Comparison of Absorbed Dose to Medium and Absorbed Dose to Water for Spine IMRT Plans Using a Commercial Monte Carlo Treatment Planning System

HTML  Download Download as PDF (Size: 1106KB)  PP. 60-66  
DOI: 10.4236/ijmpcero.2014.31010    4,770 Downloads   9,166 Views  Citations

ABSTRACT

Dose in radiation therapy has been reported as the water-equivalent dose using conventional dose calculation algorithms. The Monte Carlo (MC) algorithm employs characterization of human tissues by elemental composition and mass density. It enables more accurate dose calculation for radiation therapy treatment planning and typically reports absorbed dose to medium. Whether one should use dose to medium or tissue (Dm) in place of dose to water (Dw) for MC treatment planning remains the subject of debate. The aim of the current study is to evaluate the differences between dose-volume indices for Dm and Dw MC-calculated IMRT plans. Thirty-seven spine patients were selected for this study. The IMRT optimization and MC calculations were performed using the iPlan RT DoseTM ver 4.1.2 (Brainlab, Munich, Germany) treatment planning system (TPS) with an X-ray Voxel Monte Carlo (XVMC) dose calculation engine. Dw and Dm results for target and critical structures were evaluated using the dose-volume-based indices. Systematic differences between dose-volume indices computed with Dw and Dm were up to 5.2%, 4.2%, and 4.5% for D2, D50 and D98 indices of the clinical target volume (CTV), respectively and up to 1% for the critical structure dose indices. Our study demonstrates that employing Dm in place of Dw in MC-calculated IMRT treatment plans introduces a significant systematic difference in target DVHs. We recommend that for diffused target structures (such as spine tumors), dose to water is a better quantity for dose prescription in photon beam treatment planning using existing MC TPS. While for critical structures, it would be reasonable to report Dm always. However in future with the availability of finer spatial resolution, Dm will be the most suitable variable for both target and critical structures’ dose prescription and reporting in MC treatment planning.

Share and Cite:

M. Usmani, N. Masai, R. Oh, H. Shiomi, D. Tatsumi, H. Miura, T. Inoue and M. Koizumi, "Comparison of Absorbed Dose to Medium and Absorbed Dose to Water for Spine IMRT Plans Using a Commercial Monte Carlo Treatment Planning System," International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, Vol. 3 No. 1, 2014, pp. 60-66. doi: 10.4236/ijmpcero.2014.31010.

Cited by

[1] Report dose-to-medium in clinical trials where available; a consensus from the global Harmonisation group to maximize consistency
2021
[2] Calculating the medium correction factor for the PTW-30013 ionization chamber
2021
[3] Dose Calculation And Reporting With A Linear Boltzman Transport Equation Solver In Vertebral SABR
2021
[4] Evaluation of clinical implications in the use of dose to water versus dose to medium by using NTCP and TCP models for urinary bladder tumours
2021
[5] Dose to medium in head and neck radiotherapy: Clinical implications for target volume metrics
2019
[6] EPSM 2017, Engineering and Physical Sciences in Medicine
2018
[7] Comparison of calculated dose distributions reported as dose-to-water and dose-to-medium for intensity-modulated radiotherapy of nasopharyngeal cancer patients
Medical Dosimetry, 2018
[8] O09 CT ventilation imaging derived from breath hold CT exhibits good regional accuracy versus Galligas PET
Australasian Physical & Engineering Sciences in Medicine, 2018
[9] Comparison of Gamma Pass Rate between the Dose-to-Water and Dose-to-Medium Reporting Modes for Patient-Specific QA Using a Helical Diode Array …
2018
[10] Estimation personnalisée de la dose délivrée au patient par l'imagerie embarquée kV-CBCT et réflexions autour de la prise en charge clinique
2017
[11] Modélisation Monte-Carlo d'un accélérateur linéaire pour la prise en compte des densités pulmonaires dans le calcul de la dose absorbée en radiothérapie …
Thèse, 2016
[12] Effects of contrast materials in IMRT and VMAT of prostate using a commercial Monte Carlo algorithm
Australasian Physical & Engineering Sciences in Medicine, 2016
[13] Dose-to-Medium Vs. Dose-to-Water: Dosimetric Evaluation of Pelvis Vmat Cases Using Monaco Treatment Planning System
2016
[14] Modélisation Monte-Carlo d'un accélérateur linéaire pour la prise en compte des densités pulmonaires dans le calcul de la dose absorbée en Radiothérapie …
Thesis, Université Toulouse 3 Paul Sabatier, 2016
[15] Dose-to-medium vs. dose-to-water: Dosimetric evaluation of head and neck VMAT cases using Monaco treatment planning system
2016

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.