Unexpected Increase in Arterial to End-Tidal CO2 Gradient in a Child Undergoing Embolization of MAPCAs

Saifeldin A. Mahmoud1*, Khaled A. Sedeek2, Patrick M. McQuillan2

1Department of Anesthesiology, Albert Einstein Medical Center, Philadelphia, PA, USA

2Department of Anesthesiology and Perioperative Medicine, Penn State College of Medicine/Hershey Medical Center, Hershey, PA, USA

Copyright © 2017 by authors and Scientific Research Publishing Inc.

This work is licensed under the Creative Commons Attribution International License (CC BY 4.0).

http://creativecommons.org/licenses/by/4.0/

Received: February 7 2017; Accepted: March 26, 2017; Published: March 29, 2017

ABSTRACT

A 17-month-old infant with multiple aorto-pulmonary collateral arteries (MAPCAs) and pulmonary hypertension presented for diagnostic catheterization. On the day of the procedure, the infant was asymptomatic with oxygen saturation in the 90’s on 1.0 L/min O2 nasal cannula. His parents denied any recent illness. During the procedure, one coil was inadvertently embolized into the right lung resulting in markedly increased pulmonary artery pressures. The Pa-etCO2 gradient increased to 25 mmHg from a baseline of 2 mmHg. Therapy was initiated to reduce the PaCO2. The patient could not be weaned from mechanical ventilation due to elevated PA pressures.

Keywords:

Multiple Aorto-Pulmonary Collateral Arteries (MAPCAs), Pa-etCO2 Gradient, Ventilation-Perfusion (V/Q) Relationship

1. Introduction

The difference [gradient] between the partial pressure of carbon dioxide in arterial blood (PaCO2) from that measured at the end of expiration (etCO2) is a result of the relationship between ventilation (V) and perfusion (Q) of the lung. There is normally a gradient of 2 - 5 mm Hg [1] . When lung perfusion is impaired, the V/Q will be mismatched and cause the gradient to increase. The PaCO2 therefore, will be much higher than expected based on the etCO2. The consequence of this is a potentially unrecognized hypercarbia and respiratory acidosis. Written consent has been obtained to submit this case for publication.

2. Case Presentation

A 17-month-old about 9.4 kg male presented for diagnostic cardiac catheterization. The infant was born with Tetralogy of Fallot, moderate bilateral pulmonary artery hypoplasia, and multiple aorto-pulmonary collateral arteries (MAPCAs). A Blalock-Taussig shunt, ligation of single aortic to pulmonary artery (PA) collateral, surgical placement of RA to focalized PA conduit, and creation of fenestrated VSD were performed at six months of age.

On this admission, he had significant PA hypertension and RV systolic dysfunction. His TTE demonstrated mild RV dilation with mildly depressed systolic function. Estimated RV systolic pressure was 75 mmHg. Pre-operatively, his medication includes aspirin, digoxin, enalapril, furosemide, spironolactone and sildenafil. Cardiac catheterization was initially planned to only evaluate the response of PA pressures to sildenafil, which was begun 2 weeks earlier.

On the day of the procedure, the infant was asymptomatic. His O2 saturation was in the 90’s on a baseline of 1.0 L/min of O2 via a nasal cannula. His parents denied any recent fever or illness. Physical examination revealed no respiratory distress, wheezing, or cyanosis.

Induction of general anesthesia was achieved with inhalation of sevoflurane in 100% O2. A 22 g PIV was established, 25 mcg Fentanyl and 10 mg rocuronium were administered IV, and the airway was easily secured with a 4.5-cuffed ETT. Endotracheal tube position was confirmed with equal bilateral breath sounds, positive etCO2 and fluoroscopy by the cardiologist (Figure 1). General anesthesia was maintained with sevoflurane (1MAC). PCV mode was used with a PIP of

Figure 1. Cardiac image before coil embolization.

15 cm H2O, RR of 20/min and FiO2 of 0.21. With these ventilator settings, a VT of 100 ml was delivered and the etCO2 was 33 mm Hg. The initial ABG values were: pH 7.45, PaCO2 35 mmHg, PaO2 50 mmHg, and 24 mEq. The Pa-etCO2 gradient was 2 mm Hg.

During the procedure, multiple coils were placed in collaterals from the aorta to the left lung. One of the coils was inadvertently embolized into the right lung (Figure 2). Co-incident with the coil embolizing to the right lung, PA pressures markedly increased. An ABG at that time demonstrated a pH 7.29, PaCO2 62 mmHg, PaO2 423 mmHg, and HCO330 mEq. The etCO2 at that time was 37 mmHg with a Pa-etCO2 gradient of 25 mmHg. To reduce the PaCO2, hyperventilation was initiated with PIP of 22 cm H2O and RR of 24/min. A PaCO2 within the normal range was only attained when the etCO2 was reduced to 22 mmHg (Table 1).

Figure 2. Cardiac image after coil embolization. The circle depicts the coil in the right lung.

Table 1. PaCO2, end tidal CO2 and Pa-etCO2 gradient in mmHg in relation to the coil embolization. Note the gradient difference 30 minutes before the embolization and immediately after.

At the end of the procedure, the patient could not be weaned from mechanical ventilation. Despite the use of nitric oxide, PA pressures and the PaCO2-etCO2 gradient remained elevated. The decision was made to keep the ETT in place, and the infant was transported to the PICU for further management. Transthoracic echocardiography performed immediately after the procedure demonstrated moderate RV dilation with moderately depressed systolic function and paradoxical inter-ventricular septal motion. RV systolic pressure was supra-sys- temic with estimates of 123 mmHg. Further management of right heart failure and pulmonary hypertension was initiated with a combination of diuretics, oxygen, and a milrinone infusion.

3. Discussion

Based on the complexity of the patient’s medical condition and the nature of the procedure, this case report highlights the significance of a sudden intraoperative increase in the Pa-etCO2 gradient. This should alert the anesthesiologist to the possibility of a pulmonary blood flow incident occurred. According to Nunn and Hill [2] , the Pa-etCO2, is 4.6 ± 2.5 mm Hg in healthy, anesthetized patients. Because end-expiratory gas is a mixture from both well and poorly perfused alveoli, end-tidal PCO2 is usually less than arterial PCO2. The Pa-etCO2 gradient can be used as an index of alveolar dead space in most cases. In this case, embolization of a coil to the right lung acutely decreased the perfusion (Q) in the ventilation-perfusion (V/Q) relationship, while the ventilation (V) remained unchanged. This results in increased dead space ventilation. Using the etCO2 alone to predict the PaCO2 in such cases can lead to underestimation of the PaCO2.

Baraka el al. 1994 reports that in ASA class 1 or 2 patients during laparoscopy, no significant change in the PaCO2-etCO2 gradient after carbon dioxide insufflation. They conclude that end-tidal capnography can be used as noninvasive techniques for monitoring arterial carbon dioxide elimination during laparoscopic surgery in healthy patients [3] . The end-tidal PCO2 increases hand-in-hand with the arterial PCO2 during laparoscopic surgery [4] . Thus, end-tidal PCO2 accurately reflects the changes of the arterial PCO2 during laparoscopy. Patients with severe cardiopulmonary disease can develop significant hypercarbia associated with an increased PaCO2-etCO2 gradient [5] . In these patients, the end-tidal PCO2 is not a satisfactory index of PaCO2 measurement [6] .

The reliability of using end-tidal PCO2 as an indicator for arterial PCO2 is highly controversial. Some studies have found that the gradient is highly variable and PetCO2 cannot provide a reliable reflection of PaCO2, as in patients during craniotomies [7] , and in multisystem trauma patients [8] .

4. Conclusion

Using the etCO2 alone to predict the PaCO2 during an embolization procedure is not ideal. This case emphasizes the importance of intraoperative communication between the anesthesiologist and the cardiologist.

Cite this paper

Mahmoud, S.A., Sedeek, K.A. and McQuillan, P.M. (2017) Unexpected Increase in Arterial to End- Tidal CO2 Gradient in a Child Undergoing Embolization of MAPCAse. Open Journal of Anesthesiology, 7, 63-67. https://doi.org/10.4236/ojanes.2017.73006

References

  1. 1. Fletcher, R. and Jonson, B. (1984) Dead Space and the Single-Breath Test for Carbon Dioxide during Anaesthesia and Artificial Ventilation. British Journal of Anaesthesia, 56, 109-119.
    https://doi.org/10.1093/bja/56.2.109

  2. 2. Nunn, J. and Hill, D. (1960) Respiratory Dead Space and Arterial to End Tidal CO2 Tension Difference in Anesthetized Man. Journal of Applied Physiology, 15, 383-389.

  3. 3. Baraka, A., Jabbour, S., Hammoud, R., Aouad, M., Najjar, F., Khoury, G. and Sibai, A. (1994) Can Pulse Oximetry and End-Tidal Capnography Reflect Arterial Oxygenation and Carbon Dioxide Elimination during Laparoscopic Cholecystectomy? Surgical Laparoscopy Endoscopy & Percutaneous Techniques, 4, 353-356.

  4. 4. Puri, G. and Singh, H. (1992) Ventilatory Effects of Laparoscopy under General Anesthesia. British Journal of Anaesthesia, 68, 211-213.
    https://doi.org/10.1093/bja/68.2.211

  5. 5. Wittgen, C., Akdrus, C., Fitzgerald, S., Baudendistel, L., Dahms, T. and Kaminski, D. (1991) Analysis of Hemodynamic and Ventilatory Effects of Laparoscopic Cholecystectomy. Archives of Surgery, 120, 997-1001.
    https://doi.org/10.1001/archsurg.1991.01410320083011

  6. 6. Wahba, R. (1993) Ventilatory requirements during laparoscopic cholecyctectomy. Canadian Journal of Anesthesia, 40, 206-210.
    https://doi.org/10.1007/BF03037031

  7. 7. Russell, G. and Graybeal, J. (1995) The Arterial to End-Tidal Carbon Dioxide Difference in Neurosurgical Patients during Craniotomy. Anesthesia & Analgesia, 8, 806-810.

  8. 8. Russell, G. and Greybeal, J. (1994) Reliability of the Arterial to End-Tidal Carbon Dioxide Gradient in Mechanically Ventilated Patients with Multisystem Trauma. Journal of Trauma and Acute Care Surgery, 36, 317-322.
    https://doi.org/10.1097/00005373-199403000-00006