Article citationsMore>>
Farukhi, A.M., Stagg, B.C., Ronquillo Jr., C., Barlow Jr., W.R., Pettey, J.H., Olson, R.J., Gupta, I. and Jensen, J.D. (2014) Effect of Phaco Tip Diameter on Efficiency and Chatter. Journal of Cataract & Refractive Surgery, 40, 811-817.
http://dx.doi.org/10.1016/j.jcrs.2013.09.021
has been cited by the following article:
-
TITLE:
Retrospective Analysis of the Performance of a New-Generation Dual Pump Phaco System Using Two Sizes of Phaco Tips in Cases with Different Nuclear Hardness
AUTHORS:
Norbert Körber
KEYWORDS:
Effective Phaco Time, Total Phaco Time, Tip Size, Nuclear Density, EPT
JOURNAL NAME:
Open Journal of Ophthalmology,
Vol.6 No.3,
July
11,
2016
ABSTRACT: Purpose: To evaluate the performance of a phacoemulsification system in terms of effective (EPT) and total phaco time (TPT) using 20G and 21G phaco tips. Methods: Retrospective comparative study including 143 consecutive cataractous eyes undergoing phacoemulsification cataract surgery with the Visalis 500 device. The 20G and 21G phaco tips were used in 46 and 97 eyes, respectively. The EPT and TPT values were evaluated. Results: Median TPT was 11.25 s and 17.50 s in the 20G and 21G groups, respectively (p = 0.0011). Median EPT values were 3.15 s and 5.00 s in the 20G and 21G groups, respectively (p = 0.0032). TPT and EPT were significantly lower in 3/3+ cataract eyes compared to 4/4+ using both tips (p
Related Articles:
-
Suryakanth Dattatraya Anturlikar, Mohamed Rafiq, Gollapalle Laksminarayanashastry Viswanatha, Mohammed Azeemuddin, Agadi Hiremath Thippeswamy, Uday Kumar Veeranna Kanni, Pralhad Sadashiv Patki
-
Christian Schäferhoff, Jörg Förster, Bernd Schneider
-
Yida Li
-
Angue Tatiana Harly Mba Aki, Emmanuel Anyunzoghe, Emmanuel Anyunzoghe, Patrice Emery Itoudi Bignoumba, Stéphanie Ndjillekissa Ndjié, Christian Agaya, Emmanuel Mve Mengome
-
Bin Lin, Longlong Chen, Dongkan Li