Open Journal of Anesthesiology

Vol.8 No.4(2018), Paper ID 83867, 10 pages

DOI:10.4236/ojanes.2018.84012

 

Accuracy of the SpineNav3DTM Technology to Measure the Depth of Epidural Space: A Comparison with the Standard Ultrasound Technique in Pregnant Volunteers

 

Giorgio Capogna, Simone Baglioni, Valentina Milazzo, Arturo Vitale

 

Department of Anesthesiology, Città di Roma Hospital, Roma, Italy
Department of Anesthesiology, ASST S. Gerardo-Monza, Monza, Italy
Department of Anesthesiology, University of Catania, Catania, Italy
Statistical Consultant, Roma, Italy

 

Copyright © 2018 Giorgio Capogna, Simone Baglioni, Valentina Milazzo, Arturo Vitale et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

 

How to Cite this Article


Capogna, G. , Baglioni, S. , Milazzo, V. and Vitale, A. (2018) Accuracy of the SpineNav3DTM Technology to Measure the Depth of Epidural Space: A Comparison with the Standard Ultrasound Technique in Pregnant Volunteers. Open Journal of Anesthesiology, 8, 113-122. doi: 10.4236/ojanes.2018.84012.