Three Dimensional Expansion of Margins for Single-fraction Treatments: Stereotactic Radiosurgery Brain Cases

HTML  XML Download Download as PDF (Size: 432KB)  PP. 15-22  
DOI: 10.4236/ijmpcero.2012.12003    4,662 Downloads   11,753 Views  Citations

ABSTRACT

Purpose: To derive a clinically-practical margin formula between clinical target volume (CTV) and planning target volume (PTV) for single-fraction stereotactic radiosurgery (SRS). Methods: In previous publications on the margin between the CTV and the PTV, a Gaussian function with zero mean was assumed for the systematic error and the machine systematic error was completely ignored. In this work we adopted a Dirac delta function for the machine systematic error for a given machine with nonzero mean systematic error. Mathematical formulas for calculating the CTV-PTV margin for single-fraction SRS treatments were proposed. Results: Margins for single fraction treatments were derived such that the CTVs received the prescribed dose in 95% of the SRS patients. The margin defined in this study was machine specific and accounted for nonzero mean systematic error. The differences between our formulas and a previously published formula were discussed. Conclusion: Clinical margin formulas were proposed for determining the margin between the CTV and the PTV in SRS treatments. Previous margin’s recipes, being derived specifically for conventional treatments, may be inappropriate for single-fraction SRS and could result in geometric miss of the target and even treatment failure for machines possessing of large systematic errors.

Share and Cite:

Q. Zhang, M. Chan, Y. Song and C. Burman, "Three Dimensional Expansion of Margins for Single-fraction Treatments: Stereotactic Radiosurgery Brain Cases," International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, Vol. 1 No. 2, 2012, pp. 15-22. doi: 10.4236/ijmpcero.2012.12003.

Cited by

[1] Ramifications of Setup Margin Use During Frameless Stereotactic Radiosurgery/Therapy With Gamma Knife Icon Cone-Beam Computed Tomography (CBCT) …
Cureus, 2022
[2] Geometric uncertainty analysis of MLC tracking for lung SABR
2020
[3] Optimal Margin Application for Image-Guided, Frameless Gamma Knife Radiosurgery: A Correlational Analysis of This New Treatment Technique and Patient Specific …
2019
[4] The Effect of Setup Uncertainty on Optimal Dosimetric Margin in Linac-based Stereotactic Radiosurgery with dynamic conformal arc technique
2018
[5] A Comparison of Clinical and Radiologic Outcomes Between Frame-based and Frameless Stereotactic Radiosurgery for Brain Metastases
Practical Radiation Oncology, 2016
[6] Evaluation of setup uncertainties for single-fraction SRS by comparing two different mask-creation methods
Journal of the Korean Physical Society, 2015
[7] A new definition of biological effective dose: The dose distribution effects
Physica Medica, 2015
[8] Rotation effects on the target-volume margin determination
Physica Medica, 2015
[9] Evaluation of Setup Uncertainties for Single-Fraction SRS by Comparing the Two Different Mask-Creation Methods
arXiv preprint arXiv:1503.03921, 2015
[10] Determining Rational Planning Target Volume Margins for Intracranial Stereotactic Radiotherapy
2014
[11] A new variable for SRS plan quality evaluation based on normal tissue sparing: the effect of prescription isodose levels
The British journal of radiology, 2014
[12] The Effects of the Shape and Size of the Clinical Target Volume on the Planning Target Volume Margin
PloS one, 2014
[13] Feasibility study of real-time planning for stereotactic radiosurgery
Medical physics, 2013
[14] Forces on a saltating grain in air
Eur. Phys. J. E, 2013
[15] Three independent one-dimensional margins for single-fraction frameless stereotactic radiosurgery brain cases using CBCT
Medical physics, 2013
[16] Three independent one‐dimensional margins for single‐fraction frameless stereotactic radiosurgery brain cases using CBCT
Medical physics, 2013
[17] Feasibility study of real‐time planning for stereotactic radiosurgery
Medical physics, 2013
[18] Planning margins to CTV for image-guided whole pelvis prostate cancer intensity-modulated radiotherapy
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2012
[19] Evaluation of motion management strategies based on required margins
Physics in medicine and biology, 2012

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.