ABSTRACT

Based on electric conductivity and wide potential window of ionic liquid (IL) and electric property of single-wall car- bon nanotubes (SWCNTs), composite material of IL-SWCNTs was prepared, glucose sensor was built with this mate-rial for immobilizing glucose oxidase (GOx). It showed good response, sensitivity and stability for long time for glu-cose detection. Linear range for the detection of glucose was from 0.5 × 10–6 M to 12 × 10–6 M while detection limit was 6.26 × 10–8 M (S/N = 3).

Share and Cite:

Wang, W. , Yin, G. , Ma, X. and Wan, J. (2012) Simple Method for Preparing Glucose Biosensor Based on Glucose Oxidase in Nanocomposite Material of Single-Wall Carbon Nanotubes/Ionic Liquid. Journal of Analytical Sciences, Methods and Instrumentation, 2, 54-59. doi: 10.4236/jasmi.2012.22011.