Optics and Photonics Journal

Volume 7, Issue 8 (August 2017)

ISSN Print: 2160-8881   ISSN Online: 2160-889X

Google-based Impact Factor: 0.76  Citations  h5-index & Ranking

Test Method of Laser Detection Sensitivity Based on Every Pulse Measurement and Rearrangement

HTML  XML Download Download as PDF (Size: 278KB)  PP. 33-39  
DOI: 10.4236/opj.2017.78B006    765 Downloads   1,726 Views  

ABSTRACT

It is very important to accurately measure the detection sensitivity of laser receiving equipment. Based on the traditional test method of detection probability curve, a new test method is proposed which works through measuring laser pulse one by one. Accurate measurement systems were constructed to improve the accuracy of laser energy measurement and energy regulation. A new data processing method of detection probability curve is put forward, which based on subsection statistics. The new data processing method in effect reduces the light source instability from 6.57% to 0.67%. These works improve the test accuracy of laser receiving detection sensitivity. It is a great support for the accurate evaluation of key technical indices of laser receiving equipment, which subsequently are done by models and simulation.

Share and Cite:

Guo, H. , Ma, N. , Zhao, H. and Yao, M. (2017) Test Method of Laser Detection Sensitivity Based on Every Pulse Measurement and Rearrangement. Optics and Photonics Journal, 7, 33-39. doi: 10.4236/opj.2017.78B006.

Cited by

No relevant information.

Copyright © 2024 by authors and Scientific Research Publishing Inc.