American Journal of Analytical Chemistry

Volume 2, Issue 3 (July 2011)

ISSN Print: 2156-8251   ISSN Online: 2156-8278

Google-based Impact Factor: 1.71  Citations  h5-index & Ranking

A Validated Liquid Chromatography-Mass Spectrometry Method for the Detection and Quantification of Oxidative Metabolites of 2,2',4,4'-Tetrabromodiphenyl Ether in Rat Hepatic Microsomes

HTML  Download Download as PDF (Size: 561KB)  PP. 352-362  
DOI: 10.4236/ajac.2011.23043    4,372 Downloads   8,154 Views  Citations

Affiliation(s)

.

ABSTRACT

In the present study, we developed and validated an analytical method using ultra performance liquid chromatography-mass spectrometry (UPLC/MS) for the quantitative determination of 2,2',4,4'-tetrabromodiphenyl ether (BDE-47) metabolism by rat hepatic microsomes. BDE-47 is a brominated flame retardant that was widely used in a variety of consumer products and has subsequently been identified as a ubiquitous environmental contaminant. Hydroxy-bromodiphenyl ethers (OH-BDEs) were isolated from rat hepatic microsomes by liquid-liquid extraction. Chromatographic separation was achieved by UPLC on a C18 column with gradient elution using a mobile phase consisting of methanol and water, each containing 0.1% formic acid, at a flow rate of 0.2 mL/min. Detection and quantification were performed using a mass spectrometer in single ion recording mode with negative electrospray ionization. The UPLC/MS method was validated for linearity, limit of quantification (LOQ), accuracy, precision and recovery. The weighted calibration curves (1/X2) were linear over a concentration range of 5 - 250 nM with LOQ values between 5 nM and 50 nM for the individual OH-BDEs. Intra- and inter- day accuracy (%DEV) and precision (%RSD) values ranged from –11.7% to 9.5% and 5.9% to 16.5%, respectively. Recovery values of 70% to 90% were obtained for all OH-BDEs. The validated method allowed us to successfully analyze metabolite formation following incubation of BDE-47 with hepatic microsomes prepared from phenobarbital-treated rats. Results demonstrate that the UPLC/MS method has sufficient sensitivity and reproducibility to fully characterize the in vitro metabolism of BDE-47 and possibly other PBDEs.

Share and Cite:

Moffatt, S. , Edwards, P. , Szeitz, A. and Bandiera, S. (2011) A Validated Liquid Chromatography-Mass Spectrometry Method for the Detection and Quantification of Oxidative Metabolites of 2,2',4,4'-Tetrabromodiphenyl Ether in Rat Hepatic Microsomes. American Journal of Analytical Chemistry, 2, 352-362. doi: 10.4236/ajac.2011.23043.

Cited by

[1] Evaluation of hepatic biotransformation of polybrominated diphenyl ethers in the polar bear (Ursus maritimus)
Chemosphere, 2016
[2] Regioselective Versatility of Monooxygenase Reactions Catalyzed by CYP2B6 and CYP3A4: Examples with Single Substrates
Monooxygenase, Peroxidase and Peroxygenase Properties and Mechanisms of Cytochrome P450, 2015
[3] Legacy and emerging organic pollutants in liver and plasma of long-finned pilot whales (Globicephala melas) from waters surrounding the Faroe Islands
Science of The Total Environment, 2015
[4] Fast liquid chromatographic–tandem mass spectrometric method using mixed-mode phase chromatography and solid phase extraction for the determination of 12 …
Journal of Chromatography A, 2014
[5] Fast liquid chromatographic–tandem mass spectrometric method using mixed-mode phase chromatography and solid phase extraction for the determination of 12 mono-hydroxylated brominated diphenyl ethers in human serum
Journal of Chromatography A, Elsevier, 2014
[6] 多溴联苯醚及其代谢产物分析方法研究进展
公共卫生与预防医学, 2014
[7] Biotransformation of 2, 2′, 4, 4′-tetrabromodiphenyl ether (BDE-47) by human liver microsomes: identification of cytochrome P450 2B6 as the major enzyme involved
Chemical research in toxicology, 2013
[8] Development of a liquid chromatography-electrospray chemical ionization tandem mass spectrometry analytical method for analysis of eleven hydroxylated polybrominated diphenyl ethers.
Journal of chromatography, 2013
[9] Development of a liquid chromatography–electrospray chemical ionization tandem mass spectrometry analytical method for analysis of eleven hydroxylated …
Journal of Chromatography A, 2013
[10] Biotransformation of 2, 2′, 4, 4′-tetrabromodiphenyl ether (BDE-47) by human liver microsomes: identification of cytochrome P450 2B6 as the major enzyme …
Chemical Research in Toxicology, 2013
[11] Development of a liquid chromatography–electrospray chemical ionization tandem mass spectrometry analytical method for analysis of eleven hydroxylated …
Journal of Chromatography A, 2013
[12] Comparative oxidative metabolism of BDE-47 and BDE-99 by rat hepatic microsomes
Toxicological Sciences, 2011

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.