Validation and Comparison of Two Calibration Methods for the Measurement of Stable Iodine in the Urinary Matrix by ICP-MS: Standard Addition vs. External Calibration
Géraldine Landon, Céline Bouvier-Capely, Alexandre Legrand, Thibaud Sontag, Grégory Finance, Maïlie Saint-Hilaire, François Rebière, Xavier Millot, Valérie Renaud-Salis, Michelle Agarande
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
The French Alternative Energies and Atomic Energy Commission (CEA), Gif-sur-Yvette, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Institute for Radiological Protection and Nuclear Safety (IRSN), Fontenay-aux-Roses, France
Copyright © 2017 Géraldine Landon, Céline Bouvier-Capely, Alexandre Legrand, Thibaud Sontag, Grégory Finance, Maïlie Saint-Hilaire, François Rebière, Xavier Millot, Valérie Renaud-Salis, Michelle Agarande et al. This is
an open access article distributed under the Creative Commons Attribution
License, which permits unrestricted use, distribution, and reproduction in any
medium, provided the original work is properly cited.
How to Cite this Article
Landon, G. , Bouvier-Capely, C. , Legrand, A. , Sontag, T. , Finance, G. , Saint-Hilaire, M. , Rebière, F. , Millot, X. , Renaud-Salis, V. and Agarande, M. (2017) Validation and Comparison of Two Calibration Methods for the Measurement of Stable Iodine in the Urinary Matrix by ICP-MS: Standard Addition vs. External Calibration.
American Journal of Analytical Chemistry,
8, 245-263. doi:
10.4236/ajac.2017.84019.