Coronary Flow Reserve Assessment of Orbital Atherectomy: An Observational Pilot Study

HTML  XML Download Download as PDF (Size: 737KB)  PP. 269-275  
DOI: 10.4236/jbise.2019.124019    1,299 Downloads   2,678 Views  

ABSTRACT

The aim of this prospective observational pilot study was to observe the impact of orbital atherectomy (OA) on the coronary microcirculation via coronary flow reserve (CFR) measurements. Fifteen subjects who had successful OA and stent placement with no procedural complication were enrolled at 3 hospitals in the U.S. Baseline and hyperemic velocities were 16 ± 5.2 and 36 ± 14 cm/sec, respectively. The average CFR post-procedure was within the normal range at 2.23 ± 0.33. The observation of normal CFR following OA may be attributed to the orbital action of the device that allows for continuous flow during treatment, minimizing a bolus embolization effect which can impact microvascular function.

Share and Cite:

Dib, N. , Martinsen, B. , Anose, B. , Hodgson, J. , Kohler, R. , Nseir, G. , Antonios, N. , Shlofmitz, R. and Chambers, J. (2019) Coronary Flow Reserve Assessment of Orbital Atherectomy: An Observational Pilot Study. Journal of Biomedical Science and Engineering, 12, 269-275. doi: 10.4236/jbise.2019.124019.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.