The Verification of iPlan Commissioning by Radiochromic EBT2 Films

HTML  XML Download Download as PDF (Size: 1012KB)  PP. 1-7  
DOI: 10.4236/ijmpcero.2012.11001    5,075 Downloads   13,436 Views  Citations

ABSTRACT

Purpose: To evaluate the measured dose distributions using radiochromic EBT2 films for small fields in iPlan (BrainLab) commissioning. Methods: Radiochromic EBT2 films were irradiated with 6 MV photons on a Varian Trilogy linac using polystyrene phantoms. The measurements included dose profiles and depth doses for field sizes of 1 × 1, 2 × 2, 3 × 3, 4 × 4, and 10 × 10 cm2 etc. The dose profile measurements were taken at the depth of 5 cm. The calibration films were irradiated at dmax(1.4 cm) for doses up to 6 Gy. Films were scanned using an Epson 10,000 XL flatbed scanner with 72 dpi resolution. Pixel values were converted to doses using the established calibration-curve. The 2D dose distributions were generated from the film data analysis. In-house software was utilized to compare the measured doses from films with the treatment planning data. In addition, selected patients’ SRS fields were also measured with the EBT2 films for comparison with iPlan. An EDGETM detector was also used to check the centralaxis doses for the SRS patients’ measurements. Results and Discussion: The measured planar dose distributions achieved more than 98% and 95% passing rates with a set of 2%/2 mm dose and DTA criteria for all square fields and all patient treatment fields (<5 × 5 cm2), respectively. Agreement with measurement data with the EDGETM detector at the central axis (±1%) was found with the plan data. This is the first report for SRS small photon-field measurement using the latest radiochromic film, EBT2. The results shown in this work indicate that the use of EBT2 film provides accurate dosimetry measure-ments for small photon beams. The measurements show excellent agreement with the iPlan TP commissioning data. Conclusions: The patient-specific EBT2 film QA for iPlan SRS patients showed good results. The EBT2 films could potentially be a useful dosimeter in verification of commissioning as well as patient-specific QA for SRS cases.

Share and Cite:

M. Chan, Q. Zhang, J. Li, P. Parhar, K. Schupak and C. Burman, "The Verification of iPlan Commissioning by Radiochromic EBT2 Films," International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, Vol. 1 No. 1, 2012, pp. 1-7. doi: 10.4236/ijmpcero.2012.11001.

Cited by

[1] Report of AAPM task group 235 radiochromic film dosimetry: an update to TG-55
2020
[2] Report of AAPM task group 235 radiochromic film dosimetry: an update to TG‐55
Medical …, 2020
[3] A novel extrapolation method using OSL detectors for very small field output factor measurement for stereotactic radiosurgery
2019
[4] Verification of CyberKnife Dose Distributions with Gafchromic EBT2 Film.
Turkish Journal of …, 2018
[5] Verification of CyberKnife Dose Distributions with Gafchromic EBT2 Film
2018
[6] Film dosimetry for linear accelerator-based stereotactic radiosurgery and stereotactic body radiation therapy
2017
[7] Small-field dosimetry in megavoltage beams
2017
[8] Spectral analysis of the EBT3 radiochromic film irradiated with 6 MV X-ray radiation
Radiation Measurements, 2016
[9] A novel phantom design for brachytherapy quality assurance
2016
[10] Is It Possible to Publish a Calibration Function for Radiochromic Film?
International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 2014

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.