Journal of Minerals & Materials Characterization & Engineering, Vol. 8, No.3, pp 171-179, 2009
jmmce.org Printed in the USA. All rights reserved
171
Three-dimensional Inspection and Characterization of Plastic Micro Mixers
Fabricated by Different Packaging Techniques
R. Y. Zhang, K. Chen
*
, L. X. Wang, W. J. Zhu
Department of Materials and Energy Engineering, Guongdong University of Technology,
Guongzhou 510090, China.
*
Department of Mechanical Engineering, University of Utah, Salt Lake City, UT 84112, USA
(Ph. No.-01-801-5814150. E-mail: chen@mech.utah.edu)
ABSTRACT
An X-ray microtomography system was used for inspecting and characterizing the packaging of
micro mixers of planar configuration. Two types of micro mixers were machined from acrylic
plate by means of precision milling, and four different packaging techniques were tested for
mixer assembly: adhesive bonding, plasma bonding, screw-bolting, and xurography with double-
sided tape. The interiors of the mixers were thoroughly examined by navigating the three-
dimensional images generated by the micro CT system. It was found that plasma bonding did
not work well on acrylics. The boundaries between the acrylic plate and the plastic tape used in
tape bonding were almost indistinguishable in the X-ray images because the densities of these
two materials are very close. Of the different packaging techniques we tested tape bonding
provided good sealing without the drawback of glue or sealing materials smeared into the
mixing chamber or flow channels.
Key Words : Characterization of plastic packaging, Micro-CT (Computed Tomography), Micro
mixer
1. INTRODUCTION
X-ray based technologies have been commonly used for the inspection and characterization of a
variety of natural and man-made materials and devices. For instance, Stojanovic used the X-ray
powder diffraction technique to determine the structural and microstructural parameters of B-
doped PbTe semiconductor [1]. X-ray diffraction patterns of iron oxide powders were analyzed
172 R. Y. Zhang, K. Chen, L. X. Wang, W. J. Zhu Vol.8, No.3
in Shi and Hwang’s investigation of microwave-assisted miniaturization of chemical plants [2].
X-ray micro CT (Computed Tomography) is a very popular technique for nondestructive
inspection and characterization of structures, materials, as well as microelectronics and MEMS
devices. Nano-focus X-ray sources are commercially available in recent years, and high-
resolution Nano-CT has been developed [3] for materials research [4].
Micro mixers of planar configuration are the main components of today’s micro sensors and
micro biological, medical, and chemical reactors/analyzers, which can be batch-fabricated and
easily combined with microelectronics using the integrated circuit technology. Mixing of two or
more fluid streams is also encountered in other MEMS or microfluidic devices such as micro
pumps and valves and micro heat engines. Plastics are the favorable materials for micro mixers.
They cost less and are easier to handle and use than most other materials used in the fabrication
of microfluidic devices and components. Most plastics are transparent, thus flow visualization is
easy. This feature is important to medical devices for which easy and continuous monitoring of
fluid flows are often required. A transparent shell also makes it convenient to inspect the mixer
interiors, detect leakages, and study the flow inside the mixer.
Different plastic materials have been tested and a variety of packaging and assembly techniques
developed for the fabrication of plastic micro mixers. At present adhesive bonding and hot
embossing are the popular packaging techniques for plastic micro mixers. Innovative designs
and methods (e.g., [5-9]) have been developed to improve bonding strengths and conditions,
prevent adhesives and sealing materials smeared into the mixing chamber or flow channels,
facilitate the packaging process, and reduce deformation during packaging. Plasma bonding is
another technique that has been applied to the packaging of plastic microchips and microfluidics
fabricated from plastics as well as glass and metals. Double-sided tapes were used to bond two
glass slides in Greer et al’s manufacturing of microfluidic channels [5].
The machined or etched surfaces in plastic mixers are often opaque, prohibiting visual inspection
with a magnifier or optical microscope. Of the 3 nondestructive methods (IR transmission
imaging, X-ray tomography, and acoustic microscopy) currently used to inspect bonded devices
[10], X-ray imaging has gained increasing popularity for measuring and characterizing internal
features and flaws of MEMS and microfluidics. However detailed information (e.g., [11]) about
the application of X-ray CT for the inspection and characterization of microfluidics is rare. This
motivated the present investigation of 3D (three-dimensional) X-ray images of micro mixers
fabricated by different packaging techniques.
2. FABRICATION AND PACKAGING OF MICRO MIXERS
Two types of micro mixers of planar configuration were fabricated in the present investigation: a
T-shaped mixer and a swirling-flow mixer that involves a circular mixing chamber with two
tangential inlets and an axially directed outlet. Photos of these two micro mixers are given in
Vol.8, No.3 Three-dimensional Inspection and Characterization 173
figures 1 and 2. The pink powder in the photos is the dried red ink (Rhodamine) used in our flow
visualization experiment. The rectangular mixing chamber of the T- mixer was fabricated out of
a thin acrylic sheet and glued to another acrylic sheet to enable the mixing process to be observed
easily. Two inlet and one exit channels were drilled into the mixing chamber after the two
acrylic plates had been glued together. The diameter of the flow channels was approximately 0.4
mm. The mixing chamber was approximately 20 mm long and 4 mm wide, and its thickness was
0.4 mm before packaging.
Figure 1. A T-shaped micro mixer fabricated from thin acrylic sheets.
Figure 2. A swirling-flow micro mixer with two tangential inlets and an axially directed exit channel.
Inlet
channels
Exit
channel
x
y
z
Mixer exit
Mixing
chamber x
y
z
174 R. Y. Zhang, K. Chen, L. X. Wang, W. J. Zhu Vol.8, No.3
The swirling-flow mixer was also fabricated from acrylic sheets for easy observation of the
mixing process. The circular mixing chamber and two inlet channels were micro machined on
one acrylic plate. The exit flow channel was drilled through another acrylic plate. The mixing
chamber was approximately 0.4 mm deep before assembly and its diameter was 5 mm. The two
rectangular inlet channels were 1 mm wide and 1.5 mm deep at the mixing chamber rim. At
about 2 mm upstream of the inlet ports the dimensions of the inlet channels increased to 2 mm by
2 mm for reduced frictional loss. The diameter of the exit channel was approximately 0.4 mm.
The surfaces of the mixing chambers of the 2 mixers were inspected using a SEM before
packaging, and the roughness of the surfaces was found to be in the micron range.
Three different techniques were tested for the assembly of the swirling-flow mixer. The first
method was to bolt the two acrylic plates together with screws. Because the surfaces of
industrial acrylic sheets are extremely smooth, the two acrylic plates pressed against each other
by the screws would make intimate contact. No bonding or sealing was applied to the contact
surfaces of the two acrylic plates.
Plasma bonding was the second technique we tried to bond the two halves of the swirling-flow
mixer. While this technique has been applied successfully to the bonding of a variety of
materials, it failed to yield a good bonding between acrylic plates. Thus plasma bonding is
probably not applicable to the packaging of MEMS or microfluidics fabricated from acrylics.
In the third technique a knife plotter by Graphtec cut the geometry of the contact surface out of
25
µ
m thick, double coated tape by 3M. The tape was then sandwiched between the two acrylic
plates before the plates were tightly bolted with two screws. In the last stage the tape-bonded
acrylic mixer was placed in a 65
o
C oven for 45 minutes to ensure good bonding between the tape
and the acrylic plates. Detailed information about this micro fabrication method can be found in
Greer et al’s paper [5].
3. FACILITY FOR 3D NONDESTRUCTIVE INSPECTION
The instrument used in the present investigation of micro mixer packaging is a custom-designed,
X-ray micro-CT machine, Konoscope 40-130, by ARACOR. A microfocus X-ray tube by Kevex
(Model number KM 13006) was used for X-ray generation with a focus spot of 8
µ
m at 8 Watts,
and 20
µ
m at 16 Watts. The 2D (two-dimensional) cone-beam projections of this micro CT
system have 2048 x 2048 pixels over a 10-mm diameter. The system magnification can be
adjusted by changing the location of the specimen on a computer-controlled vertical/rotary
positioning stage, allowing for the imaging of objects up to 40 mm in diameter. The maximum
material density this micro CT system can handle is about 8 x 10
6
kg/m
3
.
Vol.8, No.3 Three-dimensional Inspection and Characterization 175
The major component of the image detection system is a Photometrics XR300 camera with
Thompson 2048 x 2048, Grade 1 CCD. It provides 16-bit digitization at 2 x 10
5
pixels per
second. A Mac Power PC with LabView and LVCam installed was used to control the
movements of the positioning stage and to acquire the CCD camera images. Three-dimensional
tomographic images were reconstructed using an SGI Octane workstation. The software MIPAV
(version 4) was used to display and navigate the 3D images. Detailed specifications and
description of this custom-designed micro CT system can be found in Lin and Miller’s paper
[12].
4. RESULTS AND DISCUSSION
The 3D images re-constructed by the micro CT system allow a nondestructive inspection of the
interiors of a micro mixer after the mixer has been packaged. The software MIPAV which was
originally developed for medical research makes it convenient to view the scanned images at
different locations from different angles, or to navigate the 3D images for animation making.
Presented in figures 3 to 5 are the orthographic views of the X-ray images of the micro mixers we
fabricated. The orientations of the Cartesian coordinates in these figures are shown in figures 1
and 2. Complete 3D data can be furnished upon request.
The three orthographic views of the T-mixer images at two different locations are shown in
figures 3(a) and 3(b). The three panels from left to right in each figure are the X-ray images on
the X-Y, Y-Z, and X-Z planes. The origins of the local coordinate systems in these two figures
are approximately at the middle points of the height (along the Y axis) and width (along the X
axis) of the mixing chamber. The X-axis passes through the two inlet ports in figure 3(a), and is
slightly upstream of the inlet ports in figure 3(b). The crosshairs in these images are the
projections of the orthographic views in the other two directions. Each pixel in these images is
about 20
µ
m. The dark areas in the images are voids (air), and the white areas are the adhesive.
The gray areas represent the acrylics. Due to the large difference in densities of these three
materials (air, adhesive, and acrylics), the boundaries between different materials are clearly
distinguishable in the X-ray images.
At the locations marked by the crosshairs in figures 3(a) and 3(b) the mixer was well sealed by
the adhesive. The white areas (the adhesive) in the central and right panels in figure 3 are in
good alignment with the mixer contour except for the lower edges of the two inlet ports in the
right panels. On the other hand the bonding was not very good as evidenced by the gaps in dark
color (voids) between the surfaces of the acrylic plate and the adhesive in the two central panels.
The finger-shape white areas attached to inlet ports in the right panels indicate some adhesive
was smeared into the mixing chamber during packaging, and partially blocked the flow
passageway in the neighborhood of the inlet ports. The left panel in figure 3(a) showed one of the
inlet channels was also partially blocked by the adhesive. In the left panel in figure 3(b) the white
176 R. Y. Zhang, K. Chen, L. X. Wang, W. J. Zhu Vol.8, No.3
area on the left has a slanted upper boundary, and the mixing chamber at this location is slightly
narrower due to the smearing of the adhesive into the chamber.
(a)
(b)
Figure 3. Orthographic views of the X-ray images at two different locations in the T-mixer.
X
Y
Z
Y
X
Z
Vol.8, No.3 Three-dimensional Inspection and Characterization 177
Figure 4. Orthographic views of the X-ray image at a selected location in the swirling-flow mixer
packaged by screw bolting alone.
Figure 5. Orthographic views of the X-ray image at a selected location in the swirling-flow mixer
packaged by tape bonding and screw bolting.
The X-ray images of the swirling-flow mixer fabricated by screw bolting are shown in figure 4
with a spatial resolution of about 20
µ
m per pixel. As can be seen in these images, the machined
mixing chamber surfaces are very smooth and flat. The thickness of the mixing chamber is also
fairly uniform. The hump at the mixer outlet in the left and central panels is the dried
Rhodamine ink (the pink powder in figures 1 and 2). It has a density close to that of acrylics. A
small gap – about 40
µ
m in width – between the two acrylic plates can be clearly seen in the left
and central panels in figure 4. Mixing of water and Rhodamine in this mixer was examined
under a bifocal microscope, and no leakage was observed. The experimental observation proved
that the high surface tensions of water and Rhodamine in small acrylic channels prevented
leakage through the small gap between the two acrylic plates. The X-ray images however show
that this packaging is not completely airtight, thus is not suitable for gas mixing.
The images of the swirling-flow mixer fabricated by tape bonding and bolted together with two
screws are shown in figure 5. A pixel size of about 40
µ
m was used in this scanning to allow a
larger volume to be scanned by the micro CT system. It was found from these X-ray images that
178 R. Y. Zhang, K. Chen, L. X. Wang, W. J. Zhu Vol.8, No.3
the tape we used had a density very close to that of acrylics. As a result the double-sided tape
and the two acrylic plates are almost indistinguishable in these images. The adhesive coatings on
the two sides of the plastic tape were compressed and deformed when the two acrylic plates were
pressed against each other by the screws, resulting in good sealing of the mixing chamber and
flow channels and intimate contact between the tape and the acrylic plates.
It is not expected the double-sided tape can hold the two acrylic plates together when the mixer is
subjected to a large pressure difference. The screws in this packaging technique therefore serve
two purposes: to secure together the two acrylic plates and to provide a compression force to the
flexible tape for good sealing. The imperfect matching of the plastic tape and the mixer contour
can be seen in a few places such as the top left corner of the mixing chamber in the left panel in
figure 5. The tape protruded into the mixing chamber is a bit blurry in this X-ray image because
the tape thickness (25
µ
m) is only half of the pixel size (40
µ
m). Scanning with smaller pixels
(which will reduce the size of the scanned volume) is needed for more accurate inspections of the
alignment and matching of the layered structures when tape bonding is used for packaging.
5. CONCLUSION
Different packaging techniques were tested to assemble micro mixers fabricated from acrylic
sheets, and the interiors of the mixers were inspected and characterized using a custom-designed
micro CT system. The image acquisition and processing hardware and software of the X-ray
scanner allowed convenient and thorough 3D inspection of every portion of the mixer. Plasma
bonding did not work well on acrylic plates. The mixer assembled by screw bolting was not
perfectly airtight, thus relying on surface tensions to prevent leakage through the small gap
between the acrylic plates. X-ray images of adhesive bonding showed good contrast between the
acrylics and the glue. Tape bonding provided good sealing, but densities of the tape and acrylics
are very close, making it difficult to distinguish these two plastic materials in X-ray images.
ACKNOWLEDGMENTS
This work was sponsored by the Chinese National Science Council through contract number
50576014. The authors also thank Dr. Chen-Luh Lin and Dr. Bruce Gale for their technical
support.
REFERENCES
[1] Stojanovic, J. N., 2006 “Structural and microstructural characteristics of B-doped PbTe
semiconductor,” Journal of Minerals and Materials Characterization and Engineering 5-2, pp.
143-153.
[2] Shi, S., Hwang, J. Y., 2005 “Miniaturize chemical plant with microwave heating
technology.” Journal of Minerals and Materials Characterization and Engineering 4-1, pp. 61-66.
Vol.8, No.3 Three-dimensional Inspection and Characterization 179
[3] Neuser, E., Suppes, A., 2007 “NanoCT visualizing internal 3D structures with
submicrometer resolution,” DIR 2007 – Internal Symposium on Digital Industrial Radiology and
computed Tomography (June 25 – 27, 2007, Lyon, France), 18 pages.
[4] Brunke, O., Neuber, D., Lehmann, D. K., 2007 “NanoCT: Visualizing of internal 3D-
structures with submicrometer resolution,” Materials Research Society Symposium Proceedings
990 Materials, Processes, Integration and Reliability in Advanced Interconnects for Micro- and
Nanoelectronics, pp. 325-331.
[5] Greer, J., Sundberg, S. O., Wittwer, C. T., Gale, B. K., 2007 “Comparison of glass etching to
xurography prototyping of microfluidic channels for DNA melting analysis,” J. Micromech
Microeng. 17, pp. 2407-2413.
[6] Qi, S., Liu, X., Ford, S., Barrows, J., Thomas, G., Kelly, K., McCandless, A., Lian, K.,
Goettert, J., Soper, S. A., 2002 “Microfluidic devices fabricated in poly(methyl methacrylate)
using hot-embossing with integrated sampling capillary and fiber optics for fluorescence
detection,” Lab on a Chip 2-2, pp. 88-95.
[7] Wang, J., Chen, G., Muck, A., Collins, G. E. 2003 “Electrophoretic microchip with dual-
opposite injection for simultaneous measurements of anions and cations,” Electrophoresis
24(21), pp. 3728-3734.
[8] Dang, F., Tabata, O., Kurokawa, M., Ewis, A. A., Zhang, L., Yamaoka, Y., Shinohara, S.,
Shinohara, Y., Ishikawa, M., Baba, Y., 2005 “High-performance genetic analysis on
microfabricated capillary array electrophoresis plastic chips fabricated by injection molding,”
Anal. Chem. 77-7, pp. 2140-2146.
[9] Kelly, R. T., Pan, T., Woolley, A. T., 2005 “Phase-changing sacrificial materials for solvent
bonding of high-performance polymeric capillary electrophoresis microchips,” Anal. Chem. 77-
11, pp. 3536-3541.
[10] Horn, G., Mackin, T., Lesniak, J., Boyce, B., 2004 “A new approach for detecting defects in
bonded MEMS devices,” Experimental Techniques 28-5, pp. 19-22.
[11] Sasov, A., 2001 “Micro-CT for nondestructive 3D reconstruction of MEMS and sensors,”
Proceedings of the SPIE - The International Society for Optical Engineering 4407, pp. 388-394.
[12] Lin, C. L., Miller, J. D., 2001 “A new cone-beam X-ray microtomography facility for 3D
analysis of multiphase materials,” 2nd World Congress on Industrial Process Tomography
(Hanover, Germany, 29
th
– 31
st
August, 2001), pp. 98 – 109.