ABSTRACT

In inductively coupled plasma mass spectrometry (ICP-MS) analysis, only a few options are available to deal with non-spectroscopic interferences. Considering that diluting the sample is impractical for traces analysis, other alternatives must be employed. Traditionally, the method of standard additions is used to correct the matrix effect but it is a time consuming method. Others methods involves separation techniques. Another way to overcome matrix interferences is to understand the mechanism involved and adjust plasma viewing conditions to reduce or eliminate the effect. In this study, the effect of various concomitant elements in ICP-MS was assessed by measuring the distribution of selected singly charged analyte ions (Al, V, Cr, Mn, Ni, Co, Cu, Zn, As, In, Ba, La, Ce, Pb), doubly charged ions (La, Ce, Ba and Pb) and oxides ions (BaO) in the presence of concomitant elements spanning a mass range from 23 (Na) to 133 (Cs) u.m.a. and different ionization energies. Concomitant elements are alkali metals, alkaline earth metals and Si. Analyte ion suppression was observed while moving the ICP across and away from the sampling interface with or without a single concomitant element. Matrix effect measures were realised, firstly, to highlight the relation between the signal extinction of an analyte and the masse of the concomitant element, and secondly to highlight the relation between the removal of the analyte signal and the first ionization energy of the element of matrix. A dependence upon both the mass of the matrix element and the mass of the analyte was observed. The suppression seems increased with increasing matrix element mass and decreased with increasing analyte mass. The effect of the mass of the matrix element was the more significant of the two factors. If space-charge effects were found to be significant for matrix elements of much lower mass, it seems diffusion also played an active part for heavier matrix elements. Finally, some evidence was found for a shift in ion-atom equilibrium for dications and for energy demand regarding oxides.

Share and Cite:

C. Mariet, F. Carrot and M. Moskura, "Effect of Sample Matrix on Radial and Axial Profiles of Ion Abundance in Inductively Coupled Plasma Mass Spectrometry," American Journal of Analytical Chemistry, Vol. 2 No. 7, 2011, pp. 739-751. doi: 10.4236/ajac.2011.27085.

Cited by

[1] Variation in the determination of platinum group metals using ICP OES induced by the effect of complex matrices and the correction method based on multivariate …
Journal of Analytical Atomic …, 2022
[2] Sustainable Valorization of Steelmaking Slag: From Metal Extraction to Carbon Sequestration
2021
[3] Space charge effects and internal standardization in a four ion lenses interface: What is changing in quadrupole inductively coupled plasma mass spectrometry?
2020
[4] Sample preparation and calibration optimization for ICP-MS analysis of copper, zinc, selenium, rubidium, strontium, magnesium, iron, molybdenum and barium?in human serum
Bulgarian Chemical Communications, 2019
[5] Matrix complexity effect on platinum group metals analysis using inductively coupled plasma optical emission spectrometry
Journal of Analytical Atomic Spectrometry, 2018
[6] Fast abiotic sorption of selenates (SeO42-) in soils: pitfalls of batch sorption data acquired by inductively coupled plasma quadrupole mass spectrometry (ICP-QMS)
Archives of Agronomy and Soil Science, 2018
[7] Análisis multilemental de minerales lateríticos pro ICP-OES. Efectos matriz, su reducción y posibles mecanismos
2018
[8] Matrix Effects using an ICP-MS with a Single Positive Ion Lens and Grounded Stop: Analyte Mass Dependent?
Journal of Analytical Atomic Spectrometry, 2017
[9] An ICP index for ICP-MS determinations–new selection rules for internal standards in ICP-MS determinations and carbon enhancement effect
Journal of Analytical Atomic Spectrometry, 2017
[10] Determination of ultra-trace elements in human plasma or serum by ICP-MS using sodium in the presence of carbon as a single calibration matrix-match component
Journal of Analytical Atomic Spectrometry, 2015
[11] FOR INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY
Diagnostic studies of ion beam formation in inductively coupled plasma mass spectrometry with the collision reaction interface, 2015
[12] Rapid Determination of Trace and Ultra Trace Level Elements in Diverse Silicate Rocks in Pressed Powder Pellet Targets by LA‐ICP‐MS using a Matrix‐Independent …
Geostandards and Geoanalytical Research, 2014
[13] Rapid Determination of Trace and Ultra Trace Level Elements in Diverse Silicate Rocks in Pressed Powder Pellet Targets by LA‐ICP‐MS using a Matrix‐Independent Protocol
Geostandards and Geoanalytical Research, Wiley Online Library, 2014

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.