A Comparison between Three-Dimensional Conformal Radiotherapy, Intensity-Modulated Radiotherapy, and Volumetric-Modulated Arc Therapy Techniques for Stereotactic Body Radiotherapy of Lung Tumors

HTML  XML Download Download as PDF (Size: 1079KB)  PP. 104-112  
DOI: 10.4236/ijmpcero.2015.42014    4,850 Downloads   6,950 Views  Citations

ABSTRACT

Introduction: The aim of this study is to dosimetrically compare 3D CRT, IMRT, and VMAT techniques that employ coplanar and non-coplanar beams for the SBRT of lung tumors. Methods and Materials: Nine (n = 9) consecutive SBRT lung patients with ten tumor sites who were previously treated at our institution were selected for this study. Six (n = 6) treatment plans were created for each PTV: 1 coplanar and 1 non-coplanar 3D CRT, IMRT, and VMAT such that 98% of PTV received 100% of prescription dose, Rx of 50 Gy in 5 fractions. The data collected from each plan included the conformity index, R50%, and homogeneity index of the target as well as the volume of normal tissue irradiated by 5 Gy, 25 Gy isodose lines, V5 and V20 of total lung, and maximum dose to organs at risk. Beam angles and arc lengths were chosen in order to achieve the lowest total lung V5, V20 and R50% values. Results: According to the observed data, the mean total lung V5 and V20 values were lowest for the non-coplanar VMAT plans, but were not statistically different from the other planning techniques. Conformity values were similar for the IMRT and VMAT plans, and significantly lower than the 3D CRT plans. R50% values were lowest for the VMAT plans and significantly lower than both IMRT and 3D CRT plans in both coplanar and non-coplanar beam arrangements. However, dose homogeneity in the PTV is significantly higher in the IMRT coplanar plans than the corresponding 3D CRT and SmartArc (SA) coplanar plans. Coplanar VMAT plans were able to produce a significant 35.5% reduction in the maximum cord dose than coplanar 3D CRT plans (p-value < 0.03). VMAT plans also reduce the volume of the normal tissue irradiated to high doses, but the reduction is insignificant. Conclusion: Overall, the VMAT plans were able to produce highly conformal plans. No single planning technique can outclass the others in all of the SBRT planning indices significantly and a thorough review of each patient plan is needed.

Share and Cite:

Narayanasamy, G. , Granatowicz, D. , Baacke, D. , Li, Y. , Gutierrez, A. , Papanikolaou, N. and Stathakis, S. (2015) A Comparison between Three-Dimensional Conformal Radiotherapy, Intensity-Modulated Radiotherapy, and Volumetric-Modulated Arc Therapy Techniques for Stereotactic Body Radiotherapy of Lung Tumors. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 4, 104-112. doi: 10.4236/ijmpcero.2015.42014.

Cited by

[1] ‌Dosimetric comparison of forward and inverse intensity modulated radiation therapy planning, and volumetric modulated arc therapy for (6and10) MV x-ray photons for …
2021
[2] A dose falloff gradient study in RapidArc planning of lung stereotactic body radiation therapy
Journal of medical physics, 2018
[3] Interplay effect in lung cancer proton therapy
2018
[4] Dosimetric evaluation of IMRT plan for homogenous and inhomogeneous medium using AAPM TG-119 protocol
Journal of Physics: Conference Series, 2017
[5] Dosimetric impact of organ motion with 4D-CT based treatment planning in lung stereotactic ablative radiotherapy
2017
[6] Volumetric-modulated arc therapy lung stereotactic body radiation therapy dosimetric quality assurance: A comparison between radiochromic film and …
Journal of medical physics, 2017
[7] Comparison of radiotherapy dosimetry for 3D-CRT, IMRT, and SBRT based on electron density calibration
2016

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.