Commissioning of TrueBeamTM Medical Linear Accelerator: Quantitative and Qualitative Dosimetric Analysis and Comparison of Flattening Filter (FF) and Flattening Filter Free (FFF) Beam

HTML  XML Download Download as PDF (Size: 2836KB)  PP. 51-69  
DOI: 10.4236/ijmpcero.2016.51006    9,127 Downloads   18,550 Views  Citations

ABSTRACT

Motive of the study is to present quantitative and qualitative analysis and comparison of beam data measurement with FF (flattening filter) and FFF (flattening filter free) beam in a Varian TrueBeamTM Medical Linear Accelerator. Critique of beam characterization and evolution of dosimetric properties for 6 MV, 10 MV, 15 MV FF beam and 6 MVFFF, 10 MVFFF FFF beam has been carried out. We performed the comparison of photon beam data for two standard FF photon energy 6 MV, 10 MV verses 6 MVFFF, and 10 MVFFF FFF beam. Determination and comparison of parameter involved PDD (Percentage depth dose), Depth dose profile, Symmetry, Flatness, Quality index, Relative output factor, Penumbra, Transmission factor, DLG (Dosimetric leaf gap), in addition to degree of Un-flatness and off-axis ratio of FFF beam. Outcomes of presenting study had shown that change of various parameters such as Percentage depth dose curves, Shape of the depth dose profile, Transmission, Value of quality index and significant rise in surface dose for FFF in comparison with FF beam. Differences in the output factor at lower and higher field sizes for FFF beam compared to that of FF beam were found. The maximum output factor deviation between 6 MV and 6 MVFFF was found to be 4.55%, whereas in 10 MV and 10 MVFFF was 5.71%. Beam quality TPR20/10 for FFF beam was found to be lesser in magnitude, 5.42% for 6 MVFFF whereas 4.50% for 10 MVFFF compared to 6 MV and 10 MV FF beam respectively. Jaw transmission and interleaf leakage for FFF beam were found to be lesser than FF beam. Also DLG for FFF beam was found to be lesser in magnitude comparable to that of flattened beam. This study is mainly inclined towards evaluation and comparison of the FF and FFF beam. It has been observed that, the outcome of a commissioning beam data generation fully complies with vendor specification and published literature.

Share and Cite:

Shende, R. , Gupta, G. , Patel, G. and Kumar, S. (2016) Commissioning of TrueBeamTM Medical Linear Accelerator: Quantitative and Qualitative Dosimetric Analysis and Comparison of Flattening Filter (FF) and Flattening Filter Free (FFF) Beam. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 5, 51-69. doi: 10.4236/ijmpcero.2016.51006.

Cited by

[1] Physical characteristics of the novel transmission-target X-ray equipment for kilovoltage radiation therapy applications
Radiation Physics and …, 2022
[2] Performance of Radiation Dosimeters in Gradient Regions at Different Dose Rates of Linear Accelerators
Iranian …, 2022
[3] Implementación de tratamientos de radiocirugía en un acelerador lineal clínico de un hospital público de mediana y alta complejidad en el Valle del Cauca, 2021
2022
[4] The Commissioning and Validation of EclipseTM Treatment Planning System on a Varian VitalBeamTM Medical Linear Accelerator for Photon and Electron Beams
2021
[5] Evaluation of Dosimetric Effect of Millennium and High Definition Multi Leaf Collimator Using 6 MV Photon Beams
Asian Pacific Journal …, 2021
[6] Monte Carlo Simulation-Based BEAMnrc Code of a 6 MV Photon Beam Produced by a Linear Accelerator (LINAC)
Atom Indonesia, 2021
[7] Commissioning Measurements of Flattening Filter and Flattening Filter Free Photon Beams Using a TrueBeam Stx® Linear Accelerator
2021
[8] Commissioning of the TrueBeam STx 6 MV FFF Beam in the RayStation Treatment Planning System for SRS and SBRT Treatments
2021
[9] 北京地方所属外照射放射治疗设备输出剂量调查分析
2020
[10] Simplified sigmoidal curve fitting for a 6 MV FFF photon beam of the Halcyon to determine the field size for beam commissioning and quality assurance
2020
[11] Simplified sigmoidal curve fitting for 6 MV FFF photon beam of Halcyon to determine field size for beam commissioning and quality assurance
2020
[12] Commissioning of Clinac IX Trilogy Linear Accelerator for Stereotactic Radiosurgery
2020
[13] Dosimetric characteristics of 6 MV photons from TrueBeam STx medical linear accelerator: simulation and experimental data
Nuclear Science and Technology, 2019
[14] Quality Assurance of Three Dimensional Dosimetry for Radiation Applications in Medicine
2019
[15] Enhancement of the Dose on 12 MV Linac with Free Flattening Filter Mode
2019
[16] A STUDY ON DEPTH DOSE OF FLATTENED AND FLATTENING FILTER FREE PHOTON BEAM OF MILLENNIUM TRUE BEAM LINEAR ACCELERATOR USED FOR …
2019
[17] Experimental dosimetry of 3D-CRT breast radiotherapy modalities, sIMRT and IMAT in chest phantom
2019
[18] Dosimetria experimental de modalidades de radioterapia de mama 3D-CRT, sIMRT e IMAT em fantoma de tórax
2019
[19] Verification of Mechanical Leaf Gap Error and VMAT Dose Distribution on Varian VitalBeam™ Linear Accelerator
Progress in Medical Physics, 2018
[20] Open beam dosimetric characteristics of True Beam medical linear accelerator with flattening filter (WFF) and flattening filter free (FFF) beam
Polish Journal of Medical Physics and Engineering, 2018
[21] A dosimetric analysis of proton beam therapy using different snouts
Journal of Radiotherapy in Practice, 2018
[22] Commissioning and evaluation of a radiochromic EBT3 film dosimetry system
Journal of Radiotherapy in Practice, 2018
[23] Development of an Analytic Software Using Pencil Beam Scanning Proton Beam
2017
[24] TureBeam low dose rate investigation for pulsed reduced dose rate IMRT
2017
[25] TrueBeam Low Dose Rate Investigation for Pulsed Reduced Dose Rate IMRT
2017
[26] Dosimetric comparison of photon beam profile characteristics for different treatment parameters
Journal of Radiotherapy in Practice, 2017
[27] Validation of Dosimetric Leaf Gap (DLG) prior to its implementation in Treatment Planning System (TPS): TrueBeam™ millennium 120 leaf MLC
Reports of Practical Oncology & Radiotherapy, 2017
[28] Evaluation of Dosimetic Characteristics of Electronic Portal Imaging Device (EPID) for Flattening Filter Free (FFF) Beams
2017
[29] 利用游離腔, 二極體及 EBT3 軟片驗證 Elekta Versa 直線加速器之百分深度劑量與剖面劑量
2017
[30] Comisionamiento de acelerador varian TrueBeam en modalidad fotones.
2016

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.