TITLE:
Diphenhydramine Compact-Cell Sensor
AUTHORS:
Mohsen M. Zareh, Mohammed I. ALahmdi, Ali A. Keshk
KEYWORDS:
Compact Cell, Diphenhydramine Determination, Drug Sensor
JOURNAL NAME:
Journal of Sensor Technology,
Vol.6 No.1,
March
31,
2016
ABSTRACT: The construction and performance characteristics of a novel (diphenhydramine) DPH CC-sensor
based on DZCE were reported in this work. The CC-membrane was prepared by incorporating of
(diazacrown ether) DZCE and/or (tetraflorophenyl borate) TFPB into a plasticized poly(vinyl
chloride) membrane. The CC-sensor revealed a Nernstian behavior over a DPH-concentration
range (9.1 × 10-3 - 6.3 × 10-7 mol ·L-1), and relatively low detection limit (1 × 10-7 mol ·L-1). The potentiometric
response was independent on the pH of the solution in the range of 7 - 10. The
DPH-CC showed a very short response time (