International Journal of Medical Physics, Clinical Engineering and Radiation Oncology

International Journal of Medical Physics, Clinical Engineering and Radiation Oncology

ISSN Print: 2168-5436
ISSN Online: 2168-5444
www.scirp.org/journal/ijmpcero
E-mail: ijmpcero@scirp.org
"Accuracy of the Small Field Dosimetry Using the Acuros XB Dose Calculation Algorithm within and beyond Heterogeneous Media for 6 MV Photon Beams"
written by Sotirios Stathakis, Carlos Esquivel, Luis Vazquez Quino, Pamela Myers, Oscar Calvo, Panayiotis Mavroidis, Alonso N. Gutiérrez, Niko Papanikolaou,
published by International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, Vol.1 No.3, 2012
has been cited by the following article(s):
  • Google Scholar
  • CrossRef
[1] The effects of intestinal air cavity on dose distribution of volume modulated arc therapy for cervical cancer
Annals of Palliative …, 2022
[2] Dosimetric comparison and validation of Eclipse Anisotropic Analytical Algorithm (AAA) and AcurosXB (AXB) algorithms in RapidArc-based radiosurgery plans of …
Medical …, 2022
[3] Dosimetry of small photon fields in the presence of bone heterogeneity using MAGIC polymer gel, Gafchromic film, and Monte Carlo simulation
Reports of Practical …, 2022
[4] Dose calculation accuracy for photon small fields in treatment planning systems with comparison by Monte Carlo simulations
Polish Journal of Medical …, 2021
[5] Impact of Dose Algorithm and Calculation Angle Interval on in Vivo Dose Verification of Stereotactic Body Radiation Therapy: A Phantom Study
2021
[6] Küçük alanların dozimetrik parametrelerinin gafkromik film ile ölçülmesi
2020
[7] Evaluation of lung heterogeneity effects on dosimetric parameters in small photon fields using MAGIC polymer gel, Gafchromic film, and Monte Carlo simulation
2020
[8] Effect of lung inhomogeneity on dose distribution during radiotherapy of patient with lung cancer
2020
[9] The impact of scanning data measurements on the Acuros dose calculation algorithm configuration
2020
[10] Evaluation of dose calculation algorithms using different density materials for in-field and out-of-field conditions
2019
[11] Entropic model for dose calculation in external beam radiotherapy and brachytherapy
2019
[12] Stereotaktik radyocerrahi ışınlamalarda inhomojenitenin doz dağılımına etkisinin araştırılması
2019
[13] Farklı anatomik bölgelere ait tedavi planları için seçilen hesaplama çözünürlük değerlerinin optimizasyon ve hesaplama üzerine etkilerinin değerlendirilmesi
2019
[14] Evaluating small field dosimetry with the Acuros XB (AXB) and analytical anisotropic algorithm (AAA) dose calculation algorithms in the eclipse treatment planning …
2019
[15] Comparison of five dose calculation algorithms in a heterogeneous media using design of experiment
2019
[16] Evaluation of dose calculation algorithms using different density materials for in-field and out-of-field conditions.
2019
[17] Méthodologie pour la détermination de la dose absorbée dans le cas des petits champs avec et sans hétérogénéités pour des faisceaux de photons de haute énergie
2018
[18] Evaluation of AAA and XVMC Algorithms for Dose Calculation in Lung Equivalent Heterogeneity in Photon Fields: A Comparison of Calculated Results with …
Journal of biomedical physics & engineering, 2018
[19] Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons
Thesis, 2018
[20] Refinement of MLC modeling improves commercial QA dosimetry system for SRS and SBRT patient‐specific QA
Medical physics, 2018
[21] Verification of Acuros XB dose algorithm using 3D printed low‐density phantoms for clinical photon beams
Journal of applied clinical medical physics, 2018
[22] Dosimetric verification of small fields in the lung using lung-equivalent polymer gel and Monte Carlo simulation
2018
[23] Underdosing of the maxillary sinus for small fields used in newer radiotherapy techniques: Comparison of thermoluminescent dosimeter and Monte Carlo data
2018
[24] Investigation of Acuros XB radiotherapy dose calculation algorithm
2018
[25] The Empirical Mode Decomposition-Decision Tree Method to Recognize the Steady-State Visual Evoked Potentials with Wide Frequency Range
2018
[26] Calculation of absorbed dose in radiotherapy by solution of the linear Boltzmann transport equations
2018
[27] Dosimetry at extreme non-charged particle equilibrium conditions using Monte Carlo and specialized dosimeters
2018
[28] Farklı tedavi algortimalarının SRS ve SBRT tedavi planlamalarına etkisinin dozimetrik incelenmesi
2017
[29] Tuning of Acuros XB source size setting for small intracranial targets
Journal of applied clinical medical physics, 2017
[30] Small field depth dose profile of 6 MV photon beam in a simple air-water heterogeneity combination: A comparison between anisotropic analytical algorithm …
2017
[31] Tuning of AcurosXB source size setting for small intracranial targets
Journal of applied clinical medical physics, 2017
[32] A comparison of two dose calculation algorithms—anisotropic analytical algorithm and Acuros XB—for radiation therapy planning of canine intranasal tumors
Veterinary Radiology & Ultrasound, 2017
[33] High performance modelling of the transport of energetic particles for photon radiotherapy
Physica Medica, 2017
[34] Small field depth dose profile of 6 MV photon beam in a simple air-water heterogeneity combination: A comparison between anisotropic analytical algorithm …
2017
[35] Collapsed cone dose calculations for heterogeneous tissues in brachytherapy using primary and scatter separation source data
Computer Methods and Programs in Biomedicine, 2017
[36] Modelo semianalítico para cálculos de dosimetría en radioterapia
2016
[37] Evaluation of the dose calculation accuracy for small fields defined by jaw or MLC for AAA and Acuros XB algorithms
Medical physics, 2016
[38] Development of Unfolding Energy Spectrum with Clinical Linear Accelerator based on Transmission Data
Journal of Radiation Protection and Research, 2016
[39] Stereotaktisten annossuunnitelmien verifiointi Compass-järjestelmällä
2016
[40] 물질투과율 측정정보 기반 의료용 선형가속기의 에너지스펙트럼 유도기술 개발
방사선방어학회지, 2016
[41] Percentage depth dose calculation accuracy of model based algorithms in high energy photon small fields through heterogeneous media and comparison with plastic scintillator dosimetry
Journal of Applied Clinical Medical Physics, 2016
[42] Percentage depth dose calculation accuracy of model based algorithms in high energy photon small fields through heterogeneous media and comparison with plastic …
Journal of Applied Clinical Medical Physics, 2016
[43] Percentage depth dose calculation accuracy of model based algorithms in high energy photon small fields through heterogeneous media and comparison with …
2016
[44] Percentage depth dose calculation accuracy of model based algorithms in high energy photon small fields through heterogeneous media and comparison …
Journal of Applied Clinical Medical Physics, 2016
[45] A Comparison between Three-Dimensional Conformal Radiotherapy, Intensity-Modulated Radiotherapy, and Volumetric-Modulated Arc Therapy Techniques …
2015
[46] Evaluation of AAA and XVMC Algorithms for Dose Calculation in Lung Equivalent Heterogeneity in Photon Fields: Comparison of Calculated Results with …
2015
[47] Comparative evaluation of modern dosimetry techniques near low‐and high‐density heterogeneities
Journal of Applied Clinical Medical Physics, 2015
[48] Evaluation of a novel reference chamber “stealth chamber” through Monte Carlo simulations and experimental data
International Journal of Cancer Therapy and Oncology, 2015
[49] Comparative evaluation of modern dosimetry techniques near low-and high-density heterogeneities
Journal of Applied Clinical Medical Physics, 2015
[50] Evaluation of AAA and XVMC Algorithms for Dose Calculation in Lung Equivalent Heterogeneity in Photon Fields: Comparison of Calculated Results with Measurements
Journal of Biomedical Physics and Engineering, 2015
[51] A Comparison between Three-Dimensional Conformal Radiotherapy, Intensity-Modulated Radiotherapy, and Volumetric-Modulated Arc Therapy Techniques for …
2015
[52] Ganesh Narayanasamy, Drew Granatowicz, Diana Baacke, Ying Li, Alonso Gutierrez
2015
[53] A comparison between three-dimensional conformal radiotherapy, intensity-modulated radiotherapy, and volumetric-modulated arc therapy techniques for …
2015
[54] Monte Carlo Based Algorithms Are More Accurate for Dose Calculations in Radiotherapy
Biomedical Science and Engineering, 2014
[55] Clinical dosimetric impact of Acuros XB and analytical anisotropic algorithm (AAA) on real lung cancer treatment plans
2014
[56] Acuros XB for dose calculations in external beam photon radiotherapy-mini review
Oncol, 2014
[57] Clinical dosimetric impact of Acuros XB and analytical anisotropic algorithm (AAA) on real lung cancer treatment plans: review
International Journal of Cancer Therapy and Oncology, 2014
[58] Importance of Treatment Planning Systems in Radiation Therapy for Cancer Treatment-A Brief Report
Journal of Radioprotection Research, 2014
[59] Photon Dose Calculation Algorithms in the Presence of Low-Density Heterogeneity-a Mini Literature Review
Applied Mathematics, 2014
[60] The accuracy of the Acuros XB algorithm in external beam radiotherapy–a comprehensive review
International Journal of Cancer Therapy and Oncology, 2014
[61] Dose-to-medium vs. dose-to-water: Dosimetric evaluation of dose reporting modes in Acuros XB for prostate, lung and breast cancer
International Journal of Cancer Therapy and Oncology, 2014
[62] A Monte Carlo approach to lung dose calculation in small fields used in intensity modulated radiation therapy and stereotactic body radiation therapy
Journal of cancer research and therapeutics, 2014
[63] Üç boyutlu yazıcı ile üretilen fantomun radyocerrahi tedavi planlarının kalite kontrolünde kullanılması
Free SCIRP Newsletters
Copyright © 2006-2024 Scientific Research Publishing Inc. All Rights Reserved.
Top