Open Journal of Radiology

Volume 7, Issue 1 (March 2017)

ISSN Print: 2164-3024   ISSN Online: 2164-3032

Google-based Impact Factor: 0.33  Citations  

Assessment of Image Quality Parameters for Computed Tomography in Sudan

HTML  XML Download Download as PDF (Size: 1113KB)  PP. 75-84  
DOI: 10.4236/ojrad.2017.71009    2,337 Downloads   5,634 Views  Citations

ABSTRACT

X-ray-computed tomography (CT) has become one of the most important investigation procedures worldwide. The study aimed to assess image quality parameters, mainly noise, and radiation doses during abdominal examination. This study examined the diagnostic parameters (kilo voltage, tube current time product, slice thickness, and pitch) and their effects on image quality as well as the radiation doses received from computed tomography scanners using phantom. The study carried out in four CT centers in Sudan. The study applied prospective and experimental methods. The study demonstrated there was a linear correlation between diagnostic parameters and image noise. The reduction in milli-ampere second and peak kilo voltage increased the image noise. Moreover increasing the pitch led to an increase in the image noise, whereas increasing the slice thickness, reduced the image noise. There was also a linear relationship between kilo voltage and radiation dose at Elnileen diagnostic center characterized by an increase kilo voltages values which led to an increase in the radiation dose by 92% and a reduction in the image noise by 83%. However, at Antalya medical center, increasing in kilo voltage values led to an increase in the radiation dose by 35% and a reduction in the image noise by 26%. Also increasing in milli-ampere second values led to an increase in the radiation dose by 49% and a reduction in the image noise by 46% in a phantom compared with an increase in radiation dose by 82% and a reduction in the image noise by 51% in patients .The study found that an optimal protocol for adult abdominal scan at Antalya medical center was 4.22HU for image noise and 10.45 mGy for radiation dose when using 120 kVp, 300 mAs, 5 mm slice thickness and pitch of 0.8. At Elnileen diagnostic center, however, the optimal protocol was 5.4 HU for image noise and 5.4 mGy for radiation dose using 130 kVp, 50 mAs, 10 mm slice thickness and pitch of 2. In addition, the quality control tests for image quality parameters carried out at the two centers were performed by using the Chat Phan phantom and all the tests were within the acceptable limits, according to Sudan Atomic Energy Commission (SAEC) Standardizations. The study concludes with a number of recommendations, such as; the necessity for an extensive collaboration among manufacturers, radiologists, technologists and physicists to find a plan to decrease patient radiation dose (ALARA Principle) from computed tomography scanner.

Share and Cite:

Elnour, H. , Ahmed Hassan, H. , Mustafa, A. , Osman, H. , Alamri, S. and Yasen, A. (2017) Assessment of Image Quality Parameters for Computed Tomography in Sudan. Open Journal of Radiology, 7, 75-84. doi: 10.4236/ojrad.2017.71009.

Cited by

[1] Automation of slice thickness measurements in computed tomography images of AAPM CT performance phantom using a non-rotational method
Polish Journal of …, 2022
[2] Computed Tomography Protocol Optimisation for Pediatric Head Trauma: Radiation Dose and Image Quality Assessment
International Journal of …, 2022
[3] Assessment of Image Quality Parameters for Computed Tomography in Tripoli, Libya
2022 IEEE 2nd International …, 2022
[4] Evaluasi Nilai CTDIVOL Dan DLP Pada Pemeriksaan MSCT Abdomen Non Kontras Selama Periode Maret 2022 Sampai Juni 2022 Di Rumah Sakit TK II Pelamonia …
Garuda Pelamonia …, 2022
[5] PENGARUH VARIASI REKONSTRUKSI SLICE THICKNESS DAN FILTER KERNEL TERHADAP KUALITAS CITRA CT-SCAN KEPALA PADA KASUS STROKE …
Humantech: Jurnal Ilmiah …, 2022
[6] Teachers and epilepsy: What they know, do not know, and need to know: A cross-sectional study of Taif City
2019
[7] What Saudi teachers know about epilepsy: A cross-sectional study of Tabuk City
Epilepsy & Behavior, 2018
[8] Assessment of Image Quality Parameters for Computed Tomography in the City of Yaoundé
2018
[9] Evaluación de la Evolución Temporal en Tumores Pulmonares Tratados con Radioterapia Estereotáctica Corporal a partir de Rasgos Extraídos de las Imágenes de …

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.