Advances in Bioscience and Biotechnology

Volume 3, Issue 1 (February 2012)

ISSN Print: 2156-8456   ISSN Online: 2156-8502

Google-based Impact Factor: 1.18  Citations  h5-index & Ranking

Portable two linearly polarized beams absorption biometric system

HTML  Download Download as PDF (Size: 542KB)  PP. 40-45  
DOI: 10.4236/abb.2012.31007    3,765 Downloads   6,445 Views  Citations
Author(s)

ABSTRACT

The purpose of this paper is to design a portable two linearly polarized beams absorption biometric system (PTLPBABS) that measures the amount of absorption of the visible collimated beam passing by the sample to know the absorbance of the sample. The measured object in the system is the gold colloid that the sample produces after a biochemical reaction. After measuring the absorbance of the gold colloid we can obtain the concentrations of the waiting-measured sample and further obtain the concentration of chemical solution contained in the sample. The two linearly polarized beams device can be used to enhance rejection of ambient light and causes PTLPBABS to have excellent data stability measured. The process of the double optical path method increases the opportunities that the visible collimated beam is absorbed, which raise the detection sensitivity of the system to a sensitivity of 0.1%.

Share and Cite:

Chen, D. (2012) Portable two linearly polarized beams absorption biometric system. Advances in Bioscience and Biotechnology, 3, 40-45. doi: 10.4236/abb.2012.31007.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.