Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom

Abstract

The Adaptive Quality Control Phantom (AQCP) is the computer-controlled phantom which positions and moves a radioactive source in the Field of View (FOV) of an imaging nuclear medicine device on a definite path to produce any spatial distribution of gamma rays to perform the QC Tests such as the Collimator Hole Angulation (CHA) of Single Photon Emission Computer Tomography (SPECT). The collimator hole angulation for seven collimators were measured with the method by using a point source and computer-controlled cylindrical positioning. In this method the displacement of the image of a point source examined as the AQCP move point source vertically away from the collimator face. The results of the high-accuracy measurement method of CHA show that the measurement accuracy for absolute angulation errors is better than ±0.024°. The Root Mean Square (RMS) of CHA for LEHR, LEHS, LEUHR, MEGP-250, MEGP-300, MEGP-360 and HEPH collimators of SMV dual heads camera were measured to be 0.290°, 0.292°, 0.208°, 0.194°, 0.181°, 0.177°, 0.150°, respectively. The Root Mean Square (RMS) of CHA for LEGP, MEGP and HEGP collimators of GE Millennium MG were measured to be 0.154°, 0.220° and 0.202° respectively. It is to be added in this connection that the measured RMS of CHA for LEHR collimator with the distance variation from the collimator’s surface +/– 1 mm has been varied +/– 0.04 degree.

Share and Cite:

H. Zeinali, E. Goodarzi, H. Ravanbakhsh, A. Sardarpour, N. Mahabadi, S. Moradkhani, F. Dolatshah and D. Rahi, "Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom," Modern Instrumentation, Vol. 1 No. 4, 2012, pp. 49-53. doi: 10.4236/mi.2012.14007.

Conflicts of Interest

The authors declare no conflicts of interest.

References

[1] R. E. Malmin, C. S. Peggy and W. R. Guth, “Collimator Angulation Error and Its Effect on SPECT,” Journal of Nuclear Medicine, Vol. 31, No. 5, 1990, 655-659.
[2] E. Busemann-Sokole, “Measurement of Collimator Hole Angulation and Camera Head Tilt for Slant and Parallel Hole Collimators Used in SPECT,” Journal of Nuclear Medicine, Vol. 28, No. 10, 1986, pp. 1592-1598.
[3] W. Chang, S. Q. Li, J. J. Williams, et al., “New Methods of Examining Gamma Camera Collimators,” Journal of Nuclear Medicine, Vol. 29, No. 5, 1988, pp. 674-683.
[4] American Association of Physicists in Medicine, “Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control,” AAPM Report No. 22, College Park, 1987.
[5] International Atomic Energy Agency, “Quality Control of Nuclear Medicine Instruments,” IAEA TECDOC 602, Vienna, 1991.
[6] M. Eckholt and H. Bergmann, “Angulation Errors in Parallel-Hole and Fan Beam Collimators: Computer Controlled Quality Control and Acceptance Testing Procedure,” Journal of Nuclear Medicine, Vol. 41, No. 3, 2000 pp. 548-555.
[7] Y. Takahashi, K. Murase, H. Higashno, et al., “SPECT Imaging with Off-Set Detector System: Comparison of Sampling Angles 2, 4 and 6 Degrees,” Annals of Nuclear Medicine, Vol. 16, No. 5, 2002, pp. 343-347. doi:10.1007/BF02988623
[8] H. Hines, R. Kayayan, J. Colsher, et al., “Recommendations for Implementing SPECT Instrumentation Quality Control. Nuclear Medicine Section—National Electrical Manufacturers Association,” European Journal of Nuclear Medicine and Molecular Imaging, Vol. 26, No. 5, 1999, pp. 527-532. doi:10.1007/s002590050421
[9] G. B. Thomas, “Calculus and Analytic Geometry,” 4th Edition, Addison-Wesley Publishing Company, Boston, 1969.
[10] B. C. Kuo, “Automatic Control Systems,” 5th Edition, McGraw-Hill Companies, New York, 1987.
[11] Web Site of National Instruments Corporation. www.NI.com.
[12] H. Z. Zeinali, M. Ghiassi-Nejad and A. Mirzai, “Design an Adaptive Quality Control Phantom to optimize QC Test Methods,” The Open Medical Imaging Journal, Vol. 1, 2007, pp. 1-6.
[13] H. Z. Zeinali and A. Mirzai, “Design an Adaptive Quality Control Phantom to optimize Integral and Differential Uniformity, Collimator Hole angulation and Center of Rotation of SPECT,” The Journal of Instrumentation, Vol. 3, 2008, pp. 1-8. doi:10.2174/157340508784356743
[14] H. Z. Zeinali, M. Ghiassi-Nejad and A. Mirzai, “Design an Adaptive Quality Control Phantom to Optimize Uniformity, Collimator Hole Angulation and Center of Rotation of SPECT,” The Journal of Instrumentation, 2008.www.iop.org/EJ/jinst

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.