Dosimetric Characteristics of 6 MV Medical Linac at BAEC

Abstract

Dosimetric characteristic is one of the essential parameters of a medical linear accelerator (LINAC), which must be obtained before clinical use. The dosimetric characteristics for 6 MV photon beam were measured and compared with the corresponding published data. The study was done using a Varian linear accelerator (Model Clinac-iX) at the Institute of Nuclear Medical Physics (INMP), AERE, Savar, Dhaka, under the Bangladesh Atomic Energy Commission (BAEC). The data is taken for 10 field sizes (2 × 2, 3 × 3, 5 × 5, 7 × 7, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm2) at same conditions. The measured Percent Depth Dose (PDD) curves were obtained for 6 MV photon beams with the field as mentioned above and compared with the calculated PDD curves. The measured depth dose (Dmax) for reference field size (FS) 10 × 10 cm2 is 15.99 mm, and the PDD at 10 cm depth (D10) is 66.87% for 6 MV photon energies that are found to be compatible with the published report BJR supplement 25. The measured PDD curves for photon energies show a good agreement with the standard PDD curves. The photon beam dosimetry data found in the current study are compatible and all the tolerances are within the clinically acceptable tolerance limit.

Share and Cite:

Roy, S. , Das, P. , Khatun, R. , Rahman, M. , Akter, S. , Kumar, T. and Ahasan, M. (2021) Dosimetric Characteristics of 6 MV Medical Linac at BAEC. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 10, 38-46. doi: 10.4236/ijmpcero.2021.101004.

1. Introduction

Dosimetric characteristics are the most important parameter of a linear accelerator (LINAC) which must be obtained prior to clinical use. Most of the developed countries are used basic treatment for linear accelerator. They can produce photon beams, electron beam or both depending on the model used. A linear accelerator can produce high energy x-rays to ensure a tumor’s shape and destroy cancer cells while canning surrounding normal tissue [1]. The Photon beams are produced with a wide spectrum of energies ranging up to the maximum energy of the electrons striking the target, with a peak at about half this maximum value [2] [3] [4]. The linear accelerators have the advantage that the photon beam dosimetric is more penetrating within the cancer cell, a diverse choice of beam energy, a smaller penumbra edge to the beam, and they provide a high dose rate and calculate accurate dose measurement [2].

The LINAC can be used for different radio-therapies (treatment) such as surgery, chemotherapy, and target therapy after completion of some satisfactory scientific methods called as pre-commissioning testing [5]. Therefore, it is essential to have a minimum data set which includes percentage depth dose (PDD), Open Beam Profile (inline, cross line & diagonal) and Beam Output factors for a series of Field Sizes (FSs) are the important dosimetric data that require during the commissioning period of a medical linear accelerator (LINAC). In this study, one basic dosimetric parameter like PDD for 6 MV photon beams with different FSs was measured. These measurements have been performed in a three dimensional (3D) computer controlled water phantom using two ionization chambers (S/N: 15050 and 15051) for the Varian Clinac iX linear accelerator machine at the Institute of Nuclear Medical Physics, Savar, Dhaka. The data obtained during the initial commissioning of LINAC can be treated as the standard data for clinical purpose. The scientific methods used for commissioning of modern Linear Accelerator are really a time-consuming procedure and need dedication in work [5]. In this work, the PDDs were compared for measured data. The objective of this work is to investigate the Varian technology evolution via a comparative study of percentage depth dose (PDD) and especially the build-up region for accelerators Varian Clinac iX. The main objective of this research, using present knowledge of relative dosimetry of photon based on different international literature, reports and guidelines and approach for possible suitable discuss their response in different fields’ size photon. In this present research, different sizes of active volume chambers, water phantom (1D, 3D) were used. A significant amount of data is accumulated for this thesis.

This data set gives the complete output factor to be used in configuring the treatment planning system at SSD = 100 cm and Depth = 10 cm. On the other hand, for the edge detector and the other two ionization chambers measurement readings were taken for field sizes of square widths from 10 cm × 10 cm for a total of 10 square field sizes.

For the two types of ionization chamber and two phantoms used and the measurement combination from two possible SSDs and Depths and also with the choice of one types of field collimation, the whole package constitutes a significantly huge data set. These data and their analyses could constitute a valuable addition toward understanding photon beam radiation measurements in medical linacs. The main advantage of removing maximum and minimum flatness and symmetry was on increasing dose and the increasing of dose rate is approximately 80% to 20% for field size of 10 × 10 cm2 and it is approximately 100% for field size of 10 × 10 cm2.

2. Materials and Methods

The measurements were conducted using Varian Clinac iX (Manufacture: Varian Medical System, USA) (Varian Oncology) at the Institute of Nuclear Medical Physics, Bangladesh Atomic Energy Commission. For this study, 1D and 3D water phantom, ionization chambers, and electrometers are used. The ionization chamber is put in a water phantom, and is connected with a myQA accept software on a computer via an optical cable. The photon beam field area is set at 10 × 10 cm2. Later, the SSD was set at a distance of 90 cm and the energy of 6 MV. The ionization chamber’s depth in a water phantom is positioned with a depth of 5cm from the water surface. The beam focus would set on, and PDD and dose profile would be analyzed using the IBA blue phantom-2 software.

The procedure would be repeated by adjusting the SSD, energy, and depth. The variations of SSD are 90, 100, and 110 cm. The depth of the water phantom is 5 cm, 10 cm, 15 cm, 20 cm, and 25 cm. The water phantom’s depth is created for analysis of percentage depth dose (PDD) and dose profile. Radiation dose profile characteristics of the radiation beam and provide essential information for radiotherapy planning. That curve interprets the distribution of radiation dose relatively on a particular radiation field. Percentage depth dose is the absorbed dose is given on the main depth as a percentage of absorbed doses at a depth of the pointer on the central axis of the area. In order to characterize the dose distribution on the central beam axis in a water phantom, the dose should be normalized to maximum dose [6].

The central axis dose distributions inside the patient or phantom are usually normalized to Dmax = 100% at a depth of dose maximum Dmax and then referred to as the PDD distributions [7]. Dose distribution at the point in the central axis of the files inside the phantom usually normalized to Dmax as 100% at a depth of maximum dose, Dmax corresponding to the reference depth. Value percentage depth dose can be defined as a result, in the form of a percentage, the dose absorbed at a certain depth where a depth d is, d0 is the reference dose along the beam axis [8].

3. Data Collection and Analysis

This data was collected using the clinac iX machine with Eclips 13.7 Version, MyQA Accept software, water phantom, etc. for 6 MV energy, SSD = 100, Dose Rate = 300, Depth = 10 cm.

Ionization chambers CC13 is used to output factor analysis and FC65-P is used to analyze the response of chambers in different field sizes concerning chamber’s active volume. Actually, using Dose-1 electrometer reading in nC, dose for 2 × 2, 3 × 3, 5 × 5, 7 × 7, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm2 field sizes is measured. The output factor shown in Table 1 is measured using different size volume Chambers and is normalized with 10 × 10 cm2 field size dose. This data set gives the complete output factor to be used in configuring the treatment planning system at SSD = 100 cm and Depth = 10 cm. The two ionization chambers measurement readings were taken for field sizes of square widths from 10 cm × 10 cm for a total of 10 square field sizes.

4. Results and Discussion

In the present study, this is a raw data calculation of 10 different field sizes with a 6 MV photons beam that was performed at the Institute of Nuclear Medical Physics (INMP), Bangladesh Atomic Energy Commission. PDDs were measured for various field sizes ranging from 5 × 5 cm2, 10 × 10 cm2, 15 × 15 cm2 and 20 × 20 cm2. Various parameters such as Dmax (1.6 cm), PDD at 5 cm, 10 cm, 15 cm and 20 cm beam were compared for selective field sizes (5 × 5, 10 × 10, 15 × 15, 20 × 20) cm2. Although the dose per pulse is the same for varying dose rates, measurement is performed with a constant dose rate of 300 MU/Min.

The least-square algorithm smoothened all percentage depth doses. After that PDDs were normalizing at a depth of maximum to 100%. Point by point comparisons of the depth dose curve is performed up to a measurement scanning depth of 35 cm. To analyze the beam dose profiles, they were normalized for all radiation field sizes and depths to 100% on the central beam axis. Figure 1 shows 6 MV photon beam’s dose profiles for a field size of 10 × 10 cm2 at different depths on the central beam axis. It can be seen in Figure 1 that the dose profiles were depended on the depth on the central beam axis; the curves of the dose profiles were widened more to more as the depth increased due to an inverse square law (ISL) that it was the relationship between the depth and the beam divergence. Next, the dose profiles were graphed for the photon beam field sizes of 5 × 5 cm2, 10 × 10 cm2, 15 × 15 cm2, 20 × 20 cm2 is shown in Figure 1.

Figure 1. Graphical representation of PDD (5 × 5, 10 × 10, 15 × 15, 20 × 20) raw data.

Table 1. Corresponding field sizes used for Output factors measurement in charge (nC).

The depth dose (Dmax) at 10 cm depth (D10) for a 6 MV photon beam for 5 × 5 cm2 FS was obtained from Figure 1. The PDD at 10cm depth (D10) is 62.8, and Dmax is 100. The maximum PDD (define the field size) 69.7%, and the minimum PDD is 62.8%. For a 6 MV photon beam, the maximum value is 1.5 cm.

The dose profile for 5 × 5 cm2 FS and 10 cm depth SSD = 100 cm are shown in Figure 2. The lateral distance between 80% and 20% of the dose on the central beam axis, called semi-darkness, is for a 6 MV photon beam. Table 2 shows Flatness and Symmetry for 5 × 5 cm2 FS.

The field flatness changes with depth. This is attributed to an increase in scatter to primary dose ratio with increasing depth and decreasing incident photon energy off-axis, as shown in Figure 3. According to this the lateral distance between 80% and 20% of the dose on the beam central beam axis, called semi semi-darkness, is for a 6 MV photon beam. The dose profile for 10 × 10 cm2 FS and 10cm depth, SSD100 cm. Table 3 shows Flatness and Symmetry for 10 × 10 cm2 FS.

It can be seen from Figure 4. that the normalized increase in relative dose curves presented the peaks at a depth of 10 cm for all field size over than field size of 15 × 15 cm2. The peak values increased with field size, the increase in relative dose for the broad field size due to the scattered photon beam of 6 MV Linear accelerator. The flatness of photon beams is extremely sensitive to change in energy of the incident beam. A small change in the photon beam penetrative quality of a photon beam’s penetrative quality results in a considerable change in the beam flatness. The field flatness changes with depth. This is attributed to an increase in scatter to the primary dose ratio with increasing depth and decreasing incident photon energy off-axis, as shown in Figure 4. According to this, the lateral distance between 80% and 20% of the dose on the central beam axes, called semi-darkness, is for a 6 MV photon beam. The dose profile for 15 × 15 cm2 FS and 10 cm depth SSD = 100 cm. Table 4 shows Flatness and Symmetry for 15 × 15 cm2 FS.

The dose profile for 20 × 20 cm2 FS and 10 cm depth SSD100 cm are shown in Figure 5. Flatness and symmetry values for 6 MV photon beam are listed in Table 5.

The dose profile for diagonal Profile(C) 20 × 20 cm2 FS and 10 cm depth SSD = 100 cm are shown in Figure 6. Flatness and symmetry values for 6 MV photon beam are listed in Table 6.

Figure 2. Graphical representation of 6 MVDose Profile(C) for 5 × 5 field size raw data.

Figure 3. Graphical representation of 6 MVDose Profile(C) for 10 × 10 field size raw data.

Figure 4. Graphical representation 6 MVDose Profile(C) for 15 × 15 field size raw data.

Figure 5. Graphical representation 6 MVDose Profile(C) for 20 × 20 field size raw data.

Figure 6. Graphical representation of diagonal Profile(C) for 20 × 20 raw data.

Table 2. Flatness and symmetry for 5 × 5 cm2 FS.

Table 3. Flatness and symmetry for 10 × 10 cm2 FS.

Table 4. Flatness and symmetry for 15 × 15 cm2 FS.

Table 5. Flatness and symmetry for 20 × 20 cm2 FS.

Table 6. Flatness and symmetry for 20 × 20 cm2 FS diagonal Profile.

5. Conclusions

The process of commissioning an LINAC for clinical use includes comprehensive measurements of dosimetric parameters where a full set of data is acquired that will be used for patient treatment planning. The essential parameters for single energy that include percentage depth dose (PDD), Open Beam Profiles (Inline, Crossline & Diagonal), and Beam Output factors for a series of Field Sizes (FSs) are the essential dosimetric data that require during the commissioning period of a medical linear accelerator (LINAC). Since the outcome of radiation treatment is directly related to the precession in the delivered dose to the patient, it depends on the accuracy of beam data used in the treatment planning process. Therefore, to operate an LINAC for treatment purposes, the measurement of percentage depth (PDD) is very crucial. The differences of flatness and symmetry contamination at 6 MV beams were less than ±2% to ±3%, respectively.

We can also analyze the golden beam data (machine first beam data which is measured by the manufacture company at the time of machine assemble) only if we had it. Then we can compare it with the measured beam data and can investigate any disagreement among them. The photon beam dosimetry data is found in current study among the linear accelerators with the same model and nominal energy. All tolerances were within the clinically acceptable tolerance limit. From clinical point of view, this study will support the radiation oncology department to avoid any delay or interruption in the treatment of patients. The PDD, Dose profile, Output Factors, beam flatness and symmetry data obtained for 6 MV photon energies in this work will be useful for initial commissioning of LINAC system. Its use for minimum error calculation and greater accuracy will be investigated in future works.

Acknowledgements

The authors would like to thank all the personnel of the Institute of Nuclear Medical Physics (INMP), Bangladesh Atomic Energy Commission for allowing us to use the VARIAN Clinac iX and give necessary support for this work.

Conflicts of Interest

The authors declare no conflicts of interest regarding the publication of this paper.

References

[1] Linear Accelerator. http://www.radiologyinfo.org/en/info.cfm?pg=linac
[2] Mayles, P., Nahum, A. and Rosenwald, J.C. (2007) Handbook of Radiotherapy Physics Theory and Practice. CRC Press, Boca Raton, 452.
https://doi.org/10.1201/9781420012026
[3] Chang, D.S., lasley, F.D., DAS, I.J., Mendonca, M.S. and Dynlacht, J.R. (2014) Dosimetry of Photon Beams in Water. In: Basic Radiotherapy Physics and Biology, Springer, New York, 77.
https://doi.org/10.1007/978-3-319-06841-1_8
[4] International Atomic Energy Agency (2000) Absorbed Dose Determination in External Beam Radiotherapy. Technical Reports Series No. 398, Vienna, 110.
[5] Das, I.J., et al. (2008) Accelerator Beam Data Commissioning Equipment and Procedures. Medical Physics, 35, 4186-4215.
https://doi.org/10.1118/1.2969070
[6] Tuathan, O. (2011) Monte Carlo Simulation of Medical Accelerator Electron Treatment Heads. Thesis, School of Physics, National University of Ireland Galway, Ireland.
[7] Podgorsak, E.B. (2010) Radiation Physics for Medical Physicists. Springer, Heidel berg, 277.
https://doi.org/10.1007/978-3-642-00875-7
[8] Kim, S., Liu, C.R., Zhu, T.C. and Palta, J.R. (1998) Photon Beam Skin Dose Analyses for Different Clinical Setups. Medical Physics, 25, 860-866.
https://doi.org/10.1118/1.598261

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.