In the radiolabelling process, the frequency of routine analytical testing must be defined with a risk assessment based on different factors (Pharmeuropa Vol. 26 No. 2, 5.19). Aim of this study is to define the criteria based on the risk assessment required for setting the frequency of analytical controls. 331 preparations of three different radiopharmaceuticals labeled with 99m Tc were examined analyzing temporal trends o f radiochemical purity values, potential correlation between radiochemical purity values and total radioactivity content of bulk as well as potential correlation between radiochemical purity values and radioactivity concentration of bulk. The analysis shows no direct correlation between percentage of radiochemical purity and total radioactivity content of preparation and between radiochemical purity values and radioactivity concentration in final bulk solution. Moreover, the routine analytical controls executed for one year on each preparation made it possible to determine how the analytical values do not have inter-operator dependency and there is no evidence of specific temporal trends over time. The study is aimed to assess any eventual criticality related with radiochemical purity tests, determine the possibility of a parametric release and the optimal frequency of tests, without reducing the level of safety of the preparations. This will produce a substantial reduction of costs and of radiation exposure to ionizing radiation of the staff dedicated to Quality Control procedures.
In Europe, the preparation and quality control of radiopharmaceuticals from kits is governed by Pharmeuropa Vol. 26 No. 2-5.19 Extemporaneous preparation of radiopharmaceutical preparations [
・ The demonstration that the test is redundant, i.e. it has not detected any out of alert limit situations, failures or other anomalies not already detected by the remaining system.
・ The product quality being assessed is assured, or directly tested by the remaining system.
The goal of this study is to assess any eventual criticality related with radiochemical purity tests, determine the possibility of a parametric release and the optimal frequency of tests on radiopharmaceuticals from kits, while ensuring the same safety standards established by the regulations for such matters and, at the same time, reducing operator exposure to ionizing radiation.
This study examined the work done at the CDI’s nuclear medicine radiopharmacy (hot lab) over a 12-month period (September 2013-August 2014), focusing on the preparation of radiopharmaceuticals from kits. During that time, 344 preparations were made, as per the
All of the preparations were subjected to the analytical controls indicated by the manufacturers in the product characteristics: in particular, the radiochemical purity has been tested with a Thin-layer chromatography technique (silica gel and storage phosphor system). The radiochemical purity tests were performed immediately after radiolabelling processes. All workers perform media fill test twice a year and zero nonconformity were observed.
Radiopharmaceutical | Number of preparations |
---|---|
DTPA?99mTc Pentetate Injection (Mallinckrodt SpA, IT) | 63 |
HDP?99mTc Oxidronate (Mallinckrodt SpA, IT) | 117 |
Myoview?99mTc Tetrofosmin for Injection (GE Healthcare SrL, IT) | 151 |
Maasol?99mTc Human Albumin Macroaggregates Injection (GE Healthcare SrL, IT) | 7 |
Ceretec?99mTc Exametazime Injection (GE Healthcare SrL, IT) | 2 |
Nanocoll?99mTc Albumin Colloid (GE Healthcare SrL, IT) | 4 |
During the preparation processes for the radiopharmaceuticals from kits in nuclear medicine, radiochemical purity can essentially be influenced by three factors:
・ Manual skill of workers;
・ Radioactivity concentration of the bulk;
・ Total radioactivity content in the bulk.
For the purpose of this study the data have been analyzed on a semi-quantita- tive level. In particular, to demonstrate that the radiochemical purity of radiopharmaceuticals from kits prepared at the nuclear medicine facility is not influenced by the aforementioned three factors this document analyzes the following:
・ Trends over time for radiochemical purity values;
・ Potential correlation between the values for radiochemical purity and the total radioactivity of the preparation [
・ Potential correlation between the same radiochemical purity values and radioactivity concentration [
This study refers to the DTPA, HDP and Myoview preparations. For the other three types of preparations, given the limited statistics, it is not possible to put forward similar assessments.
No values outside of the specifications were identified and they were routinely administered to patients.
Figures 1-3 show the trends as a function of time for radiochemical purity values, for the period under examination. For the simple moving average, the period for the trendline is equal to 5. The charts included are for DTPA, HDP and Myoview. For the remaining three kits, the limited number of preparations completed makes any analysis of trends over time impossible. For Maasol, Cere-
DTPA | HDP | Myoview | Maasol | Ceretec | Nanocoll | |
---|---|---|---|---|---|---|
Min | 95.0% | 95.0% | 90.5% | 95.5% | 96.0%; 97.4% | 95.6%; 98.2%; 99.9%; 98.6% |
Max | 99.2% | 99.2% | 98.8% | 99.0% | ||
Median | 96.5% | 96.2% | 96.1% | 98.6% | ||
Legal limit | 95.0% | 95.0% | 90.0% | 95.0% | 95.0% | 95.0% |
tec and Nanocoll, it can be merely noted that no values outside of the quality specifications were found during the analytical controls.
As can be seen from the charts, it is not possible to identify any specific trends over time during the observation period.
The following Figures 4-6 show, for DTPA, HDP and Myoview radiopharmaceuticals, the distribution of radiochemical purity in relation to the total radioactivity content of the bulk.
The charts also contain the linear interpolation parameters and the values of the correlation index (R) squared. (The radioactivity data has been adjusted to the values used in standard clinical practice.).
As shown, there is no correlation between the variation in the radiochemical purity values and the radioactivity content of the bulk. Therefore, it is possible to
rule out that the percentage of radiochemical purity might decrease with increased radioactivity levels, at least for the radioactivity intervals contained in the preparations created at CDI S.p.A.’s nuclear medicine hot lab.
The following charts show, for DTPA, HDP and Myoview radiopharmaceuticals, the distribution of radiochemical purity in relation the radioactivity concentration (MBq/ml) of the bulk preparation.
The charts also contain the linear interpolation parameters and the values of the correlation index (R) squared (
As clearly shown by the charts, the values of radiochemical purity do not correlate with the radioactivity concentration levels of the preparation. Therefore, it is possible to rule out that the percentage of radiochemical purity might decrease with increased radioactivity concentration, at least for the radioactivity intervals contained in the preparations created at CDI S.p.A.’s nuclear medicine hot lab (Figures 7-9). (The radioactivity concentration data has been adjusted to the values used in standard clinical practice.).
The data shows that it is not possible to identify specific trends over time apparently related to manual skill workers.
Furthermore, this parameter is independent of the radioactivity concentration of the final preparation and the total radioactivity levels. The frequency of quality
controls proposed by the Italian regulatory framework and by Pharmeuropa Vol. 26 No. 2-5.19 Extemporaneous preparation of radiopharmaceutical preparations seems excessive in centers that comply with the highest quality standards. More specifically, such centers are those where the preparations are completed in grade A isolators (Annex 1―EU Guide to Good Manufacturing Practice [
The data reported herein has been obtained after a year of controls performed regularly on all the preparations.
Given such initial statistical information and the working conditions, it may be considered to apply the concept of parametric release, in compliance with the PIC/S document entitled “Recommendation on Guidance on Parametric Release―Appendix 2” [
This would result in considerable savings in the solvents and consumables needed to perform quality controls (estimated at around 78%) [
In addition, reduced control frequency means a reduction in the extremity doses operators are exposed to during quality controls [
Ria, F., Albini, G., Battista, S., Salvatore, V., Messere, R., Bergantin, A., Redaelli, I., Bonfanti, P., Martinotti, A.S., Gandolfo, P. and Papa, S. (2017) Study to Define Frequency in Rou- tine Analytical Controls in the Radiola- belling Process. Open Access Library Jour- nal, 4: e3602. https://doi.org/10.4236/oalib.1103602