Comparison of Depth Dose Distributions Using Cerenkov Fiber-Optic Dosimeter and Monte Carlo Simulation for HDR Brachytherapy

HTML  XML Download Download as PDF (Size: 570KB)  PP. 1499-1502  
DOI: 10.4236/jamp.2016.48157    1,570 Downloads   2,369 Views  

ABSTRACT

In this study, we fabricated a Cerenkov fiber-optic dosimeter (CFOD) without any scintillator to measure Cerenkov radiation signals owing to gamma-rays. The relative depth dose (RDD) distributions of Ir-192 HDR brachytherapy source were obtained by using the CFOD based on a subtraction method and the RDD curve was compared with the simulation result of Monte Carlo N-particle extended transport code (MCNPX). Finally, we demonstrated that the CFOD can be used to measure real-time dose information for HDR brachytherapy.

Share and Cite:

Shin, S. , Song, Y. , Kim, M. , Kim, H. , Yoo, W. , Jang, K. and Lee, B. (2016) Comparison of Depth Dose Distributions Using Cerenkov Fiber-Optic Dosimeter and Monte Carlo Simulation for HDR Brachytherapy. Journal of Applied Mathematics and Physics, 4, 1499-1502. doi: 10.4236/jamp.2016.48157.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License