Reliability Study of a New Electromechanical Device Designed to Measure the Relative Dorsal Mobility of the First Ray of the Foot

HTML  XML Download Download as PDF (Size: 1357KB)  PP. 383-392  
DOI: 10.4236/ojo.2022.1210039    101 Downloads   672 Views  

ABSTRACT

Introduction: A new electromechanical instrument has been developed to measure relative dorsal mobility of the first ray in an objective and reliable way by simulating ground reaction forces during gait. This device equally applies a standardized, electronically controlled, and precise force under the first metatarsal head M1 as well as under the heads of the lesser metatarsals M2 to M5. The relative dorsal mobility between these two bearings is then measured. The purpose of this study is to assess the intra- and inter-examiners reliabilities of the measurements obtained with this device. Methods: The protocol included two examiners and 36 feet (18 volunteers with healthy feet and no history of forefoot disorders). A total of nine measurements were performed on each foot separated into three sets of three trials for the assessment of inter-rater and intra-rater reliability. For this purpose, the interclass correlation coefficient (ICC), the error of measurement (SEM) and the Bland and Altman (B&A) graphical analysis were computed. Results: Excellent ICC values (≥0.91) were obtained with the novel device for inter-rater and intra-rater reliability when using the FRRM calculation. The B&A analysis presented a bias between examiners of -0.25 mm ranging from -1.66 to 1.18 mm. Conclusion: This study demonstrated the capability of the developed device to reliably measure the relative dorsal mobility of the first ray of the foot. This is a promising first step for further studies to better understand, qualify and quantify first ray hypermobility.

Share and Cite:

Reymond, N. , Praz, Q. , Schoinas, S. , Sockalingam, N. , Acker, A. , Dubois-Ferrière, V. , Passeraub, P. and Assal, M. (2022) Reliability Study of a New Electromechanical Device Designed to Measure the Relative Dorsal Mobility of the First Ray of the Foot. Open Journal of Orthopedics, 12, 383-392. doi: 10.4236/ojo.2022.1210039.

Cited by

No relevant information.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.