American Journal of Analytical Chemistry
Vol.5 No.3(2014), Article ID:43090,11 pages DOI:10.4236/ajac.2014.53023

Flow Injection Amperometric Determination of Phenol and Chlorophenols at Single Wall Carbon Nanotube Modified Glassy Carbon Electrode

Negussie Negash1, Hailemichael Alemu2*, Merid Tessema1

1Department of Chemistry, Addis Ababa University, Addis Ababa, Ethiopia

2Department of Chemistry and Chemical Technology, National University of Lesotho, Roma, Lesotho

Email: *hm.alemu@nul.ls, *hmalemu@yahoo.com

Copyright © 2014 Negussie Negash et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. In accordance of the Creative Commons Attribution License all Copyrights © 2014 are reserved for SCIRP and the owner of the intellectual property Negussie Negash et al. All Copyright © 2014 are guarded by law and by SCIRP as a guardian.

Received November 16, 2013; revised December 21, 2013; accepted December 30, 2013

KEYWORDS

Single Wall Carbon Nanotube Modified Electrode; Phenol and Chlorophenols; Amperometry; Flow Injection Analysis

ABSTRACT

Single wall carbon nanotube modified glassy carbon electrode (SWCNT/GCE) was used for flow-injection analysis (FIA) for phenolic compounds (phenol (P), 4-chlorophenol (CP), 2,4-dichlorophenol (DCP), 2,4,6-trichlorphenol (TCP) and pentachlorophenol (PCP)). Experimental variables such as the detection potential, flow rate and pH of the carrier solution, 0.1 M sodium acetate, were optimized. Under these conditions, the designed electrode showed a very good performance for the amperometric measurements, with no need to apply a cleaning or pre-treatment procedure. The operational stability was tested with 20 repetitive injections of each analyte and was found to be good. The analytical performance of the SWCNT/GCE electrode under flow through conditions was tested and was found to be impressive. When it is compared with other enzymatic and non-enzymatic sensors, it shows wider dynamic range for the detection of phenolic compounds with low limits of detection. These results suggest that the method is quite useful for monitoring and analyzing phenols and chlorophenols.

1. Introduction

Modification of conventional electrode surfaces is necessary in sensor constructions as they can be used as a sensing platform. In electrochemistry, carbon based electrodes have been widely used, because of their low cost, low background current, wide potential window and biocompatibility [1,2]. Among several forms of carbon used as electrochemical sensors, carbon nanotubes (CNT) stand out as the most promising material for the development of sensors due to their fast electron transfer kinetics and enhanced electronic properties. CNTs also display high electrical conductivity, chemical stability and mechanical strength. CNT modified electrodes exhibit low limit of detection and fast responses due to their high surface area, low over voltage and rapid electrode kinetics [2-7].

The two main classes of CNTs are single-walled carbon nanotubes (SWCNTs) and multi-walled carbon nanotubes (MWCNTs). SWCNTs are sp2 hybridized carbon that possesses a cylindrical nano-structure that has only one single layer of graphene cylinder [8]. SWCNTs can be classified as either semi-conducting or metallic allotropes depending on the chirality [2]. MWCNTs have a complex structure with each carbon layer having different chiralities and electronic properties and consist of several layers of graphite cylinders that are concentric tubes encircling one another [4]. CNTs have attracted increasing interest in the application of CNTs based sensors in the detection and determination of phenols and phenolic compounds [3,9-12]. Phenolic compounds include a large class of compounds that are common in nature. They are widely distributed in fruits and vegetables with varying phenolic content [13]. They are partly formed as a result of biodegradation of natural compounds like humic acids, tannins and lignins. Phenolics are a complex group of substances that provide flavor and color characteristic of food, since most of them occur in food products. Many of the benefits associated with consumption of phenolrich foods are associated with their antioxidant activities which are determined by their reactivity as hydrogen or electron donating agent [14-16]. Phenols and phenol derivatives are commonly used in the manufacture of dyes, plastics, drugs, explosives, detergents, paper mills and in the petroleum industry. They are used to manufacture chemicals that are used as preservatives for woods, textile and leathers [17] and have a considerable application in agriculture as herbicides, insecticides and fungicides. Most phenolic compounds, especially chlorophenols and nitro phenols, are toxic, highly resistant to biological degradation and persistent in the environment. Chlorinated phenols such as 2-chlorophenoll, 2,4-dichlorophenol, 2,4, 6-trichlorophenol and pentachlorophenol are toxic and possess carcinogenic and immunosuppressive properties [18]. As a result, the United State Environmental Protection Agency (EPA) and the European Union have included these phenolic compounds in their list of primary pollutants [19-21].

The applications of phenol and its derivatives represent potential source of pollution and are usually found in waste waters’ effluents from different factories using phenolic compounds and manufacturing petrochemicals [17, 22]. Phenol contamination of water bodies also has serious environmental implications because of the damaging effects it has on aquatic organisms [22] including algae and aquatic spermatophytes. Phenolic derivatives are released to the environment, and their identification and quantifications are of great importance in environmental monitoring.

Various methods, such as high-performance liquid chromatography with different detectors, liquid chromatography-mass spectrometry, gas chromatography mass spectrometry, capillary electrophoresis and fluorometry, have been reported for the determination of phenols and phenol derivatives [15,17,18,23-30]. These conventional methods are sensitive, reliable, specific and precise. In spite of their advantage, most of these techniques need expensive instrumentations, require skilled technicians and are time-consuming, and some require derivatization procedures.

The sensitive, rapid and precise determination of phenols and its derivatives is of growing interest in environmental control and protection. Flow injection analysis (FIA) has been widely used for the determination of phenolic compounds because of its outstanding features of high sampling rate, small sample volumes, good accuracy and precision of results, high reproducibility, high versatility, high robustness and easy automation [31-33]. A number of research out-puts were published on the amperometric detection of phenol and its derivatives using tyrosnase (Tyr) [31,34], laccase [35,36], horse-radish peroxidase [37,38], as well as poly phenol oxidase based electrochemical sensors coupled with flow-injection analysis for the determination of phenol and phenol derivatives. These enzymes were either individually immobilized or a composite multienzyme system immobilized on conventional electrodes to determine phenolic compounds [39]. The analytical performances of enzyme based biosensors strongly depend on the immobilization procedure and accuracy, sensor-to-sensor reproducibility and operational life times and are drastically affected by enzyme stability [40]. The poor biocompatibity and low capacity for immobization of these enzymes also result in inefficient performance of some biosensors [41].

To our understanding, CNT modified electrodes for the determination of phenols and chlorophenols as an electrochemical sensor for flow injection amperometric analysis system have not been reported. The simplest electrochemical flow detector, SWCNT modified GC electrode, was developed successfully for the determination of phenols and chlorophenols. The parameters such as operational potential, pH and flow rates were investigated and optimized.

2. Experimental

2.1. Reagents and Apparatus

SWCNT was obtained from Saarchem Holparo Analytical (Pty) Ltd. Phenol, 4-chlorophenol, 2,4-dichlorophenol, 2,4,6-trichlorophenol, and pentachlorophenol were purchased from Fluka (Buchs, Switzerland). Acetic acid and DMF were obtained from (Merck). O-nitrophenol was from Riedelde Haen and other chemicals were obtained from SAAR-CHEM, Republic of South Africa (RSA). Helium gas (Air Products, RSA) was employed for the de-aeration of the carrier solution. All chemicals were analytical grade and were used without further purifications. Ultrapure water of resistivity 18.2 MW·cm was obtained from ELGA PURELAB Option-Q (UK) water purification system and was used throughout.

Electrochemical measurements were performed using Epsilon Electrochemical Analyzer (BASi) attached to BAS C2 cell stand (BASi Instrumentation, USA). A conventional three electrode cell was used for measurements, with a bare glassy carbon electrode (3 mm diameter, BASi) or SWCNT/GCE used as the working electrode, Ag/AgCl (3 M NaCl) (BASi) as the reference electrode and a platinum wire auxiliary electrode (BASi). The flow system consisted of a BAS pump (Model PM-92E; USA), a sample injection valve, and a thin-layer electrochemical detector (BASi LC-4C, USA), with a flow-through cell. The electrochemical detector was connected to the Epsilon Electrochemical Analyzer. Two GCE modified with SWCNT (in series configuration) were used as working electrodes, while Ag/AgCl electrode (3 M NaCl) (RE-1; BAS) served as the reference electrode. The counter electrode was the bulk plate of the cell, made of stainless steel. The thickness of the layer in the cell was 0.0127 cm. The pH of the solutions was measured using HANNA HI 8314 pH-meter.

2.2. Fabrication of SWCNT/GCE

The modified working electrode, SWCNT/GCE, was prepared by dispersing SWCNT in DMF (1 mg: 1 mL ratio) and sonicated with ultrsonication water bath for an hour [42]. Before surface modification, the bare glassy carbon electrode was polished with a 0.3 and 0.05 mm alumina slurry and washed with distilled water several times and sonicated for 3 minutes. The SWCNT film was prepared by casting a 20 μL of the dispersed solution over the bare GCE and dried overnight at room temperature.

2.3. Measurements

For the voltammetric measurements, 10 mL solution that contain 1 × 10−4 M phenol and chlorophenols, in a 0.1 M acetate buffer supporting electrolyte, were transferred to the cell. The cyclic voltammetry experiments were run after stirring the spiked solution for 30 seconds. The measurements were performed by scanning the potential from 0 to 1.2 V and back at a scan rate of 50 mV·s−1 for all the studied phenols. A 0.1 M sodium acetate buffer was also used as a carrier solution in the flow through experiments after de-aerating the solution with helium gas. A 20 μL of the sample solutions were injected at every experimental run.

3. Results and Discussion

3.1. Electrochemical Characterization of Phenol at SWCNT/GCE

The electrochemical characteristic of phenol at a bare glassy carbon electrode was studied using cyclic voltammetry. Figure 1 shows subsequent current/potential curves recorded for 1 × 10−4 M phenol in a 0.1 M sodium acetate buffer (pH 6.0) in the potential range of 0 to 1.2 V. Due to the electro-deposition of non-conductor oxidation product of phenol i.e. polymers [43-45], the oxidation peak at the bare glassy carbon electrode decreased markedly with successive cycles. The response disappeared totally after three cycles owing to the adsorption of phenol oxidation products which decrease the available electrode surface area. Normally the oxidation of phenol produces conducting monomers and non conducting polymer as shown by Scheme 1, which is responsible for such electrode fouling [46,47].

In the absence of phenol, only with the sodium acetate buffer solution, the bare GCE Figure 1(a), and SWCNT/ GCE, Figure 1(c) showed no oxidative or reduction peak, while the back ground current at SWCNT/GCE is much larger than the bare electrode. This indicates that the SWCNTs increased the effective surface area of the modified electrode [9] and also proved that SWCNTs have been successfully immoblized onto the bare electrode surface. For 1 × 10−4 M phenol solution of Figure 1(b), the voltammogram at the bare GCE showed irreversible behavior with Epa = 0.851 V [46], and at the SWCNT/ GCE there was also an irreversible peak, Figure 1(d), with Epa = 0.686 V with about twelve times intensified peak current compared to the bare GCE. This indicates that SWCNTs can greatly enhance the electron transfer rate, with significantly reduced over-potential. This implies that the SWCNTs show excellent electrocatalytic effect towards phenol oxidation.

On the other hand ten successive cyclic voltammetric curves of phenol recorded at SWCNT/GCE as shown in Figure 2 demonstrate a quite interesting behavior. During the first positively going scan, phenol undergoes oxidation in a single step, peak 1a, at Ep1a = +0.69 V. On the negative going scan, two reduction peaks occurred, peak 3c, at Ep3c = +0.36 V and peak 2c, at Ep2c = +0.10 V.

The cathodic peak corresponds to the reduction of the two phenol oxidation products, ortho-quinone to orthophenol (catechol), peak 3c, and para-quinone to paraphenol (hydroquinone), peak 2c, Scheme 1 [12,46]. The second scan in the positive potential direction shows two anodic peaks: peak 2a at Ep2a = +0.13 V, and peak Ep3a = +0.39 V, indicating the reversibility of the two peaks [46, 12]. This shows that the electroxidation path way of phenol at the modified electrode is different from that of the

Figure 1. Cyclic voltammograms: 0.1 M acetate buffer pH 6 base electrolyte at bare GCE (a); 0.1 M acetate buffer pH 6 base electrolyte at SWCNT/GCE (c); 1 × 10−4 M phenol at bare GCE (b); and 1 × 10−4 M phenol at SWCNT/GCE recorded at Scan rate, 50 mV∙s−1; pH 6.

Scheme 1. Phenol electro-oxidation and polymerization processes [12].

Figure 2. Overlay of 10 successive cyclic voltammograms for 1 × 104 M phenol at 50 mV∙s1.

bare GCE. At the modified electrode, unlike at the bare GCE, the oxidation peak current of phenol 1a shows a small decrease in intensity after the first scan, but is then practically remains constant for the subsequent 8 scans. The peak potential observed ranged between 0.685 - 0.690 V versus Ag/AgCl and this makes it suitable to be used as an electrochemical flow through detector for the determination of phenol and chlorophenols in flow injection system.

3.2. Flow Injection Analysis

The electrochemical performance of the glassy carbon electrode modified with SWCNT was studied in 0.1 M acetate buffer solution using flow injection analysis. The parameters optimized to determine phenol and chlorophenols were: potential, flow rate and pH. For comparison, similar experiments were carried out using bare GCE during potential optimization.

3.2.1. Effect of Potential on the Response of SWCNT/GCE Electrode

The influence of the applied potential on the amperometric responses of the two electrodes for phenol and chlorophenols was examined using flow injection system. Figure 3 shows the hydrodynamic voltammograms obtained at bare GCE and SWCNT/GCE for 1 × 10−4 M phenol and chlorophenols, respectively. The influence of the applied potential on phenol and chlorophenols oxidation was studied over a potential range of 0.1 - 1.0 V. The maximum oxidation potential observed was between 0.6 and 0.8 V. This was similar to the maximum current observed between 0.65 to 0.80 V for 2,4,6-trichlorophenol [47]. As it is shown in Figure 3, the effect of the SWCNT/GCE can be clearly seen by the significant increase of the peak current compared to the bare GCE. Such effect is attributed to the fast electronic transfer offered by SWCNT which enhances the oxidation of phenols and chlorophenols [8,48]. The bare glassy carbon electrode responded favorably only to the first three injections with low amperometric current and exhibited a poorly defined response as reported [9].

3.2.2. Effect of pH on the Response of SWCNT/GCE

The effect of pH on the performance of GCE modified with SWCNT was evaluated by using 0.1 M acetate buffer solution over the pH range of 3 - 10 for phenol and chlorophenols. For each analyte the amperometric response was recorded at a fixed potential. The result shows that, (Figure 4), the amperometric response increases with increasing pH until it reaches the range of pH 6 to 7.5, where the higher peak current values were observed. The highest peak current for 2,4,6-trichlorophenol was observed at about pH 9. The current decreases as the pH increases. The optimum pH range that gave maximum current is similar to that previously reported for both enzymatic and mediated electrochemical reactions at carbon nanotubes [49], a screen-printed sensor based on immobilization of laccase, peroxidase, and tyrosinase for monitoring phenols [39,50] and a wide optimum pH of 5 - 8 for free enzyme [51]. It seems that the SWCNT/GCE composite has not altered the optimum pH of phenol and chlorophenols oxidation significantly. For further studies, pH of 7 for phenol and pentachlorophenol, pH 6 for 4-chlorophnol and 2,4-dichlorophenol and pH 9 for trichlorophenol were selected, respectively.

3.2.3. Effect of Flow Rate on the Response of SWCNT/GCE

One of the basic parameters which influence the performance of flow injection analysis is flow rate. The flow rate affects detection limit, sensitivity, and accuracy of analysis. The influence of flow rate on the amperometric signal was examined in the 0.1 - 2.0 mL∙min−1 range for all analytes at SWCNT/GCE. To investigate the effect of flow rate, 20 µL of 1 × 10−4 M phenol and chlorophenols were injected at various flow rates. The applied potential

Figure 3. Dependence for 1 × 10−4 M solutions of the current response in FIA on the applied working potential at GCE (B) and SWCNT/GCE (A), 2,4,6-trichlorphenol (TCP), 2,4-dichlorophenol (DCO), 4-chlorophenol (CP), pentachlorophenol (PCP) and phenol: Flow rate, 1.0 mL∙min−1; carrier solution: sodium acetate buffer (0.1 M); pH 6, injection volume 20 µL.

Figure 4. Variation of amperometric current responses with pH at the SWCNT/GCE for 1 × 10−5 M phenol and 1 × 10−4 M for the others; and Eapplied for phenol (0.725 V), for 2,4- dichlorophenol (0.75 V), for 2,4,6-trichlorophenol, (0.7 V), and for pentachlorophenol, (0.8 V), vs Ag/AgCl; flow rate 1 mL/min, injection volume 20 µL.

was constant for each analyte. As it is shown in Figure 5, similar trends were observed for phenol and all chlrophenols. It can be seen that the peak current increases in the range of 0.1 - 1 mL∙min−1. At low flow rates, the mass transfer of phenolic analytes to the electrode surface is relatively low [52]. Also a broader peak and low resolution were observed. For relatively high flow rates there was a decrease in the flow-injection peak current which is similar to the usual behavior for enzyme-based FIA [50]. At higher flow rates, above 1.2 mL∙min−1, the electron transfer and mass transfer rate are not significantly different. Moreover as the sample passes faster over the electrode the smaller is the fraction of analyte oxidized [36] and as a result the amperometric current

Figure 5. Flow rate dependence of the peak current in FIA: (a) for 1 × 10−5 M Phenol at pH 7, Eapp, 0.725 V, (b) 1 × 10−4 M Pentachlorophenol at pH, 7; (c) 2,4,6-Trichlorophenol with at pH 9; (d) 2,4-Dichlorophenol pH 6; (e) 4-Chlorophenol, at pH 6; Eapp as in Figure 4; Carrier, 0.1 M sodium acetate.

decreases. The optimum flow rate of the carrier solution for acceptable sharp peaks and sensitivity were in the range of 0.9 and 1.2 mL·min−1.

3.2.4. Stability of SWCNT/GCE

The performance of the SWCNT/GCE as a flow injection detector was found to be remarkably good and the amperometric response of the sensor showed a good stability, since after 20 injections the response remains the same. Figure 6(a) shows the amperometric response measured for 2,4,6-trichlorophenol. At each step 20 µL of: 8 × 10−6 M phenol and 4-chlorophenol, 1.5 × 10−5 M 2,4-dichlorophenol, 1 × 10−4 M 2,4,6-trichlorophenol and 1 × 10−5 M pentachlorophenol, were injected. Figure 6(b) shows the operational stability of the amperometric peak current,

(a)(b)

Figure 6. Amperometric flow injection response of 2,4,6- trichlorophenol (a) and stability plot (b) obtained for 20 repetitive injections of 1 × 10−5 M phenol and pentachlorophenol; 6 × 10−5 M 4-chlorophenol; 1.5 × 10−5 M 2,4-dichlorophenol and 1 × 10−4 M 2,4,6-trichlorophenol at a flow rate: 1.1 mL∙min−1 for phenol, 1 mL∙min−1 for 4-chlorophenol and 2,4-dichlorophenol, 1.2 mL∙min−1 2,4,6-trichlorphenol and pentachlorophenol; Other conditions are as in Figure 5.

ip, for 20 successive injections of each analyte. The response had a relative standard deviation ranging between 0.94% and 3.20%, as shown in Table 1. These results are due to the fact that flowing liquid solutions continuously clean the surface of the cell and remove the reaction products and impurities leached from the electrode [33, 53]. This experimental results are in good agreement with the stability of CNT modified electrodes [53,54] and different enzymztic biosensor [36,50]. Such additional advantages of the flow injection method based on the SWCNT/GCE modified detector broaden the possibilities of determining different phenol derivatives.

3.2.5. Analytical Performance of the SWCNT/GCE Detector in the Flow Injection System

In flow injection analysis, the flow-through detection cell monitors the concentration-time profile of the analyte. The behavior of SWCNT/GCE modified electrode was investigated in flow injection for the determination of phenol and different chlorophenls under the optimized conditions. Increasing concentrations of phenol and chlorophenols were injected to the carrier solution (0.1 M sodium acetate). Figure 7(a) displays the amperometric response of the SWCNT/GCE for 4-chlorophenol and it can be seen that the peak current increases with increasing the concentration of 4-chlorophenol. This wellshaped amperometric peaks indicate that the surface sensing properties of the modified electrode were efficient and reproducible mass transport towards the electrode surface was observed which goes in line with [53, 55]. The amperometric peak heights increase with increase in the concentration of 4-chlorophenol. Figure 7(a) was used to construct the analytical calibration curves of the SWCNT/GCE composite electrode under the optimum experimental conditions for each analyt. Figure 7(b) shows the calibration curve for 4-chlorophenol. Table 2 summarizes the analytical characteristics of the calibration graphs. The limits of detection were calculated on the 3 sb/m, where m is the slope of the linear calibration graph for each analyte, and sb was estimated as the stan-

Table 1. Reproducibility of the stability measurements obtained with SWCNT/GCE in flow-injection analysis.

Table 2. Analytical characteristics of the calibration plots for different phenolic compounds in flow-injection system with the SWCNT/GCE detector.

(a)(b)

Figure 7. Amperometric responses for different concentrations of 4-chlorophenol (a-k) of concentrations 2, 4, 6, 8, 9, 10, 12, 14, 16, 18, 20, 22 mM, respectively (a), and the calibration curves (b), Eapp, 0.8 vs Ag/AgCl; flow rate 1.0 mL∙min−1, pH, 6; carrier solution 0.1 M sodium acetate.

dard deviation (n = 10) of the response obtained from the repetitive analyte injections at the lowest concentration. Using this criterion, detection limits ranging between 0.011 and 0.31 mM were achieved. As summarized in Table 2, the SWCNT/GCE system used as a detector exhibits a good linear response between the amperometeric peak currents and concentrations of phenol and chlorophenols with low detection limits for all phenolic compounds investigated. The analytical performance of the SWCNT/GCE determined in the flow through system has been compared with other enzyme/enzymeless electrochemical sensors reported recently as shown in Table 3. Characteristics such as, range of linearity, sensitivity and limit of detection achieved and type of electrode were compared. It can be clearly seen from Table 3 that the SWCNT/GCE sensor offers advantages as compared to the other detector designs. Apart from the simplicity in its preparation its dynamic range for the detection of phenolic compounds is wider.

3.2.6. Interference Studies

The amperometric response of the SWCNT/GCE modified electrode for to 5 × 10−6 phenol was not affected by the addition of oand p-nitrophenol up to the concentration of to 5 × 10−4, which is in a ratio of 1:100. Above this concentration the peak current increases. Similarly the addition of the same concentrations of 4-chlorophenol, 2,4-dichlorophenol, 2,4,6-trichlorophenol and pentachlorophenol to 5 × 10−6 M phenol has no effect on the amperometric responses of the electrode. With further increase in the concentration of chlorophenols added an increase in the response current was also observed. These can be attributed to their similarity in the optimum operational conditions which resulted in the oxidation of phenolic OH group. In fact, the selectivity of most commonly used amperometric biosensors such as tyrosinase, laccase, and peroxidase, modified electrodes for phenolic compounds are relatively higher due to the use of low applied potentials [35]. The SWCNT/GCE modified electrode needs higher potential to be able to oxidize the phenol; therefore the risk of electro-oxidation interfering compounds in the sample is higher.

3.2.7. Analytical Application

The analytical applications of the modified electrode were investigated within the linear range. With phenol as the analyte the estimation of content was carried out by injecting a 2 μM phenol to deionizzed water. Increased concentrations of phenol were injected and the amperometric responses were recorded which were used for recovery calculations. As shown in Table 4 the recoveries were between 99.7% and 108.36%, which confirm the designed electrode used in the method is reliable for quantitative determination of phenol and chlorophenols.

4. Conclusion

In this paper, we presented one simple and low cost electrochemical sensor based on SWCNT modified GCE for the analysis of phenol and chlorophenols using flow injection amperometric methods. The SWCNT modified electrode has a great role in minimizing of surface fouling due to its high electrocatalytic activity. The analytical performance obtained after optimizing experimental variables was impressive, and the electrode has a good operational stability, linear range and detection limits. Although the sensor has limited selectivity for these analytes, the electrode is simple in its preparation.

Acknowledgements

Negussie Negash is grateful to the Graduate School of

Table 3. Comparison of analytical performance of the calibration curves of SWCNT/GCE with different modified electrodes for phenol, 4-chlorophenol, 2,4,6-trichlorophenol, and pentachlorophenol.

Table 4. Phenol determination.

Addis Ababa University, for financial support and the National University of Lesotho for giving him the chance to use the resources and facilities of their research laboratory.

REFERENCES

  1. I. E. Mülazımoğlu, A. D. Mülazımoğlu and E. Yılmaz, “Determination of Quantitative Phenol in Tap Water Samples as Electrochemical Using 3,3’-Diaminobenzidine Modified Glassy Carbon Sensor Electrode,” Desalination, Vol. 268, No. 1, 2011, pp. 227-232. http://dx.doi.org/10.1016/j.desal.2010.10.033
  2. C. B. Jacobs, M. J. Peairs and B. J. Venton, “Review: Carbon Nanotube Based Electrochemical Sensors for Biomolecules,” Analytica Chimica Acta, Vol. 662, No. 2, 2010, pp. 105-127. http://dx.doi.org/10.1016/j.aca.2010.01.009
  3. K. Balasubramanian and M. Burghard, “Biosensors Based on Carbon Nanotubes,” Analytical and Bioanalytical Chemistry, Vol. 385, No. 3, 2006, pp. 452-468. http://dx.doi.org/10.1007/s00216-006-0314-8
  4. S. K. Vashist, D. Zheng, K. Al-Rubeaan, J. H. T. Luong and F.-S. Sheu, “Advances in Carbon Nanotube Based Electrochemical Sensors for Bioanalytical Applications,” Biotechnology Advancement, Vol. 29, No. 2, 2011, pp. 169-188. http://dx.doi.org/10.1016/j.biotechadv.2010.10.002
  5. D. W. H. Fam, A. I. Palaniappan, A. I. Y. Tok, B. Liedberg and S. M. Moochhala, “A Review on Technological Aspects Influencing Commercialization of Carbon Nanotube Sensors,” Sensors and Actuators B, Vol. 157, No. 1, 2011, pp. 1-7. http://dx.doi.org/10.1016/j.snb.2011.03.04
  6. K. Scida, P. W. Stege, G. Haby, G. A. Messina and C. D. García, “Recent Applications of Carbon-Based Nanomaterials in Analytical Chemistry: Critical Review,” Analytica Chimica Acta, Vol. 69, No. 1, 2011, pp. 6-17. http://dx.doi.org/10.1016/j.aca.2011.02.025
  7. P. Yanez-Sedeno, J. Riu, J. M. Pingarron and F. X. Rius, “Electrochemical Sensing Based on Carbon Nanotubes,” TrAc-Trends in Analytical Chemistry, Vol. 29, No. 9, 2010, pp. 939-953. http://dx.doi.org/10.1016/j.trac..2010.06.00
  8. A. J. Ahammad, J. Lee and M. A. Rahman, “Electrochemical Sensors Based on Carbon Nanotubes,” Sensors, Vol. 9, No. 1, 2009, pp. 2289-2319. http://dx.doi.org/10.3390/s90402289
  9. J. Wang, R. P. Deo and M. Musameh, “Stable and Sensitive Electrochemical Detection of Phenolic Compounds at Carbon Nanotube Modified Glassy Carbon Electrodes,” Electroanalysis, Vol. 15, No. 23-24, 2003, pp. 1830-1834. http://dx.doi.org/10.1002/elan.200302772
  10. Q. Zhao, L. Guan, Z. Gu and Q. Zhuang, “Determination of Phenolic Compounds Based on the Tyrosinase-Single Walled Carbon Nanotubes Sensor,” Electroanalysis, Vol. 17, No. 1, 2005, pp. 85-88. http://dx.doi.org/10.10.10.1002/elan.200403123
  11. A. S. Arribas, E. Bermejo, M. Chicharro, A. Zapardiel, G. L. Luque, N. F. Ferreyra and G. A. Rivas, “Analytical Applications of Glassy Carbon Electrodes Modified with Multi-Wall Carbon Nanotubes Dispersed in Polyethylenimine as Detectors in Flow Systems,” Analytica Chimica Acta, Vol. 596, No. 2, 2007, pp. 183-194. http://dx.doi.org/10.1016/j.aca.2007.06.014
  12. S. Sundaram and S. K. Annamalai, “Selective Immobilization of Hydroquinone on Carbon Nanotube Modified Electrode via Phenol Electro-Oxidation Method and Its Hydrazine Electro-Catalysis and Escherichia Coli Antibacterial Activity,” Electrochimica Acta, Vol. 62, 2012, pp. 207-217. http://dx.doi.org/10.1016/j.electacta.2011.12.044
  13. M. Naczk and F. Shahidi, “Phenolics in Cereals, Fruits and Vegetables: Occurrence, Extraction and Analysis,” Journal of Pharmaceutical and Biomedical Analysis, Vol. 41, No. 5, 2006, pp. 1523-1542. http://dx.doi.org/10.1016/j.jpba.2006.04.002
  14. A. Simic, D. Manojlovic, D. Segan and M. Todorovic, “Electrochemical Behavior and Antioxidant and Prooxidant Activity of Natural Phenolics,” Molecules, Vol. 12, No. 10, 2007, pp. 2327-2340. http://dx.doi.org/10.3390/12102327
  15. C. Proestos, D. Sereli and M. Komaitis, “Determination of Phenolic Compounds in Aromatic Plants by RP-HPLC and GC-MS,” Food Chemistry, Vol. 95, No. 1, 2006, pp. 44-52. http://dx.doi.org/10.1016/j.foodchem.2004.12.016
  16. A. S. Arribas, M. Martınez-Fernandez and M. Chicharro, “The Role of Electroanalytical Techniques in Analysis of Polyphenols in Wine,” TrAc Trends in Analytical Chemistry, Vol. 34, 2012, pp. 78-96. http://dx.doi.org/10.1016/j.trac.2011.10.015
  17. A. Penalver, E. Pocurull, F. Borrull and R. M. Marce, “Solid-Phase Microextraction Coupled to High-Performance Liquid Chromatography to Determine Phenolic Compounds in Water Samples,” Journal of Chromatography A, Vol. 953, No. 1-2, 2002, pp. 79-87. http://dx.doi.org/10.1016/S0021-9673(02)00113-9
  18. G. W. Muna, V. Quaiserova-Mocko and G. M. Swain, “The Analysis of Chlorinated Phenol Solutions by Capillary Electrophoresis Coupled with Direct and Indirect Amperometric Detection Using a Boron-Doped Diamond Microelectrode,” Electroanalysis, Vol. 17, No. 13, 2005, pp. 1160-1170. http://dx.doi.org/10.1002/elan.200403230
  19. “Phenols in Federal Register: Environment Protection Agency Method 604,” Part VIII, 40 CFR Part 136, 1984, pp. 58-66.
  20. “Phenols by Gas Chromatography, Capillary Column Technique: Environment Protection Agency Method 804,” Part 1, Washington DC, 1995, pp. l-28.
  21. D. Puig and D. Barcelo, “Determination of Phenolic Compounds in Water and Waste Water,” TrAc Trends in Analytical Chemistry, Vol. 15, No. 8, 1996, pp. 362-365. http://dx.doi.org/10.1016/0165-9936(96)00057-X
  22. J.-S. Park, M. T. Brown and T. Han, “Phenol Toxicity to the Aquatic Macrophyte Lemna Paucicostata,” Aquatic Toxicology, Vol. 106-107, 2012, pp. 182-188. http://dx.doi.org/10.1016/j.aquatox.2011.10.004
  23. M. Krizman, D. Baricevic and M. Prosek, “Determination of Phenolic Compounds in Fennel by HPLC and HPLCMS Using a Monolithic Reversed-Phase Column,” Journal of Pharmaceutical and Biomedical Analysis, Vol. 43, No. 2, 2007, pp. 481-485. http://dx.doi.org/10.1016/j.jpba.2006.07.029
  24. M. Tasioula-Margari and O. Okogeri, “Simultaneous Determination of Phenolic Compounds and Tocopherols in Virgin Olive Oil Using HPLC and UV Detection,” Food Chemistry, Vol. 74, No. 3, 2001, pp. 377-383. http://dx.doi.org/10.1016/S0308-8146(01)00176-5
  25. F. Zhou, X. Li and Z. Zeng, “Determination of Phenolic Compounds in Wastewater Samples Using a Novel Fiber by Solid-Phase Microextraction Coupled to Gas Chromatography,” Analytica Chimica Acta, Vol. 538, No. 1-2, 2005, pp. 63-70. http://dx.doi.org/10.1016/j.aca.2005.02.009
  26. Y. Liang, W. Cao, W.-J. Chen, X.-H. Xiao and J.-B. Zheng, “Simultaneous Determination of Four Phenolic Components in Citrus Honey by High Performance Liquid Chromatography Using Electrochemical Detection,” Food Chemistry, Vol. 114, No. 4, 2009 pp. 1537-1541. http://dx.doi.org/10.1016/j.foodchem.2008.11.024
  27. E. Hurtado-Fernández, M. Gómez-Romero, A. CarrascoPancorbo and A. Fernández-Gutiérrez, “Application and Potential of Capillary Electroseparation Methods to Determine Antioxidant Phenolic Compounds from Plant Food Material,” Journal of. Pharmaceutical and Biomedical Analysis, Vol. 53, No. 5, 2010, pp. 1130-1160. http://dx.doi.org/10.1016/j.jpba.2010.07.028
  28. A. M. Naley, “Analysis of Phenols in Sea Water by Fluorometry: Direct Analysis of the Water Phase,” Bulletin of Environmental Contamination and Toxicology, Vol. 3, No. 4, 1983, pp. 494-500. http://dx.doi.org/10.1007/BF01622283
  29. Y. Tsuruta, S. Watanabe and H. Inoue, “Fluorometric Determination of Phenol and p-Cresol in Urine by Precolumn High-Performance Liquid Chromatography Using 4-(N-Phthalimidinyl)benzenesulfonyl Chloride,” Analytical Biochemistry, Vol. 243, No. 1, 1996, pp. 86-91. http://dx.doi.org/10.1006/abio.1996.0485
  30. M. Trojanowicz, “Recent Developments in Electrochemical Flow Advantages Detections—A Review Part II. Liquid Chromatography,” Analytica Chimica Acta, Vol. 688, No. 1, 2011, pp. 8-35. http://dx.doi.org/10.1016/j.aca.2010.12.024
  31. B. Fuhrmann and U. Spohn, “An Enzymatic Amplification Flow Injection Analysis (FIA) System for the Sensitive Determination of Phenol,” Biosensor and Bioelectronics, Vol. 13, No. 7-8, 1998, pp. 895-902. http://dx.doi.org/10.1016/S0956-5663(98)00061-X
  32. F. Maya, J. M. Estela and V. Cerdà, “Flow Analysis Techniques as Effective Tools for the Improved Environmental Analysis of Organic Compounds Expressed as Total Indices,” Talanta, Vol. 81, No. 1-2, 2010, pp. 1-8. http://dx.doi.org/10.1016/j.talanta.2010.01.028
  33. E. J. Liorent-Martinez, P. Ortega-Barrales, M. L. Fernandez de Cordova and A. Ruiz-Mendina, “Trends in FlowBased Analytical Methods Applied to Pesticide Detection: A Review,” Analytica Chimica Acta, Vol. 684, No. 1-2, 2011, pp. 30-39. http://dx.doi.org/10.1016/j.aca.2010.10.036
  34. Y.-C. Tsai and C.-C. Chiu, “Amperometric Biosensors Based on Multiwalled Carbon Nanotube-Nafion-Tyrosinase Nanobiocomposites for the Determination of Phenolic Compounds,” Sensors and Actuators B, Vol. 125, No. 1, 2007, pp. 10-16. http://dx.doi.org/10.1016/j.snb.2007.01.032
  35. B. Haghighi, L. Gorton, T. Ruzgas and L. J. Jönsson, “Characterization of Graphite Electrodes Modified with Laccase from Trametes versicolor and Their Use for Bioelectrochemical Monitoring of Phenolic Compounds in Flow Injection Analysis,” Analytica Chimica Acta, Vol. 487, No. 1, 2003, pp. 3-14. http://dx.doi.org/10.1016/S0003-2670(03)00077-1
  36. R. S. Freire, N. Duran and L. Kubota, “Development of a Laccases-Based Flow Injection Electrochemical Biosensor for the Determination of Phenolic Compounds and Its Application for Monitoring Remediation of Kraft EI Paper Mill Effluent,” Analytica Chimica Acta, Vol. 463, No. 2, 2002, pp. 229-238. http://dx.doi.org/10.1016/S0003-2670(02)00417-8
  37. S. Korkut, B. Keskinler and E. Erhan, “An Amperometric Biosensor Based on Multiwalled Carbon NanotubePoly(pyrrole)-horseradish Peroxidase Nanobiocomposite Film for Determination of Phenol Derivatives,” Talanta, Vol. 76, No. 5, 2008, pp. 1147-1152. http://dx.doi.org/10.1016/j.talanta.2008.05.016
  38. S. K. Ozoner, F. Yilmaz, A. Celik, B. Keskinler and E. Erhan, “A Novel Poly(Glycine methacrylate-co-3-thienylmethyl methacrylate)-polypyrrole-carbon Nanotube-Horseradish Peroxidase Composite Film Electrode for the Detection of Phenolic Compounds,” Current Applied Physics, Vol. 11, No. 3, 2011, pp. 402-408. http://dx.doi.org/10.1016/j.cap.2010.08.010
  39. R. Solna and P. Skladal, “Amperometric Flow-Injection Determination of Phenolic Compounds Using a Biosensor with Immobilized Laccase, Peroxidase and Tyrosinase,” Electroanalysis, Vol. 17, No. 23, 2005, pp. 2137-2146. http://dx.doi.org/10.1002/elan.200403343
  40. A. Sassolas, L. J. Blum and B. D. Leca-Bouvier, “Immobilization Strategies to Develop Enzymatic Biosensors,” Biotechnology Advances, Vol. 30, No. 3, 2012, pp. 489- 511. http://dx.doi.org/10.1016/j.biotechadv.2011.09.003
  41. J. Kulys and R. Vidziunaite, “Amperometric Biosensors Based on Recombinant Laccases for Phenols Determination,” Biosensors and. Bioelectronics, Vol. 18, No. 2-3, 2003, pp. 319-325. http://dx.doi.org/10.1016/S0956-5663(02)00172-0
  42. L.-S. Duan, F. Xie, F. Zhou and S.-F. Wang, “The Electrochemical Behavior of Acetaminophen on Multi-Walled Carbon Nanotubes Modified Electrode and Its Analytical Application,” Analytical Letters, Vol. 40, No. 14, 2007, pp. 2653-2663. http://dx.doi.org/10.1080/00032710701588341
  43. M. A. Heras, S. Lupu, L. Pigani, C. Pirvu, R. Seeber, F. Terzi and C. Zanardi, “A poly(3,4-ethylenedioythiophene)- poly(styrne sulphonate) Composite Electrode Coating in the Electrooxidation of Phenol,” Electrochimica Acta, Vol. 50, No. 7-8, 2005, pp. 1685-1691. http://dx.doi.org/10.1016/j.electacta.2004.10.029
  44. M. Ferreira, H. Varela, R. M. Torresi and G. TremiliosiFilho, “Electrode Passivation Caused by Polymerization of Different Phenolic Compounds,” Electrochimica Acta, Vol. 52, No. 2, 2006, pp. 434-442. http://dx.doi.org/10.1016/j.electacta.2006.05.025
  45. T. Spataru and N. Spataru, “Voltammetric Detection of Phenol at Platinum-Polytyramine Composite Electrodes in Acidic Media,” Journal of Hazardous Materials, Vol. 180, No. 1-3, 2010, pp. 777-780. http://dx.doi.org/10.1016/j.jhazmat.2010.04.058
  46. T. A. Enache and A. M. Oliveira-Brett, “Phenol and ParaSubstituted Phenols Electrochemical Oxidation Pathways,” Journal of Electroanalytical Chemistry, Vol. 655, No. 1, 2011, pp. 9-16. http://dx.doi.org/10.1016/j.jelechem.2011.02.022
  47. M. Pohanka and P. Skladal, “Electrochemical Biosensors —Principles and Applications”, Journal of Applied Biomedicine, Vol. 6, No. 2, 2008, pp. 57-64.
  48. J. Wang, “Carbon Nanotube Based Electrochemical Biosensors: A Review,” Electroanalysis, Vol. 17, No. 1, 2005, pp. 7-14. http://dx.doi.org/10.1002/elan.200403113
  49. A. S. Santos, A. C. Pereira, M. D. P. T. Sotomayor, C. R. T. Tarley, N. D. Lauro and T. Kubota, “Determination of Phenolic Compounds Based on Co-Immobilization of Methylene Blue and HRP on Multi-Wall Carbon Nanotubes” Electroanalysis, Vol. 19, No. 5, 2007, pp. 549- 554. http://dx.doi.org/10.1002/elan.200603759
  50. S. Campuzano, B. Serra, M. Pedrero, F. J. M. Villena and J. M. Pingarrón, “Amperometric Flow-Injection Determination of Phenolic Compounds at Self-Assembled Monolayer-Based Tyrosinase Biosensors,” Analytica Chimica Acta, Vol. 494, No. 1-2, 2003, pp.187-197. http://dx.doi.org/10.1016/S0003-2670(03)00919-X
  51. T. E. Barman, “Enzyme Handbook,” Springer, New York, 1985.
  52. J.-S. Wang, P.-Y. Chen, T.-T. Huang and M.-S. Lin, “Enzymeless Flow Injection Analysis of 2,4,6-Trichlorophenol Based on Preoxidation by Ammonium Cerium (IV) Nitrate,” International Journal of Electrochemical Science, Vol. 7, No. 10, 2012, pp. 9113-9121.
  53. S. K. Ozoner, M. Yalvac and E. Erhan, “Flow Injection Determination of Catechol Based on Polypyrrole-Carbon Nanotube-Tyrosinase Biocomposite Detector,” Current Applied Physics, Vol. 10, No. 1, 2010, pp. 323-328. http://dx.doi.org/10.1016/j.cap.2009.06.017
  54. D. Vega, L. Agui, A. Gonzalez-Cortes, P. Yanez-Sedeno and J. M. Pingarron, “Electrochemical Detection of Phenolic Estrogenic Compounds at Carbon Nanotube-Modified Electrodes,” Talanta, Vol. 71, No. 3, 2007, pp. 1031-1038. http://dx.doi.org/10.1016/j.talanta.2006.05.071
  55. K. Schachl, H. Alemu, K. Kalcher, J. Jezkova, I. Svanvara and K. Vytras, “Amperometric Determination of Hydrogen Peroxide with a Manganese Dioxide-Modified Carbon Paste Electrode Using Flow Injection Analysis,” Analyst, Vol. 122, No. 9, 1997, 985-989. http://dx.doi.org/10.1039/a701723e
  56. A. Remes, A. Pop, F. Manea, A. Baciu, S. J. Picken and J. Schoonman, “Electrochemical Determination of Pentachlorophenol in Water on a Multi-Wall Carbon NanotubesEpoxy Composite Electrode,” Sensors, Vol. 12, No. 6, 2012, pp. 7033-7046. http://dx.doi.org/10.3390/s120607033
  57. R. S. Freire, N. Durán and L. T. Kubota, “Electrochemical Biosensor-Based Devices for Continuous Phenols Monitoring in Environmental Matrices,” Journal of Brazilian Chemical Society, Vol. 13, No. 4, 2002, pp. 456- 462. http://dx.doi.org/10.1590/S0103-50532002000400008
  58. J. W. Zhao, D. H. Wu and J. F. Zhi, “A Novel Tyrosinase Biosensor Based on Biofunctional ZnO Nanorod Microarrays on the Nanocrystalline Diamond Electrode for Detection of Phenolic Compounds,” Bioelectrochemistry, Vol. 75, No. 1, 2009, pp. 44-49. http://dx.doi.org/10.1016/j.bioelechem.2009.01.005

NOTES

*Corresponding author.