Direct Analytical Method to Calculate Photopeak Efficiency and Photopeak Attenuation Coefficient of NaI(Tl) Well-Type Detector

HTML  XML Download Download as PDF (Size: 614KB)  PP. 115-124  
DOI: 10.4236/wjnst.2016.62012    3,418 Downloads   5,064 Views  

ABSTRACT

In this paper full-energy peak (photopeak) efficiency and photopeak attenuation coefficient of 3'' × 3'' NaI(Tl) well-type scintillation detector were calculated using gamma-rayisotropic radiating point sources (with photon energy: 0.245, 0.344, 0.662, 0.779, 0.964, 1.1732, 1.333 and 1.408 MeV) placed outside the detector well. These energies were obtained from 152Eu, 137Cs and 60Co. The relations between the full energy peak efficiency and photopeak attenuation coefficients, were plotted vs. photon energy at different sources to detector distance, and it found that the full energy peak efficiency decreased by increasing the distance between the source and the detector.

Share and Cite:

Al-Mugren, K. , Abbas, M. , El-Bayoumi, E. and Aly, N. (2016) Direct Analytical Method to Calculate Photopeak Efficiency and Photopeak Attenuation Coefficient of NaI(Tl) Well-Type Detector. World Journal of Nuclear Science and Technology, 6, 115-124. doi: 10.4236/wjnst.2016.62012.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.