
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2013, 2, 15-18
Published Online February 2013 (http://www.scirp.org/journal/ijmpcero)
http://dx.doi.org/10.4236/ijmpcero.2013.21003
Copyright © 2013 SciRes. IJMPCERO
Technical Note: The Uses of I’mRT MatriXX in Electron
Beams
Mutian Zhang, Sicong Li, Hua Deng, Sumin Zhou
Department of Radiation Oncology, University of Nebraska Medical Center, Omaha, USA
Email: mutianzhang@unmc.edu
Received Ocotber 20, 2012; revised November 22, 2012; accepted November 30, 2012
ABSTRACT
Purpose: The objective of this study is to investigate the properties of I’mRT MatriXX device in electron beams, and to
validate MatriXX in electron dosimetry and quality assurance (QA). Methods: The measurements were conducted us-
ing MatriXX in electron and photon beams from Siemens linacs. The MatriXX was placed horizontally on the linac
tabletop. Solid Water layers were used for buildup. For all the measurements, the linac gantry angle was 0˚, and the
source-to-surface distance was 100 cm from the Solid Water surface. The electron cone factors, cutout factors, and
beam profiles were measured and compared with thimble ionization chamber results. Results: The effective water
equivalent depth of MatriXX measurement point is larger than 4 mm. When measuring at the respective depths of
maximum dose, MatriXX has different responses to different beam energies. The cone factors measured by MatriXX
are nearly identical or close to those derived by ionization chambers. Beam profiles (flatness and symmetry) can be eas-
ily determined using MatriXX and are comparable to water tank results. The planar dose map of electron cutout blocks
can be visually observed, and the cutout factors can be conveniently measured. Conclusions: The MatriXX needs
separate dose calibration factors for electron and photon beams. MatriXX can be used to measure electron cutout factors
and beam profiles, thus has the potentials in electron beam dosimetry and routine linac and patient-specific QA tests.
Keywords: Electron Beam; MatriXX; Dosimetry; Quality Assurance
1. Introduction
The I’mRT MatriXX (IBA Dosimetry GmbH, Germany)
device consists of a two-dimensional (2D) array of ioni-
zation chambers. There are 1,020 vented parallel plate
ion chambers on the array detector, arranged in 32 × 32
grid. The chamber center-to-center distance is 7.62 mm,
and the active area is 24.4 × 24.4 cm2. MatriXX has been
validated for 2D dose measurements [1], and is increas-
ingly used in photon beam dosimetry and patient-specific
quality assurance (QA) [2-4]. The application of Ma-
triXX is also extended to QA checks for proton therapy
[5].
In this work, we report our investigation on the feasi-
bility of using MatriXX in electron beam dosimetry and
routine linac QA or patient-specific treatment QA. This
note is the expansion of an abstract submitted to the 2010
American Association of Physicists in Medicine annual
meeting [6].
2. Methods
The measurements were conducted using electron and
photon beams from Siemens ONCOR and PRIMUSTM
linacs (Siemens Healthcare, Germany). The linacs could
produce 6 MV and 23 MV photons, and electron beams
with multiple energies between 5 MeV and 21 MeV. The
MatriXX was placed horizontally on the linac treatment
couch, supported by 5 or 6 cm Solid Water (Gammex
Inc., USA) blocks (Figure 1). The MatriXX was posi-
tioned using the linac light field. On the MatriXX surface,
30 cm × 30 cm Solid Water layers served as beam
buildup with 1 mm thickness resolution. The linac gantry
angle was 0˚, and the source-to-surface distance was set
at 100 cm from the Solid Water surface.
The MatriXX was previously calibrated for the photon
beams. Before each use, the MatriXX was powered on
for 30 minutes, and irradiated with at least 500 cGy until
stable readings were achieved. For each reading, 100
monitor units were delivered. The measurements of each
data point were repeated three times. When taking the
readings, the calibration factor of 6 MV photon beam
was used, so that the MatriXX responses to different
beam energies could be compared. The chamber array
was placed at the depth of maximum dose (dmax) of the
corresponding beam energy for the measurements of
dose response, cone factor or cutout factor. The MatriXX
measurements were compared with beam data acquired
using calibrated PTW semiflex thimble chambers (PTW