Journal of Biomedical Science and Engineering

Volume 8, Issue 4 (April 2015)

ISSN Print: 1937-6871   ISSN Online: 1937-688X

Google-based Impact Factor: 0.66  Citations  h5-index & Ranking

Local Pulse Wave Velocity Estimation in the Carotids Using Dynamic MR Sequences

HTML  XML Download Download as PDF (Size: 529KB)  PP. 227-236  
DOI: 10.4236/jbise.2015.84022    4,009 Downloads   5,105 Views  Citations

ABSTRACT

The current study presents a new protocol for local pulse wave velocity (PWV) measurement using dynamic MR sequences, which have a high temporal resolution (TR < 6 ms). MR images were obtained at two positions along the common carotid artery, separated by a distance of 5 cm. In each phase of a MR series, carotid region was automatically extracted and then its area distension waveform could be obtained. Sixteen volunteers with no symptoms of cardiovascular diseases were studied. For local PWV estimation, three delay estimation principles were tested and produced the following values: intersecting tangents method (M1): 4.72 ± 1.40 m/s, second derivative method (M2): 4.94 ± 1.68 m/s and cross-correlation method (M3): 5.03 ± 1.17 m/s. The cross-correlation method showed a relative high reliability as its least standard deviation.

Share and Cite:

Darwich, M. , Langevin, F. and Darwich, K. (2015) Local Pulse Wave Velocity Estimation in the Carotids Using Dynamic MR Sequences. Journal of Biomedical Science and Engineering, 8, 227-236. doi: 10.4236/jbise.2015.84022.

Cited by

[1] Development of non-invasive, optical methods for central cardiovascular and blood chemistry monitoring.
2022
[2] Non-Invasive Methods for PWV Measurement in Blood Vessel Stiffness Assessment
2021
[3] Measurement of local pulse wave velocity for carotid artery by using an ultrasound-based method
2020
[4] Local pulse wave velocity: theory, methods, advancements, and clinical applications
2019
[5] Krutost žila u hipertoničara s hipertenzivnom krizom mjereno oscilometrijskom metodom
2019
[6] Accelerometric patch probe for cuffless blood pressure evaluation from carotid local pulse wave velocity: design, development, and in vivo experimental study
2019
[7] Carotid Local Pulse Wave Velocity Measurement using Dual-Element Accelerometric Patch Probe
2018
[8] Non-Invasive Assessment of Local Pulse Wave Velocity as Function of Arterial Pressure
2018
[9] Measurement of carotid blood pressure and local pulse wave velocity changes during cuff induced hyperemia
2017
[10] Experimental validation of dual PPG local pulse wave velocity probe
2017
[11] Calculating Pulse Transit Times and Pulse Wave Velocities using Novel Processing of Surface Peri-Arterial Accelometric Data
ProQuest Dissertations Publishing, 2017
[12] Novel Methods for Pulse Wave Velocity Measurement
Journal of medical and biological engineering, 2015

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.