(1)
where c is the velocity of light (2.997 × 108 m/s), Vp is the inverse square root of the transmission-line dielectric constant: for a polyethylene transmission-line Vp = 0.65 to 0.67. The symbol n is the integer number of ripples in a given sweep
Figure 1. (a) VNA 2-port transmission and reflection, (b) S11 swept frequency analysis of circular ring chamber (control; solid line), healthy blood (dotted line), and COVID-19 blood (dashed line). (c) Portable 1-port VNA and non-contact antenna measurement (S11 and S11 phase) of COVID-19.
range and l is the transmission-line length. With c and 2 being constant, the ratio of Vp/l enables the reflection plane to be identified and followed, particularly at the zero that is closest to the blood resonant frequency where the greatest signal response is obtained.
In 2021, Elsheakh et al. [9] reported on a portable/handheld Agilent N9918A 1-port VNA S11 measurement of continuous flowing human blood which is passed over a microwave resonating planar microstrip antenna. In this case the antenna does not come into direct contact with the blood, and the VNA has less dynamic range and available output power compared to the Agilent PNA-L VNA [8]. Figure 1(c) shows a schematic of the microwave set-up. In this arrangement both healthy blood and blood containing COVID-19 are sequentially passed over the antenna (resonating in the 2.45 GHz industrial, scientific and medical (ISM) frequency band), both of which are placed within a chamber. As a control, S-parameter (S11 and S11 phase), measurements are taken without blood flow, then with healthy human blood flow, and COVID-19 contaminated human blood flow. In these three measurements the length of the transmission-line between the VNA and antenna induces a ripple in the frequency response according to the test chamber impedance [49] [50]. At present, logistic regression analysis of theses measurement have yielded a binary outcome (positive or negative COVID-19). Using 66 samples the authors claim a COVID-19 detection accuracy of 63.3% for S11 and 60.6% for S11 phase measurements both of which compare favorably with current molecular polymerase chain reaction tests.
3. Inactivation of Influenza Viruses by Mono-Mode Microwave Feed-Horn Applicator
Inactivation of microorganism using mono-mode microwave feed-horn irradiation has been reported by a number of research groups [31] [32] [33] [34] including a US Patent [35]. Generally the microwave feed-horn applicator comprises a frequency source, amplifier, and feed-horn that illuminate the virus suspension. However there are three means of energy illumination. Figures 2(a)-(c) shows three such means of illumination: direct broadband swept frequency illumination [35] Figures 2(a), direct narrowband frequency illumination Figures 2(b), and parallel reflectarray and focusing reflectarray Figures 2(c). In the case of Figure 2(b) and Figure 2(c) [31] [33] it has been established that the optimum operating frequency for influenza virus inactivation is in the frequency range of f0 = 8.0 to 8.5 GHz. Examples of virus inactivation by microwave feed-horn illumination by Hung et al. and [32] and Yang et al. [33] are given in Section 3.1.
Virus Inactivation by Microwave Feed-Horn Illumination
In 2014, Hung et al. [32] has used a microwave focusing reflect array (Figure 2(c)) operating at a power of 1 W continuous wave (CW) at 8 GHz to inactivate influenza virus subtype H3N2. The virus was immersed in a culture fluid suspension within a 10 cm2 flat Petri dish with a focal point distance of 178 mm from
Figure 2. (a)-(c) Three examples of microwave feed-horn used for illmination of virus. (a) Direct broadband swept frequency illumination [35], (b) direct narrowband frequency illumination [32], and parralel reflect array, and fousing array illumination [33] (c).
the focusing array. Their findings revealed, that for a 15 minute irradiation exposure time with a power density of 27 mW∙cm−2 that an inactivation efficiency of 93 % was achieved within the focal area could be obtained.
Yang et al. 2015 [33] used direct microwave illumination (Figure 2(a)) to inactivate influenza virus subtype H3N2 in the 8.0 to 8.4 GHz range. The input power to the feed-horn was 3.6 W CW at 8.4 GHz. For a similar exposure time of 15 minutes to that of Hung [32], a calculated average illumination power density of 81 mW∙cm−2 was found to produce a subtype H1N1 and H3N2 suspension inactivation efficiency in the region 90% to 100%.
4. Inactivation of Bacteria and Virus within Domestic Multi-Mode Microwave Ovens
This section outlines how multi-mode domestic microwave oven can be used for the thermal and non-thermal inactivation studies of homogeneous microorganism populations. In each case the microwave power source employs a free-running cavity-magnetron frequency is f0 = 2.45 ± 0.1 GHz. Normally the magnetron manufactures stated power, is calibrated using the open water bath method, where the temperature rise of water (1000 ml) is measured over time for a given power level setting [29] [30] [51] [52] [53], based on Equation (2).
(2)
where P is power (W, or J∙s−1), m is the mass (g) of water, C is heat capacity (4.184 J∙g−1∙K−1) of water, ΔT is the change in water temperature (final temperature minus the initial temperature), and t is the heating time measured in seconds. Given this method of power calibration for a domestic microwave oven, at full power the cavity-magnetron operates in the CW mode, or duty-cycle (D) = 100%. For power settings below full power, the cavity-magnetron is sequentially pulsed on and off. For example, a power setting that employs D = 50%, the on-period (ton) is equal to the off-period (toff). Thus for the ton period, the microorganism suspension undergoes dielectric volume heating, and in toff the microorganism suspension cools under conduction processes throughout the suspension volume. For the purpose of this work, the terms: non-thermal microwave and thermal microwave inactivation are used. Figure 3 shows the first series of ton and toff pulses, with the suspension not cooling to its initial temperature. Under these conditions the suspension experiences a step-like incremental increase in temperature over time [53] [54].
The study of microorganism membrane disruption normally requires sublethal conditions, the technique of non-thermal [16] (or, athermal [18] ) microwave irradiation has between developed for this purpose, where the aim is to remove, or lower, the induced thermal heat thereby attempting to isolate the microwave irradiation effect (Figure 3). This is achieved by cooling (26˚C - 29˚C [16] [18] ) the vessel containing the microorganism suspension in the microwave oven and using a D < 100%. For air forced cooling of organic compounds see [55] [56] [57]. The thermal heat is thus removed at the containing vessel wall and by conduction thought out the suspension in the toff period. In practice, most domestic microwave oven control circuits have a toff period limited to 15 to 30 seconds to protect the cavity-magnetron’s cathode [52]. This may lead to a situation where a selected minute process time at medium to low power results in Ton = Toff which equates to 30 seconds of microwave irradiation and 30 seconds no irradiation.
Figure 3. Schematic of thermal and non-thermal microwave heating within a domestic microwave oven. Thermal microwave dielectric volume heating (solid line) and non-thermal dashed line.
4.1. Multi-Mode Microwave Oven Batch Processing of Microorganisms
This section describes single batch in-vitro processing of microorganisms. The vessels considered here are: beaker, flat Petri dish, kitchen pads, syringe towels, and N95-like respirators. Figure 4(a) and Figure 4(b) shows the location of beaker, flat Petre dish, and syringe within a microwave oven. Depending upon the reference paper (13 - 19, 23 - 28) the inactivation is given either as a Log10 reduction (Log10(A/B), or percentage reduction (A − B * 100/A) as a function of experiment conditions. Where the inactivation is reported as total, an equivalent log reduction of 7 is given [58].
4.2. Multi-Mode Microwave Oven Batch Treatment of E. Coli
In 1992, Fujikawa et al. [13] studied the inactivation kinetics of a gram-negative E. coli suspension within a RE-S650 - Sharp Corporation microwave oven. The study was performed as function suspension geometry (cylindrical beaker (200 ml) or flat Petri dish (109 mm diameter × 17 mm high), phosphate buffer and pH suspension, exposure time (0 to 4 minutes), a fixed rated power level of 100, 200, 300 W with sample placed on a 2.46 rpm glass rotating carousel (Figure 3(a) and Figure 3(b)). They found for each unit volume of bacterial suspension, the temperature (T) and bacterial survival rate after a given microwave exposure are a functions of suspension pH, suspension, location, geometry (beaker of flat Petri dish), and exposure time (see Table 1).
Figure 4. (a)-(c) Cartoon of multi-mode microwave ovens for inactivation of microorganism. (a) Batch location of microorganism: beaker [13] [14] [18], (b) flat Petri dish [16], syringe [15] [19] towels [17], and (c) continuous flow tube [20] [21] [22].
Table 1. Multi-mode microwave thermal inactivation of gram-negative E. coli [13] [14] [15] [16].
*Leakage of nucleic acid in suppression, cell membrane disruption via puncturing of the cell membrane.
Woo, Rhee and Park 2000 [14] used a MR301M-LG Electronics, Inc., microwave oven for inactivation of E. coli and B. subtilis suspensions. After each microwave thermal inactivation experiment they measured the nuclei acid within the suspension and imaged the bacteria outer surface using a Scanning Electron Microscope (SEM) and Transmission Election Microscope (TEM). The E. coli suspension temperature and inactivation data is given in Table 1.
Later in 2006, Park et al. [15] studied inactivation of E. coli inoculated kitchen sponge (3.17 × 3.37 × 0.69 cm), kitchen non-metallic scrubbing pad (9.06 × 7.07 × 0.65 cm), and 10 ml-plastic syringe within a RE-S630D0 - Sharp Corporation microwave oven (Figure 3(a) and Figure 3(b)). The material of the sponge foam and scouring pads are not given, but given related patent information at the time the martial used is most likely to be open-cell polyurethane foam and non-woven nylon, respectively [59]. The inactivation was performed as a function time at a fixed cavity-magnetron rated power of 1 or 1.1 kW (reported here as 1050 ± 0.050 W). In addition, thermal temperature measurements were performed to estimate the E. coli suspension temperature throughout each inactivation experiment. Their results are given in Table 1.
Asay et al. 2008 [16] studied thermal and non-thermal (26˚C ± 1˚C) microwave inactivation of 10 ml E. coli suspension within a flat Petri dish, placed on a rotating glass carousel within a ER-754BTC Toshiba microwave oven (Figure 3(b)). The power was set to # 5 out of 9 levels, where 9 is maximum cavity-magnetron rated power of 650 W). The temperature and inactivation as a function of both power and time are shown in Table 1. In addition, propidium iodide staining (the uptake is inversely correlated with cell viability due to loss of membrane integrity) and β-galactosidase assay (an indicator of thermal protein denaturation) were performed. The thermal microwave inactivation results are shown in Table 1, while the non-thermal results are given in Table 2.
In 1998, Tanaka et al. reported on microwave oven sterilization of moist towels that where inoculated using (1 ml aliquots on 7 × 7 cm gauze) of S. aureus, P. aeruginosa and C. albicans. The microwave oven (Corona Co., Ltd., Tokyo) [17] had a rated microwave power of 500 W and irradiation time of 1 minute for a single moist towel and 2 minutes for three moist towels. Section 6 gives further information on this study.
In 2006, Benett et al. [18] reported on microwave non-thermal disruption of E. coli suspension within a 50 ml beaker that was cooled by an ice bath (20˚C to 29˚C) when microwave irradiated within a Panasonic microwave oven (Mn-5853C; rated cavity-magnetron power = 800 W). Microwave irradiation exposure was performed at a medium-low power setting for: 15, 20, 25, 30, 35, and 40 s. After exposure, propidium iodide staining was performed to investigate membrane integrity. Colonies where also counted for cell viability. The sample temperature and disruption data is given in Table 2. Their viability experiment revealed a decrease in viable cell count from ~9.8 × 108 at 15 s; ~9 × 108 at 20 and; 6.4 × 108 cell ml−1 at 25 s. This was followed by a leveling-off cell count in the 30, 35, and 40 s exposure times with a mean value of ~5.3 ± 0.5 × 108 cell ml−1.
In 2016, Siddharta et al. [19] reported upon the microwave oven inactivation of HCV and HIV contaminated syringes and cigarette filter papers with the aim
Table 2. Multi-mode microwave non-thermal membrane disruption of gram-negative E. coli [16] [18].
* Exposure time where cell viability count leveled-off.
to mimic re-use within a population of People Who Inject Drugs (PWID). A Bosch microwave oven (type: MM817ASM, cavity-magnetron rated power of 800 W) was used for this purpose, along with insulin syringes (40 ml, with needle removed), and slim cigarette filters (6 × 15 mm). For the syringe, a HCV and/or HIV suspension is used to spike the syringe. While to mimic storage of used cigarette filters, the virus suspension was allowed to be soaked-up (approximately 0.5 ml [60] ) by the filter and then dried in household foil. Table 3 provides the temperature and inactivation data for an irradiation time of 3 minutes at 90, 180 and 360 W for contaminated syringe and cigarette filter papers.
4.3. Microwave Oven Continues Flow Gram-Positive and Gram-Negative Microorganism Inactivation
In 2010, Wu and Yao [20] reported on a modified microwave oven continuous flow inactivation process for bacteria (gram-positive B. subtilis var niger and gram-negative P. fluorescens) and mold fungi (gram-positive A. versicolor). A commercial microwave oven (Midea Inc, Foshan, Guangdong Province, China) was used for this purpose (Figure 4(c)). This works reports the magnetron operating power conditions: full CW power = 700 W (100% duty cycle (30 s in 30 s), duty cycle 12 s in 30 s for medium power (384 W), and duty cycle 5.1 s in 30 s at the low power setting (119 W). Both airborne and liquid-borne microorganism suspensions were microwave irradiated at pre-set microwave power levels of 119, 384 and 700 W for exposure times between 1 and 3 minutes. Table 4, gives the microorganism inactivation (survival rate) as a function experiment conditions. Both SEM and TEM imaging were performed on the microorganisms after microwave irradiation.
Further in 2014, Wu and Yao [21], reported on airborne and liquid-borne survival rate/inactivation of MS2 bacteriophage (a surrogate for the 2009 pandemic H1N1 influenza virus) using the same modified microwave oven (Figure 2(c)) and power level setting as reported in [20]. The aerosol oven exposure times used were: 1.7, 2, 3, and 5 minutes. Table 5 gives the microorganism inactivation
Table 3. Multi-mode microwave thermal decontamination of HCV and HIV spiked insulin syringe and cigarette filter [19].
Table 4. Multi-mode microwave thermal inactivation of gram-positive and negative microorganisms under continuous flow conditions [20].
The gray text denotes that P. fluorescens is classified as gram-negative bacteria. *SEM and TEM images showed visible damages to the microwave-irradiated liquid-borne.
Table 5. Multi-mode microwave thermal inactivation of MS2 bacteriophage under continuous flow conditions [21].
*SEM and TEM images showed visible damage to MS2 bacteriophage.
(survival rate) as a function experiment conditions. Scanning electron microscope and TEM imaging of 700 W irradiated microorganisms highlighted the presence of visible damage to both cell membrane and elements within the cytoplasm.
In 2019, Wang et al. [22], reported on continuous flow airborne and liquid-borne inactivation of gram-negative E. coli. The continuous flow tube having a 100 mm diameter, 1 mm wall thickness positioned within a microwave oven (M1-L213B, China), Figure 2(c). The microorganisms were irradiated at a microwave power level of 700 W between with flow adjusted to produces exposure times between 0 and 20 minutes. Table 6 gives the E. coli inactivation as a function
Table 6. Multi-mode microwave thermal inactivation of gram-negative E coli continuous flow (airborne and liquid-borne [22].
of airborne and liquid-borne microwave power and exposure time. In their analysis they used the calorimetric calculation (Equation (3)) to measure the heat involved in raising the temperature of air and water.
(3)
where Q is the specific heat (J), m is the mass (g) of the substance, C is heat capacity (J∙g−1∙K−1) of the substance, and ΔT is the temperature change of the substance.
5. MGS Decontamination of Virus Inoculated N95-Like Respirators
In a response to the 2009 and 2019 influenza virus pandemics, experimental studies in to MGS decontamination of bacteria and influenza virus inoculated N95-like respirators have been reported [23] - [28]. The aim of these studies was to kill the microorganisms whilst keeping the physical and visual properties of the respirators intact with a view to re-use the respirators. Like the decontamination of HCV and HIV inoculated syringes and cigarette filters [19] there is a temperature limited processing widow for N95-like respirators. Below 70˚C it is found that there is insufficient: virus inactivation, whereas at the boiling point of water (100˚C) and prolonged irradiation times there is a loss of respirator filter void, loft, structural failure at respirator component interface, plus metal noise clip burning. To improve the understanding of the MGS decontamination process window, a thermodynamic analysis of the process along with dielectric considerations have been reported [29] [30].
As detailed in Table 7, the thermodynamic analysis can be carried out using Equations (2) and (3) to estimate the energy input and energy output (Q), and liquid-phase water conversion into steam for six bacteria and virus inoculated N95-like respirators studies. It was found that many of the experimental variables (rated and applied power) were not reported sufficiently; data mining of supporting research papers and commercial information was undertaken. With this approach, an estimation of the process energy budget and its pathway between raising temperature of the liquid-phase water and the production of steam was made for each microwave oven. The analysis of N95-like respirators outcome with the associated process energy input also allowed meaningful
Table 7. Multi-mode thermal MGS decontamination of inoculated N95-like respirators.
The grey shading average measured rather than a rated cavity-magnetron power.
recommendations to made in choice of a suitable microwave oven for MGS decontamination purposes. This approach can be applied in the assessment of future microwave oven designs for MGS decontamination processes, along with redesign rules of N95-like respirators that enables them to survive the MGS decontamination process.
6. Discussion
The paper has reviewed microwave in-vitro contact and non-contact detection of COVID-19, microwave disruption and inactivation of microorganisms (bacteria, fungi, and virus), and MGS decontamination of N95-like respirators. A discussion on these microwave processes now follows.
6.1. In-Vitro Microwave Rapid Detection of COVID-19
Proof of principle of in-vitro microwave rapid test measurements have been established, further work is required to define if the tests may be implemented in a blood bank within hospitals and health centers. As these tests require blood to be taken from the consenting patient, injection hesitancy [61] may prove to a prohibitive barrier if the tests are used to control travel at international airport, seaports, and land-boarders.
6.1.1. VNA Choice
Notwithstanding injection hesitancy, to enhance the probability of the uptake of these invasive tests, there is a need to demonstrate a predictive accuracy comparable to current molecular polymerase chain reaction and lateral flow tests, and at completive cost. With regard to cost, the choice of whether to use a portable/handheld or bench top VNA is a significant factor, as many portable/handheld VNA have low dynamic range (typically 95 dB) and output power (typically 0 dBm) as compared to a bench top VNA that have a typical dynamic range of 125 dB and output power of 0 to 10 dBm. It is, therefore, important to compare directly the measurement resolution of different VNA’s within future in-vitro rapid detection experiments.
6.1.2. Microwave Frequency Band Selection
There is now considerably evidence that healthily human blood and COVID-19 infected blood has microwave frequency dependency response away from the 2.45 GHz. Human blood relaxation measurements [39] - [46], swept frequency rapid detection experiments [8] [9], microwave feed-horn illumination experiments along with microorganism confined acoustic vibrations theory [31] [32] [33] [34] [35] indicate the frequency band of interest is between 8.0 to 10 s of GHz.
6.1.3. Contact or Non-Contact Antenna Design
The research into biosensors is a rapidly expanding field of interest. For microwave detection of COVID-19, the question whether contact or non-contact planar microstrip antenna provides the most detection sensitivity and selectivity is an important area of future research.
6.2. Mono-Mode Feed-Horn Inactivation of Microorganism
It is generally accepted that the microwave electric field energy couples to the dipole of polar molecules. Recently, it has been proposed that cytoplasm RNA (ribonucleic acid) within the virus spheroid volume collectively act as single dipole in the 8 GHz frequency range. Thereby allowing microwave energy to couple into the mechanical (acoustic) vibrational lowest mode (l = 1) [31] [32] [33] [34] [35]. This energy transfer phenomenon is termed Microwave Resonant Absorption (MRA) [30] [31], or Structure Resonance Energy Transfer (SRET) [33] [34] [35]. As this method of cell destruction becomes more widely accepted, longitudinal modes within rod-like shape viruses have also been examined in the 6 to 40 GHz range [34] [35]. Inherently, the mechanism assumes that an internal force displacement must overcome the rupture pressure of a solid boundary membrane, however further experimental and theoretical work is required to account for cell lysis [14] [16] [18]. A further feature of note is that MRA (SRET) closely resembles the simple theory of diocotron mode (l = 1) [62]: is association thereby provides additional investigation.
6.3. Microorganism Suspension Size, Geometry, and Location within Microwave Oven
Since the multi-mode microwave oven was first patented for treating food-stuff [63], it has been known that dielectric material (polar molecule, such as water and food-stuff), size, geometry, and location within the ovens cavity presents a number of non-even heating challenges [64] [65]. It is standard practice that once heated; the food to be eaten should stand for a few minutes for conduction and convection processes to transmit heat throughout the food [66]. However, Geedipalli et al. [54] showed that food placed upon glass rotating carousel upon improved food temperature uniformity by 40%.
In 2002, Houšová and Hoke [67] studied the dielectric volume heating of small amounts of water (<250 ml) within four different domestic microwave ovens and found it to be oven-specific due to the oven’s cavity size and position of the water load within the oven cavity. Above 250 ml of water, no dependence upon geometry of glass beaker and flat Petri dish could be derived for these domestic microwave ovens. An example of which is the Moulinex microwave oven; type FM 2915Q that has a rated power output 850 W, with a non-rotating glass shelf. They based their findings on Equation (4) [22] [37] [38] [53] [68] [69].
(4)
where Pv is power density in the load (W∙m−3), the 2πfε0ε'' term represents the dielectric conductivity (σ) of the load, where f is the microwave frequency (2.45 GHz), ε0 is the permittivity of free space (8.854 × 10−12 F∙m−1), ε'' is the dielectric loss of the load, and E is the electric field strength inside the load (V∙m−1) at the local position (x, y, z) within the multimode cavity. From this they calculated the relative absorbed power (Pa/Pmax), where Pa is the amount of heated water, and Pmax is the power absorbed by 1000 ml of water.
In this context the aliquots of microorganisms suspension reported in Table 1 and Table 2; i.e., 100 ml [13], 100 ml [14], 10 to 100 ml [15], 10 ml [16], and 50 ml [18] inactivation outcome is most likely to be oven-specific.
Given the theoretical maximum cavity unloaded Q-factor (Qu) may be defined as in Equation (5) [52] [68].
(5)
where Vc is the cavity volume, δ is the electrical skin depth of the cavity wall conductivity (typically 5 microns) and Ac is the cavity wall area. Thus, Qu is defined as the ratio of stored energy in cavity to the energy loss to the cavity walls. Under load conditions the cavity Q-factor falls and is redefined as Ql (Equation (6)) where Vl is the load volume.
(6)
A worked example is given here. Introducing a small load volume (Vl ~ 100 ml (1 × 10−4 m3) into the Moulinex oven (cavity volume = 0.0238 m3, surface area = 0.5 m2) reveals a Q-factor reduction from 19,040 to 18,960, or a 0.42% reduction. Under these load conditions, it is reasonable to assume that the ovens electric field pattern is not significantly altered and remains in a high Q mode where the electric field intensity alternates between high (hot-spots) and low (cool-spots) every ½ wavelength (~6.1 cm). This would indicate that the central vertical axis of glass shelf were the load is placed experiences a hot-spot. Indeed multi-mode cavity, 3-dimensional (x, y, z) multiphysics models provide evidence for this assumption [53] [54] [69]. By extension, it is assumed that the medical syringe inactivation data reported in [15] [19] and towels [17] are comparable to each other and the flat Petre dish inactivation date.
6.4. Respirators Metal Noise Clips and Bubbles within a Microwave Oven
In some MGS decontamination studies it has been recorded that a metal nose-clip sometimes causes local burning damage in certain respirators models. It has been proposed that the mechanism for the burning damage is due to the metal noise clip acting like a microstrip operating in a quasi-transverse electromagnetic (TEM) mode [30]. Under these electromagnetic conditions and as the MGs decontamination proceeds with increasing temperature, the microstrip surrounding air is replaced by water vapor and ultimately steam causing the microstrip phase velocity to decreases with time. Thus the microstrip standing wave reduces in physical length resulting in short- and open-circuit nodes to momentarily align with protrusion and sharp edges of the metal nose clip causing high voltage stress points that lead to local heating and burning of the respirator polymer material.
6.5. Bubble Formation within a Microwave Oven
In this work the dielectric properties of the microorganism and suspension fluid are considered to have similar values at room temperature. However after a period of dielectric volume heating of the virus suspension the liquid-phase approaches saturation temperature, where bubble nucleation is promoted, and which grow in size, and then float to the free-surface where heat is lost from the bulk of the liquid-phase. As the bubbles grow in number a heterogeneous two-phase mixture develops, were the dissolved nature of the air bubbles alters the balk dielectric constant and is defined as (εm) where the subscript m is the liquid-bubble mixture. One way to calculate the binary mixture effective dielectric constant is to use a modified Looyenga cubic mixing formula [30] [53] [70] [71]. Equation (7) gives Loo-yenga’s binary formula where the subscripts denote the liquid and gas vapor phase and ag is the vapor fraction.
(7)
For MGS decontamination of thermally sensitive N95-like respirators that begin to deform at temperatures close to 100˚C, it is worth noting that knowledge of bubble formation is beneficial as these bubbles should help to prevent superheating.
6.6. Non-Thermal and Thermal Microwave Oven Inactivation Mechanism
Microwave dielectric volume heating of polar molecules undergoes polarization and relaxations in their dipoles [36] [37] [64] [65] generating a faster mechanism than thermal heat transfer due to conduction of heating from the outside of the vessel that hold the water. However, within the physical and engineering sciences is has been uncertainty if microwave irradiation alone can cause death in biological systems [72], whereas with more certainty thermal heating is known to cause microorganism inactivation > 50˚C [13] - [18] [53] [54] [55] [56] [58] [73] [74]. It is also known that inactivation of microorganism heterogeneous populations tend to follow a non-first-rate law [73] and microorganism homogeneous populations follow a first-order-rate law. In this work the inactivation of microorganism homogeneous populations has been reviewed and assumed to have a first-order-rate of inactivation, [C] = [Co]e−kt. Where [C] is the reduction microorganism concentration, [Co] is the initial microorganism concentration, k is the rate constant, and t is time.
A comparison of thermal and non-thermal microwave disruption and inactivation (≥4 log10) of microorganisms is shown in Figure 5, where temperature is plotted on the horizontal axis and exposure time on the vertical axis. The microwave data is taken from Tables 1-3. In addition, thermal heat inactivation data is taken from [58] is also plotted. For the non-thermal microwave disruption studies, even though applied microwave power levels of 200 to 650 W are used, the dataset is in the low temperature range (20˚C to 29˚C). Whereas, the thermal microwave inactivation data is in the higher temperature range (50˚C to 86˚C) and is separated into two groups that are characterized by their applied
Figure 5. Thermal microwave, non-thermal microwave, and thermal treatment of microorganisms plotted as a function of time and temperature. The inactivation data point represents a ≥4 log10 reduction. E coli suspension; non-thermal microwave blue circles. E. coli thermal microwave (open squares) plus HCV and HIV (open triangle). E. coli soaked kitchen sponge and pad, and used syringe (red squares; 1.05 ± 0.050 kW). Thermal inactivation of SARS-CoV-2 suspension data (open stars with trend-line) is taken from [58].
microwave power (suspension fluid, syringe, and cigarette filters (100 to 650 W), and inoculated sponge, kitchen pad and syringe (1 to 1.1 kW)). The culture fluid suspension dataset, that includes HCV and HIV, has a 7-point trend-line of −1.1113 and temperature intercept of 74˚C. In comparison the soaked sponge, kitchen scouring pad, and used syringe dataset has a higher temperature intercept of 88˚C indicating a greater microorganism resilience on/in these fomites with respect to the liquid suspension. Note, the syringe requires 4 minutes for E. coli inactivation compared with the 1 and 2 minutes for kitchen sponge and kitchen pad. The thermal inactivation 8-point dataset trend-line is −1.3889, or approximately 12.5 times slower rate than the thermal microwave data, indicating a difference in inactivation mechanisms.
Figure 6 uses the same temperature x time domain as in Figure 5 to provide a comparison of thermal microwave inactivation (≥4 log10) of microorganism suspension, syringe, cigarette filter, plus inoculated N95-like respirators. The microwave data is taken from Table 1, Table 3 and Table 7. Here it is seen that the data point spatial spread is separated into three groups. That is, the E. coli suspension and HCV and HIV inoculated syringe, and cigarette filter at the lower 60˚C to 70˚C range with an associated power of 1 to 1.1 kW, Inoculated kitchen sponge, kitchen scouring pad and plastic syringe in the 80˚C to 90˚C and associated 1 to 1.1 kW, and the N95-like respirators temperature being defined by the boiling point of water (100˚C) at one atmospheric pressure with an associated power of 0.75 to 1.25 kW. It is concluded from this study that three temperature groups
Figure 6. Thermal microwave inactivation (≥4 log10 reduction) of microorganisms plotted as a function of time and temperature. E. coli suspension; thermal microwave open squares circles. HCV and HIV suspension open square with triangle (360 W). E. coli inoculated sponge, kitchen pad, and syringe: red squares (1.05 ± 0.050 kW). MGS decontamination of N95-like respirators: open circle (0.75 and 1.8 kW), close blue circle (1.1 and 1.25 kW), and closed red circle (1.1 to 1.15 kW).
do not relate to the reported cavity-magnetron rate power and applied power. Indeed, the variance in the power groupings appears to be both oven-specific and fomite-specific.
Figure 7 provides an energy phase-space projection of process energy budget (kJ) on the horizontal-axis against the specific energy density of 1 ml of water liquid-phase (kJ∙ml−1) on the vertical-axis. Within the microorganism suspension dataset (beaker, sponge, scrubbing pads and MGS), thermodynamic analysis is used to provide further process clearly. In this projection least square trend-line can be fitted to the beaker, sponge, and MGS data points: starting from the vegetative bacteria and finishing with bacteria spores, with virus and their surrogates in the middle. As expected the bacteria spores are the most resilient to the thermal microwave irradiation stress. Using this knowledge, and as the trend-line progresses, the variance only increases around MGS decontamination points, indicating the possible presence of independent variables, such as respirator-type; water reservoir type (open water bath, bag or sterilizer), oven type (greatest variance is where 2 cavity-magnetrons are employed). Note that the beaker and sponge data points fit on a trend-line with an R2 value of 0.896 indicting a strong association. With this analysis, it is reasonable to propose that the high degree of water absorption and retention (approximately 90% by volume) within the soaked sponge volume behaves like the beaker microorganism suspensions where the suspension
Figure 7. The microwave inactivation (≥4 log10) data presented in Figure 5 and Figure 6, is now presented using a thermodynamic analyses approach, were the process energy budget is plotted against specific energy density of liquid suspension (kJ∙ml−1). E. coli beaker; open circles (100 to 300 W [13] ). E. coli and B. subtitles beaker; open diamond (600 W [14] ). E. coli, MS2, B. cereus and B. cereus spores inoculated sponge; open squares (1.05 ± 0.050 kW [15] ). E. coli inoculated kitchen scrubbing pad: open squares (1.05 ± 0.050 kW CW [15] ). MGS decontamination of N95-like respirators using: open triangles (0.75 to 1.8 kW [23] - [28] ). Least square tend-line excludes the two outliers (red filled square and triangle).
is retained by means of beaker containing wall. Finally there does not appear to be a duty cycle dependency which would suggest that cell wall disruption is masked by the thermal microwave irradiation.
As regards to the outliers: the scrubbing pad (red square) is excluded from the trend-line due to its poor water and absorption and retention properties (approximately 26% by volume), in addition the 2 cavity-magnetron data point (red triangle) is excluded due to the high rated power used (1800 W) and the bi-directional applied power (top and bottom) [30]. Not shown in Figure 7 is the 1 to 10 ml of microorganism suspension used to inoculate towels [17], syringes and cigarette filters [15] [19], due to the inevitable outcome being a relative and disproportionate high energy density value, compared to the beaker, sponge and MGS outcomes. This is because the low volumes used were intended to simulate real-life health care environments and within populations of PWID: where a relative high fraction of the fomite (glass, plastic, or gauze) undergoing dielectric volume heating.
The use of Looyenga’s heterogeneous binary formula may assist in the understanding of microwave sterilization of used syringes [15] [19] and used towels [17], where the volume (mass) fraction of suspension liquid (al) is smaller than both the fomite volume and mass. This aspect of Looyenya’s formula may be used as an exclusion criterion, where the fomite fraction is likely to be 95% to 99%. For this reason, the inactivation (≥4 log10) data obtained for: used towels, scrubbing pads, syringes, and cigarette filters are collected separately and presented in Table 8 that includes an estimation of suspension plus the unknown additional water content for the moist towel. Notwithstanding the lack of moisture content knowledge within the towel, the available data shows that the empty used syringe provides the highest process energy budget. This limited data suggests the near-empty plastic syringe has a low rate of microwave energy absorption compared to the other fomites containing water. From a HCV and HIV infection prevention perspective, it would appear that placing the contaminated near-empty syringes in an open water bath prior to microwave treatment; thereby help standardized sterilization protocols within PWID populations.
Table 8. Inactivation (≥4 log10) data obtained for used: towel, scrubbing pad, syringe, and cigarette filter.
7. Summary and Outlook
This paper has reviewed three interactions between microwave energy and microorganisms (bacteria, fungi, and viruses). The interactions are characterized by their electrical power, mode of power delivery (mono-mode 50-Ohm transmission-line, the feed-horn illumination, and multimode microwave oven), and microorganism outcome (in-vitro contaminated human blood detection, cell-wall disruption, and inactivation). In the latter case, the purpose of inactivation was not only to kill the virus, but also to re-use the decontaminated fomite (HCV and HIV contaminated syringe and cigarette filter, and influenza virus contaminated N95-like respirator). For this dual requirement, the applied energy in the form of dielectric volume heating needs be to sufficient to kill the virus but not sufficient to damage the fomite.
Given these requirements, the three microwave energy processes relate to the following. For in-vitro virus-contaminated blood detection: 0 to 10 dBm @ 2 to 8.5 GHz; for feed-horn illumination and inactivation: 1 to 6.5 W @ 8 to 26 GHz at the feed-horn illumination an applied power density of 10 to 100 mW−2 at the virus suffice. The third energy process that is within a multi-mode microwave oven uses 100 s of W @ 300 to 1800 W @ 2.45 GHz for inactivation, with temperature control between 70˚C to 100˚C for reuse purposes.
Mechanisms for the three microwave interactions are also discussed. Swept frequency analysis of transmission-line terminated with a non-50-Ohm virus load, chemical cellular lysis, and microwave resonant absorption causing cell wall rupture. In addition, the examination of the thermodynamics of input and output process energy budget has been performed with the aim of revealing the inactivation mechanisms in different microorganism suspension volumes and the inactivation of microorganisms on and within different fomites (kitchen sponge, scouring pads syringes cigarette filters and N95-like respirators). To this end, the size of microorganism suspension, geometry, and location within microwave oven have been discussed. Looyena’s heterogeneous binary mixture formula may help in the understanding of how fomites influence the microwave inactivation outcome.
The overall conclusion from this study is that the mechanism of microwave interaction with microorganisms is complex and is clearly dependent on the microorganism type, along with the treatment time and the level of energy imparted by the microwave source. As regards the outlook of the work, there are many research questions that remain to be investigated, two of which are listed here.
1) In the case of the microwave swept frequency of healthy and infected blood studies, it would useful to undertake a lumped (inductor, capacitor, and resister) linear circuit analysis [49] [50] of the non-biological components (transmission-line and test chamber) and compare them to the physiochemical components of the healthy and infected blood. Here is hoped that the LCR elements may provide and insight into the shape and depth of the zeros.
2) The reduced information within the energy phase-space projection of the multi-mode microwave oven inactivation data provides a degree of clarity in the microorganism inactivation process. Here it may useful to use multi-variant tools, such as principal component analysis [75], to differentiate further between the independent variables (number of cavity-magnetrons used and the direction of illumination, fomite, and microorganism). The score information could then be used to standardize future multi-mode microwave oven studies.