ABSTRACT

This paper summarizes preliminary rat studies aimed at identifying the effectiveness of using electrically stimulated silver as a bactericidal agent for indwelling residual hardware devices (RHD). A variety of bactericidal indwelling devices were designed, fabricated and surgically inserted into the medullary cavity of live rats. The rats were inoculated with Staphylococcus aureus to try and induce osteomyelitis. A total of 37 surgeries were performed by implanting the rats with both control and potentially bactericidal devices. As surgical procedures and devices were improved, it appeared that the implants produced antibiotic effects in the animals. All of the control animals and all of the animals where the device failed tested positive for Staphylococcus aureus growth. Of the rats with operational bactericidal devices (those that survived the surgery and incubation period), half tested negative for Staphylococcus aureus. The device designs are discussed in this paper along with the test procedures, operating practices and results. A statistical analysis of the results, which shows a very high confidence level in the effectiveness of electrically stimulated silver as a bactericidal agent/antibiotic, is also presented.

Share and Cite:

Wysk, R. , Sebastianelli, W. , Shirwaiker, R. , Bailey, G. , Charumani, C. , Kennett, M. , Kaucher, A. , Abrahams, R. , Fuller, T. , Royer, P. , Voigt, R. and Cohen, P. (2010) Prophylactic Bactericidal Orthopedic Implants – Animal Testing Study. Journal of Biomedical Science and Engineering, 3, 917-926. doi: 10.4236/jbise.2010.39122.

Cited by

[1] Antibacterial efficacy and cytotoxicity of low intensity direct current activated silver–titanium implant system prototype
BioMetals, 2017
[2] Effects of Electrically Activated Silver–Titanium Implant System Design Parameters on Time-Kill Curves Against Staphylococcus aureus
Journal of Medical and Biological Engineering, 2016
[3] Modified Pharmacokinetic/Pharmacodynamic model for electrically activated silver-titanium implant system
2015
[4] Effects of cathode design parameters on in vitro antimicrobial efficacy of electrically-activated silver-based iontophoretic system
Journal of Materials Science: Materials in Medicine, 2015
[5] An Antimicrobial Dual-metal Implant System Activated by Low Intensity Direct Current.
2015
[6] Nanomaterials and synergistic low‐intensity direct current (LIDC) stimulation technology for orthopedic implantable medical devices
Wiley Interdisciplinary Reviews: Nanomedicine and Nanobiotechnology, 2013
[7] Determining Optimal Current Intensity and Duration for Electrically Activated Silver-Based Prophylactic Hip Implant Prototype Design
ASME 2013 Summer Bioengineering Conference. American Society of Mechanical Engineers, 2013
[8] Evaluation of a Multi-metal Approach to Iontophoretic Antibacterial Prophylaxis and Treatment in Orthopaedic Implants.
2013
[9] The characterization of the antibacterial efficacy of an electrically activated silver ion-based surface system
2011

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.