Background: A recent survey of in-hospital reprocessing in Tanzanian hospitals identified bag-valve masks (BVM) as a commonly reused single-use device. In low- and middle-income countries (LMIC), in-hospital reprocessing supports neonatal resuscitation strategies by helping to maintain adequate supplies of BVM. However, there is a need for device-specific protocols defining reprocessing procedures and inspection criteria to overcome variations in reprocessing practices between hospitals. The purposes of this study were: 1) to complete a comprehensive design review and identify challenges to reprocessing BVMs; and 2) to investigate three different residual bioburden analysis methods for assessing the efficacy of decontaminating a disposable BVM. Methods: New, unused bag-valve-masks were contaminated with Staphylococcus epidermidis and Artificial Mucus Soil to simulate the worst case soiling conditions. Devices underwent one of five disinfection protocols, including one currently used in a LMIC hospital. Three analytical (two quantitative and one qualitative) methods were selected to evaluate residual bioburden on the device following decontamination. Results: Of all protocols tested, only the positive control and the Soap and Bleach protocols met disinfection targets. Most cleaning outcomes were consistent from trial to trial for each protocol. However, cleaning outcomes varied greatly for the Alcohol Wipe protocol. For the residual bioburden analyses, the two quantitative methods produced similar results, but the qualitative measurement exhibited increased variability. Conclusion: While this study revealed positive disinfection outcomes for the Tanzanian hospital decontamination protocol, more studies are required to support these findings. Design features of the BVM mask presented challenges to cleaning and drying during different decontamination protocols, as seen in the variability in the Alcohol Wipe protocol performance. These findings support the case for a device-specific protocol for the BVM. Given proper hospital personnel training and available resources, in-hospital reprocessing could support neonatal resuscitation strategies and other demands for manual resuscitation by helping to maintain adequate supplies of BVM.
Regulatory agencies define a single-use device (SUD) as a medical device that is designated by the manufacturer for use during a single medical procedure on a single patient and is intended to be discarded after the procedure [1 - 4]. However, used SUDs are not discarded in all circumstances; rather, they are sometimes reprocessed for reuse using specific methods for cleaning and disinfection. Recent trends indicate regulated reprocessing is often performed by third-party reprocessors who are independent from healthcare facilities [4 - 8], but in-hospital reprocessing has been reported for many different types of SUDs and remains prevalent in many low- and middle-income countries (LMIC) [9 - 13].
The current investigation was motivated by a recent survey of in-hospital reprocessing in Tanzanian hospitals that identified bag-valve masks (BVM) as a commonly reused SUD [
Recognized challenges with in-hospital reprocessing include variations in reprocessing practices between hospitals and a need for device-specific protocols defining reprocessing procedures and inspection criteria [8 , 10 , 12 , 13 , 22 , 23]. In the Tanzanian survey [
The purposes of the current study were: 1) to complete a comprehensive design review and identify challenges to reprocessing BVMs; and 2) to investigate three different residual bioburden analysis methods for assessing the efficacy of decontaminating a disposable BVM.
BVM designs have basic common features, including a soft polymer mask to conform to the patient’s face, a deformable ventilation bag, a non-rebreathing valve connecting the mask to one end of the bag, and an air intake valve at the opposite end of the bag. Worldwide, self-inflating BVM are the most common manual ventilation device used in neonatal and adult intensive care units [15 , 18 , 26]. As described by Davies, et al. [
BVM can either be reusable or disposable. For the purposes of the current study, disposable BVM, hereafter referred to as Test BVM (
Several design features of the Test BVM were considered reprocessing challenges, including small crevices near the valve attachment, contours on the outside surface, and tight folds inside the mask (
For testing, a total of five Test BVM were purchased. Each mask was cut into two equal halves (
A BVM is considered an oronasal mask typically covering a patient’s mouth and nose [
Worst-case contamination conditions were achieved by fully submerging the Test BVM mask halves into a soil solution consisting of standard mucus test soil simulating mucus exposure from a cystic fibrosis patient [
The simulated mucus soil, termed Artificial Mucus Soil, was prepared according to an international standard for validation of cleaning methods for reusable medical devices [
Decontamination requires cleaning of the device to the point where visible bioburden is removed. According to FDA regulatory classifications, BVM are semi-critical reprocessed devices due to contact with mucous membranes (but not sterile tissue); therefore, the decontamination protocols for cleaning must remove visible bioburden and achieve high-level disinfection to eliminate microorganisms [
Five decontamination protocols, including three experimental protocols and positive and negative controls, were applied to the Test BVM masks following contamination. The negative control was designed to yield a high bioburden and included masks that did not undergo any decontamination (
The three experimental decontamination protocols were chosen based on hospital reprocessing observations at three hospitals in Tanzania [
Three analytical methods were selected to evaluate the residual bioburden on the Test BVM following each decontamination protocol (
Name | Method | Markers | Metric | Measure |
---|---|---|---|---|
Metric 1 | 1ChannelCheckTM | carbohydrates, protein, hemoglobin | color change on test strip pads | qualitative |
Metric 2 | 2ATP Complete® | adenosine triphosphate (ATP) | relative luminescence units (RLU) | quantitative |
Metric 3 | Standard Plate Count | bacterial growth | colony forming units (CFU) | quantitative |
1Healthmark Industries Company, Inc, Fraser, MI; 2Ruhof Corporation, Mineola, NY.
that were swabbed 24 hours after contamination. Sample collection for Metric 2 involved swabbing approximately one-fourth of the inside surface area due to the manufacturer recommendation to use a small sampling area.
Metric 1 was a commercial method that provided a quick (~2 minutes), qualitative assessment of cleanliness by detecting the presence of residual carbohydrates, protein, and hemoglobin on sample swabs. Test strips provided by the manufacturer featured three colored pads that indicated the presence of physical bioburden in the swabbed region. According to the manufacturer, the detection limits for the test strips are ≥210 ug/ml for carbohydrate, ≥120 ug/ml for protein, and ≥0.25 ug/ml for hemoglobin. For this study, decontamination targets were met when both trials showed no indicator color change on any test strip pad, which was consistent with reduction of residual carbohydrates, protein, and hemoglobin below detection limits during decontamination.
Metric 2 was a commercial method that provided a quick (~10 seconds), quantitative assessment of cleanliness by detecting the presence of residual ATP on sample swabs. Reagent vials provided by the manufacturer emit bioluminescence, which correlates to certain ATP levels and was detectable as light emission when inserted into a handheld device also provided by the manufacturer. According to the manufacturer, the detection limit is 0.2 mg protein per swab, and a surface can be considered “clean” if the RLU (relative luminescence units) value displayed is less than 100. For this study, decontamination targets were met when both trials had RLU less than 100.
Metric 3 was a commonly utilized microbiological technique (standard plate count) that provided a quantitative assessment of disinfection by detecting the presence of residual bacteria on sample swabs, which present as CFUs on agar plates. Plate counts of CFUs were repeated in triplicate for swabs from each Test BVM half mask, averaged, and then divided by the plate dilution to obtain the concentration of bacteria. Overall log reduction of CFUs for a given decontamination protocol was calculated as the difference in bacterial concentration following the decontamination protocol (CP) relative to the negative control sample (CNC) (Equation (1.1)).
Sample collection involved swabbing the designated inside surface area of each half mask. Following instructions for use for Metric 1, swabs were placed into sterile test tubes with 10 ml of sterile Millipore (ASTM Type I) water and vortexed for 1 minute, followed by full immersion of the provided test strips into the solution and manual agitation for 10 seconds. The test strips were removed and held horizontally for 90 seconds prior to reading results. The test strips were compared to the color chart provided by the manufacturer, and the presence or absence of residue was recorded. Following instructions for use for Metric 2, swabs were placed in provided reagent vials and gently shaken for 3 seconds prior to reading results by inserting individual vials into the hand-held unit by the manufacturer and recording the displayed RLU value. Following standard microbiological methods for Metric 3, swabs were placed into sterile test tubes with 10 ml of sterile Millipore (ASTM Type I) water and vortexed for 1 minute. A ten-fold dilution series was prepared, plated onto agar (Tryptic Soy Agar, Remel, Lenexa, KS), and incubated for 24 hours at 37˚C prior to reading results by manually counting CFUs for each plate.
Careful review of the Test BVM identified several design features that were considered reprocessing challenges (
As anticipated from the design review, the BVM mask design features negatively impacted the performance of the decontamination protocols. Based on visual examination of the half masks, tight folds and rounded cavities on the interior surface of the mask retained more water and bioburden than other areas of the mask. This was most prevalent after the mask was removed from the soil solution and allowed to dry, when noted mask regions contained pools of the soil solution. Such areas were more difficult to reach during manual cleaning with the alcohol wipes compared to protocols based on mask submersion in cleaning solutions. This was reflected in the inconsistent results for the Alcohol Wipe protocol (
Protocol | Metric 1 | Target met? | |||||
---|---|---|---|---|---|---|---|
Protein Present? | Carbohydrates Present? | Hemoglobin Present? | |||||
Trial 1 | Trial 2 | Trial 1 | Trial 2 | Trial 1 | Trial 2 | ||
Negative Control | Yes | Yes | Yes | Yes | No | No | No |
Water Rinse | No | Yes | Yes | Yes | No | No | No |
Alcohol Wipe | No | No | No | Yes | No | No | No |
Soap and Bleach | No | Yes | No | Yes | No | No | No |
Positive Control | No | No | No | No | No | No | Yes |
Protocol | Metric 2 | Target met? | |
---|---|---|---|
ATP Value (RLU) | |||
Trial 1 | Trial 2 | ||
Negative Control | 9999 | 3967 | No |
Water Rinse | 9999 | 7112 | No |
Alcohol Wipe | 2343 | 8948 | No |
Soap and Bleach | 0 | 6 | Yes |
Positive Control | 0 | 0 | Yes |
Protocol | Metric 3 | Target met? | |
---|---|---|---|
Log Reduction in Bacteria | |||
Trial 1 | Trial 2 | ||
Negative Control | 0 | 0 | No |
Water Rinse | 3.7198 | 2.5179 | No |
Alcohol Wipe | 8.2576 | 3.3010 | No |
Soap and Bleach | 8.2576 | 7.1613 | Yes |
Positive Control | 8.2576 | 7.1613 | Yes |
All three methods for assessing residual bioburden required training of personnel based on instructions for use (Metrics 1 and Metrics 2) or standard microbiological techniques (Metric 3). Metric 1 required little training outside of the swabbing technique. However, Metric 1 was a qualitative assessment judged by the user performing the test and results were highly subjective due to differences in individual ability to identify a change in color. Metric 2 required ample training time for: 1) proper swabbing technique; 2) preparing the swab and reagent vial for insertion into the handheld device; and 3) operating the device. Metric 3 required time-intensive training for: 1) cell culturing techniques and use of lab equipment; 2) growing stock inoculum and culturing of bacteria; 3) proper inoculation of the BVM masks; and 4) growing bacterial samples on agar plates.
Based on these preliminary data, the Alcohol Wipe and Water Rinse protocols (
All three methods for assessing residual bioburden required some use of consumable materials and/or durable equipment. Metric 1 required use of consumable materials (swabs, test strips) to provide a qualitative assessment of residual bioburden (proteins, carbohydrates, hemoglobin) based on color change. Metric 2 required use of consumable materials (swabs, reagent vials) and durable equipment (refrigerator for reagent vials, hand-held device for measuring the RLU of emitted bioluminescence) to provide a quantitative assessment of residual ATP. Metric 3 required use of consumable materials (swabs, bacteria and growth media, agar plates, cell spreaders, pipette tips) and durable equipment (incubator, pipettes, biological hood) for measuring bacterial CFUs to provide a quantitative assessment of residual bacterial concentration.
This study completed a comprehensive design review of a disposable BVM to identify potential challenges to reprocessing and investigated three different methods for assessing residual bioburden on the BVM masks following decontamination. Overall, only the positive control protocol (
The BVM mask geometry negatively impacted the performance of the decontamination protocols. Flaps of pliable materials, tight folds and rounded cavities on the interior surface of the mask were challenging to clean and contributed to large variations in cleanliness for Metric 2 and Metric 3 in the Alcohol Wipe protocol. Although only the BVM mask was evaluated in the decontamination protocols, the design review identified additional challenges associated with the modular connection of the non-rebreathing valve, as well as the complex valve geometry having small crevices, dead-end chambers, and a filter.
All three methods required some level of training based on instructions for use (Metric 1 and Metric 2) or standard microbiological techniques (Metric 3). Metric 1 seemingly required the least training but generated the least consistent results. The time required for assessment after sample collection ranged from approximately 1 - 3 minutes for Metric 1 and Metric 2 and approximately 1 day for Metric 3.
Several drawbacks were noted for each of the residual bioburden assessment methods. Metric 1 required subjective interpretation of color change and was incompatible with protocols involving bleach, which induced a color change falsely indicating the presence of hemoglobin. This inaccuracy would be problematic if hemoglobin was a routine contaminant for a given SUD. Possibilities for overcoming color interpretation include use of a colorimeter to compare the test strip to the color standard or taking pictures of the test strips in a uniform, well-lit environment and completing hue analysis. Drawbacks for Method 2 and Method 3 center on the need for consumables and durable equipment and extended training time, as mentioned in the results.
Several study limitations are noted. These results may not be generalizable, as only one model of disposable BVM was evaluated in this study. This underscores the need for device-specific validation of reprocessing protocols, as design features vary between BVM and can impact the effectiveness of decontamination protocols. While the design review involved the entire Test BVM (full mask and non-rebreathing valve), the analytical methods for residual bioburden analysis (Metrics 1 - 3) were evaluated using masks halves (n = 2). This approach provided for efficient screening of the analytical methods, but it is recognized that additional evaluation of the full BVM (mask and valve) and statistical comparisons are needed for definitive conclusions related to the decontamination protocols. Finally, all analyses were performed by a team of five trained bioengineering graduate students with faculty supervision, which does not represent the personnel likely to conduct decontamination of disposable BVM in a healthcare setting.
This study describes an initial experimental approach for validating BVM decontamination protocols and generating data for objective assessment of reprocessing and reuse practices. The data support positive decontamination outcomes using the bleach-based in-hospital reprocessing protocol currently in use in some Tanzanian hospitals [
The authors declare no conflicts of interest regarding the publication of this paper.