Open Journal of Pediatrics
Vol. 2  No. 2 (2012) , Article ID: 19855 , 6 pages DOI:10.4236/ojped.2012.22021

A lightweight mainstream capnometer with very low dead space volume is useful monitor for neonates with spontanous and mechanical ventilation: Pilot study*

Daijiro Takahashi1,2, Miyu Matsui1, Takehiko Hiroma1, Tomohiko Nakamura1

1Division of Neonatology, Nagano Children’s Hospital, Nagano, Japan

2Department of Pediatrics, Fukuda Hospital, Kumamoto, Japan

Email: tnakamura@naganoch.gr.jp

Received 6 February 2012; revised 8 May 2012; accepted 20 May 2012

Keywords: Capnography; Neonate

ABSTRACT

Objects: The purpose of this study was to observe a correlation between PETCO2 and PaCO2 in intubated neonates under intermittent mandatory ventilation with spontaneous breathing. Material and methods: A total of 55 paired PETCO2 measured by mainstream capnometry and PaCO2 values were obtained from 4 intubated neonates in our neonatal intensive care units at Nagano Children’s Hospital, Nagano, Japan. Results: PETCO2 and PaCO2 were significantly correlated (r2 = 0.928, p < 0.0001). For samples in ventilated neonates with spontaneous breathing, maximum PETCO2 and mean PETCO2 correlated strongly with PaCO2 (maximum PETCO2: r2 = 0.9401, p< 0.0001; mean PETCO2: r2 = 0.8587, p < 0.0001). Although PaCO2 also correlated with minimum PETCO2 (r2 = 0.2884, p < 0.01) in ventilated infants with spontaneous breathing, a significant difference was seen with maximum PETCO2 (p < 0.05) and mean PETCO2 (p < 0.05) in the correlation coefficient r between PaCO2 and PETCO2. Conclusion: Present study showed that a good correlation exists between PETCO2 and PaCO2 in intubated neonates under intermittent mandatory ventilation with spontaneous breathing. Lightweight with low amounts of dead space mainstream capnometry can be used as noninvasive monitor in incubated neonates with spontaneous breathing.

1. INTRODUCTION

Capnography, which displays the level and waveform of CO2 in exhaled air, is a simple technique that appears to accurately indicate arterial PCO2 (PaCO2) and provides information on cell metabolism, blood perfusion, and alveolar ventilation [1-3]. The use of end-tidal CO2 (PETCO2) for monitoring and as a tool for verifying endotracheal tube (ETT) position is another common practice in the operating room and in adult and pediatric intensive care units [3]. This procedure was also introduced to neonatal intensive care units (NICUs), but there are limitations of the technique in smaller babies, especially for extremely low birth weight infants due to issues such as the weight of sensors or water droplets within circuits, dead space, and leakage from tracheal intubation tubes. Recently, a lightweight mainstream capnometer was developed. We have previously reported a strong correlation between PETCO2 and PaCO2 under controlled ventilation when tidal volume/body weight (TV/BW) was 6 - 15 mL/kg and the leakage rate was <60% in rabbits. Furthermore, under conditions of 6 mL/kg of tidal volume (TV), PaCO2 was significantly increased by a dead space increase of only 1 mL, representing >7% of TV [4].

Capnometry is expected as one of the non-invasive monitor in NICUs. Although several investigators have demonstrated good relationships between values for PETCO2 and PaCO2 in infants [5-8], the value of capnometry in estimating PaCO2 has been questioned during anesthesia with spontaneous ventilation [9,10]. On the other hand, while TV is larger than spontaneous breaths, PETCO2 is close to the PaCO2 and to that observed with voluntary maximal expiration [11]. The purpose of this study was to observe a correlation between PETCO2 and PaCO2 in intubated neonates under intermittent mandatory ventilation (IMV) with spontaneous breathing.

2. MATERIAL AND METHODS

A total of 55 paired PETCO2 and PaCO2 values were obtained from the 4 neonates who had been admitted to the NICU at Nagano Children’s Hospital between April and July 2009. Neonates deemed as non-viable by the attending physician were excluded, as were those with conditions such as acute shock, infection, or hemodynamic instability. Before enrollment into this study, informed consent was obtained from the parents or guardian of each infant. Study subjects comprised 4 neonates. Mean (±standard deviation (SD)) gestational age and birth weight were 36.7 ± 2.1 weeks and 2446 ± 487 g, respectively. Data on demographics, clinical characteristics, and laboratory findings for subjects were collected by referring to the clinical laboratory records.

The infants were ventilated mechanically using a timecycled pressure-limited ventilator (Calliope®; Metran, Saitama, Japan). Peak inspiratory pressure (PIP), fraction of inhaled oxygen (FiO2), inspiratory time, positive end expiratory pressure (PEEP) and respiratory rate were settled to provide the optimal arterial PaO2 and PaCO2 as defined by the neonatologists and were not manipulated for the purposes of the study.

Mainstream PETCO2 was measured via a capnograph connected to the proximal end of the endotracheal tube (Cap-One®; TG-970P, Nihon-Kohden, Tokyo, Japan). Data was continuously recorded on a laptop computer using the software programmed by LabVIEW (National Instruments, Texas, USA) through CO2 monitor (OLG-2800; Nihon-Kohden, Tokyo, Japan) each patients. We distinguished between spontaneous breaths and ventilator breaths on the basis of capnography for 20 s at the time of blood gas analysis. For each 20-s sample period, we determined the maximum, mean and minimum values of PETCO2 (Figure 1). Measurements of PETCO2 that did not show an alveolar plateau due to a large amount of leakage were excluded.

The tidal volume was measured by mainstream capnography (CO2SMO 8100, Fukuda Denshi, Tokyo, Japan). The leakage ratio was calculated using the following equation:

Blood samples were drawn from indwelling arterial lines into a 0.1-mL heparinized syringe to prevent coagulation. Blood sampling was performed by heel puncture when arterial line was not placed. Measurements were then immediately made using a bedside blood gas analyzer (ABL 700; Radiometer, Copenhagen, Denmark) for PaCO2. Blood gas analysis was performed for the purposes of evaluation of the patient (including PaO2, PaCO2, electrolytes or lactate, etc.) only. Calibrations were performed automatically for the blood gas analyzer and the accuracy of the capnography was checked by 5% CO2 gas cylinder.

All statistical analyses were conducted using SPSS Statistics version 17.0 (SPSS, Chicago, Illinois). To determine whether PETCO2 were representative of PaCO2, the relationship between PETCO2 and PaCO2 was analyzed by simple linear regression. The standard technique of Fisher’s Z transformation was performed to determine whether a significant difference existed in the correlation coefficient r between PaCO2 and each group of PETCO2.

Furthermore, Bland-Altman plots were performed to assess measurements of PETCO2. Bland-Altman plots demonstrate “good agreement” not only when differences between methods are consistent across all measurements but also when the differences are small. In a situation in which the difference between measurements is expected to change based on a third variable, Bland-Altman plots lose importance. Precision of PETCO2 and the agreement between PETCO2 and PaCO2 were assessed by bias, SD and calculating the 95% confidence interval (CI) for the bias (bias = PETCO2 – PaCO2). Values of p < 0.05 were determined to be significant.

This study was carried out under the control of the Ethics Committee of Medicine and Medical Care, Nagano Children’s Hospital, Nagano, Japan.

3. RESULTS

Mean TV/BW and leakage ratio were 7.3 ± 1.7 mL/kg and 9.5% ± 12.0%, respectively (Table 1). All patients were treated using sedative drugs.

PETCO2 and PaCO2 were significantly correlated (r2 = 0.928, p < 0.0001). In the Bland-Altman plot test, the mean difference (bias) and SD of the differences for

Figure 1. Continuous recording of PETCO2. For each 20-s sample period, we determined maximum A, mean B and minimum C PETCO2.

Table 1. Baseline characteristics.

PETCO2 was –0.88 ± 2.69 mmHg (95% CI for the bias, −1.61 to –0.16 mmHg) (Figure 2). We chose the maximum for PETCO2 on the basis of capnograms for each 20-s period at the time of blood gas analysis.

Due to breath-to-breath variation, we evaluated three measurements of PETCO2 to determine which one most consistently and accurately predicted PaCO2. From 55 measurements, we have selected 24 paired PETCO2 and PaCO2 values which were obtained at the time when spontaneous breathing was present.

For samples in ventilated infants with spontaneous breathing, maximum PETCO2 and mean PETCO2 correlated strongly with PaCO2 (maximum PETCO2: r2 = 0.9401, p < 0.0001; mean PETCO2: r2 = 0.8587, p < 0.0001). Although PaCO2 also correlated with minimum PETCO2 (r2 = 0.2884, p < 0.01) in ventilated infants with spontaneous breathing, a significant difference was seen with maximum PETCO2 (p < 0.05) and mean PETCO2 (p < 0.05) in the correlation coefficient r between PaCO2 and PETCO2 (Figures 3(a)-(c)).

Bland-Altman analysis showed that PETCO2 underestimated PaCO2 by a mean difference (bias) of –0.175 ± 2.31 mmHg (95% CI for the bias, –1.15 to 0.799 mmHg) in the maximum PETCO2, –5.01 ± 3.55 mmHg (95% CI for the bias, –6.50 to –3.51 mmHg) in mean PETCO2, and –14.6 ± 8.82 mmHg (95%CI for the bias, –18.3 to –10.9 mmHg) in minimum PETCO2 (Figures 3(d)-(f)).

4. DISCUSSION

Advances in the treatment of neonatal respiratory failure, including exogenous surfactant [12,13], inhaled nitric oxide (iNO) [14,15], and a growing repertoire of assisted ventilation strategies [16] have decreased morbidity and mortality rates. Patient monitoring has played a critical role in the safe and effective application of these advanced therapies.

Pulse oximetry provides a noninvasive method of assessing oxygenation and continuous surveillance of the partial pressure of arterial oxygen [17]. Maintaining PaCO2 within the desired range by frequent arterial sampling can increase the need for multiple transfusions in the NICU [18], highlighting the need for methods of continuous non-invasive monitoring of CO2 levels. Both hypocarbia and hypercarbia are detrimental to extremely low birth weight infants and have been implicated as

(a)(b)

Figure 2. The relationship between PETCO2 and PaCO2 (a) and Bland-Altman plot shows bias against average values of PETCO2 and PaCO2 in ventilated infants (b). PETCO2 and PaCO2 were significantly correlated (r2 = 0.928, p < 0.0001).

(a)(b)(c)(d)(e)(f)

Figure 3. The relationship between PETCO2 and PaCO2 (a), (b) and (c) and Bland-Altman plot shows bias against average values of PETCO2 and PaCO2 in ventilated infants with spontaneous breathing (d), (e) and (f). Maximum PETCO2 and mean PETCO2 correlated strongly with PaCO2. Conversely, PaCO2 did not correlate with minimum PETCO2.

causative factors in periventricular leukomalacia, intraventricular hemorrhage and chronic lung disease [19-21]. Critical event analyses have documented that hypoxemia secondary to depressed respiratory activity is a principal risk factor for near misses and death [22,23].

Monitoring of PETCO2 is a simple and noninvasive technique that appears to accurately indicate PaCO2 in a variety of clinical situations [1,24]. However, levels of PETCO2 and PaCO2 depend on ventilation, cardiac output, CO2 output, and pulmonary function; a change in any of these will cause a change in PETCO2 [25]. For instance, a growing degree of difference between PETCO2 and PaCO2 can indicate the severity of pulmonary embolism [26] or even the effects of thrombolytic therapy [27].

PETCO2 varied appreciably from breath to breath. The reliability of this value under conditions of significant ventilation perfusion inequality or heterogeneous tidal volumes has thus been questioned [28]. In many cases, spontaneous breaths of variable tidal volumes far outnumbered ventilator breaths, but still contributed relatively little to alveolar minute ventilation. We found that maximum PETCO2 during each sampling period showed the best correlation with PaCO2. A number of investigators have suggested that larger tidal volumes are necessary to measure PETCO2 accurately, as small (e.g., spontaneous) breaths may fail to “wash out” the anatomic dead space [11,29].

Weinger et al. [30] also found a wide range in differences between PETCO2 and PaCO2 over time. However, they measured mean peak PETCO2 over a period of 8 min. Although averaging maximal PETCO2 values over a longer time span might improve the stability and reliability of PaCO2 - PETCO2, assessing the respiratory status to measure PETCO2 over a long time would be difficult. The respiratory apparatus might change second to second.

Takano et al. [31] reported that reliable PETCO2 was obtained when a vital capacity maneuver was performed on each nonintubeted patient, indicating that full expiration to the maximal expiratory position is necessary for precise estimation of PaCO2.

Comparison with spontaneous breathing, end tidal CO2 measured from ventilated breath is close to the PaCO2 and to that observed with a voluntary maximal expiration. This PCO2 gradient between ventilator and spontaneous breathing indicates a large dead-space to tidal volume ratio; much of the expired CO2 appearing with spontaneous breathing is diluted, with dead space air lowering the concentration at any point during expiration.

Although our prospective observational study revealed a good correlation and agreement between PaCO2 and maximum PETCO2, the present study included only a very small number of participants, and the conditions of patients were not constant. Furthermore, in children with congenital cyanotic heart disease, right-to-left intracardiac shunting reportedly causes an obligatory difference between PaCO2 and PETCO2 [32,33]. Our subjects included patients with congenital heart disease because they showed stable cardiorespiratory status in this study.

This study consisted of only a small number of participants, the statistical implications of repeated measurements in the same infants, and the conditions of the cases were not constant. In addition, we did not adjust for cardiac output their measurements, though differences in cardiac output are known to affect PETCO2 measurements [25]. Therefore, we would like to study these issues in the future.

5. CONCLUSION

Our results indicate that a good correlation exists between PETCO2 and PaCO2 in intubated neonates under intermittent mandatory ventilation with spontaneous breathing. Furthermore, maximum PETCO2 correlated strongly with PaCO2 compared to minimum PETCO2. Lightweight with low amounts of dead space mainstream capnometry can be used as noninvasive monitor in incubated neonates with spontaneous breathing.

6. ACKNOWLEDGEMENTS

The authors acknowledge the help and support of Mr. Dainobu, Mr. Inoue, and Mr. Yamamori from Nihon Kohden.

REFERENCES

  1. Burton, G.W. (1966) The value of carbon dioxide monitoring during anaesthesia. Anaesthesia, 21, 173-183. doi:10.1111/j.1365-2044.1966.tb02596.x
  2. Burton, G.W. (1969) Measurement of inspired and expired oxygen and carbon dioxide. British Journal of Anaesthesia, 41, 723-730. doi:10.1093/bja/41.9.723
  3. Bhende, M.S. (2001) End-tidal carbon dioxide monitoring in pediatrics—Clinical applications. Journal of Postgraduate Medicine, 47, 215-218.
  4. Takahashi, D., Hiroma, T. and Nakamura, T. (2011) PETCO2 measured by a new lightweight mainstream capnometer with very low dead space volume offers accurate and reliable noninvasive estimation of PaCO2. Research and Reports in Neonatology, 1, 61-66.
  5. Meyer, R.E. and Short, C.E. (1985) Arterial to end-tidal CO2 tension and alveolar dead space in halothaneor isoflurane-anesthetized ponies. American Journal of Veterinary Research, 46, 597-599.
  6. Geiser, D.R. and Rohrbach, B.W. (1992) Use of end-tidal CO2 tension to predict arterial CO2 values in isoflurane-anesthetized equine neonates. American Journal of Veterinary Research, 53, 1617-1621.
  7. Riker, J.B. and Haberman, B. (1976) Expired gas monitoring by mass spectrometry in a respiratory intensive care unit. Critical Care Medicine, 4, 223-229. doi:10.1097/00003246-197609000-00002
  8. McEvedy, B.A., McLeod, M.E., Kirpalani, H., Volgyesi, G.A. and Lerman, J. (1990) End-tidal carbon dioxide measurements in critically ill neonates: A comparison of side-stream and mainstream capnometers. Canadian Journal of Anesthesia, 37, 322-326. doi:10.1007/BF03005583
  9. Hagerty, J.J., Kleinman, M.E., Zurakowski, D., Lyons, A.C. and Krauss, B. (2002) Accuracy of a new low-flow sidestream capnography technology in newborns: A pilot study. Journal of Perinatology, 22, 219-225. doi:10.1038/sj.jp.7210672
  10. Bhat, Y.R. and Abhishek, N. (2008) Mainstream endtidal carbon dioxide monitoring in ventilated neonates. Singapore Medical Journal, 49, 199-203.
  11. Kugelman, A., Zeiger-Aginsky, D., Bader, D., Shoris, I. and Riskin, A. (2008) A novel method of distal end-tidal CO2 capnography in intubated infants: comparison with arterial CO2 and with proximal mainstream end-tidal CO2. Pediatrics, 122, e1219-e1224. doi:10.1542/peds.2008-1300
  12. Lee, K., Khoshnood, B., Wall, S.N., Chang, Y., Hsieh, H.L. and Singh, J.K. (1999) Trend in mortality from respiratory distress syndrome in the United States, 1970-1995. Journal of Pediatrics, 134, 434-440. doi:10.1016/S0022-3476(99)70200-3
  13. Rodriguez, R.J. and Martin, R.J. (1999) Exogenous surfactant therapy in newborns. Respiratory Care Clinics of North America, 5, 595-616.
  14. Wessel, D.L., Adatia, I., Van Marter, L.J., Thompson, J.E., Kane, J.W., Stark, A.R., et al. (1997) Improved oxygenation in a randomized trial of inhaled nitric oxide for persistent pulmonary hypertension of the newborn. Pediatrics, 100, E7. doi:10.1542/peds.100.5.e7
  15. Cornfield, D.N., Maynard, R.C., de Regnier, R.A., Guiang, S.F., Barbato, J.E. and Milla, C.E. (1999) Randomized, controlled trial of low-dose inhaled nitric oxide in the treatment of term and near-term infants with respiratory failure and pulmonary hypertension. Pediatrics, 104, 1089-1094. doi:10.1542/peds.104.5.1089
  16. Plavka, R., Kopecky, P., Sebron, V., Svihovec, P., Zlatohlavkova, B. and Janus, V. (1999) A prospective randomized comparison of conventional mechanical ventilation and very early high frequency oscillatory ventilation in extremely premature newborns with respiratory distress syndrome. Intensive Care Medicine, 25, 68-75. doi:10.1007/s001340050789
  17. Dziedzic, K. and Vidyasagar, D. (1989) Pulse oximetry in neonatal intensive care. Clinics in Perinatology, 16, 177-197.
  18. Madsen, L.P., Rasmussen, M.K., Bjerregaard, L.L., Nohr, S.B. and Ebbesen, F. (2000) Impact of blood sampling in very preterm infants. Scandinavian Journal of Clinical and Laboratory Investigation, 60, 125-132. doi:10.1080/00365510050184949
  19. Giannakopoulou, C., Korakaki, E., Manoura, A., Bikouvarakis, S., Papageorgiou, M., Gourgiotis, D., et al. (2004) Significance of hypocarbia in the development of periventricular leukomalacia in preterm infants. Pediatrics International, 46, 268-273. doi:10.1111/j.1442-200x.2004.01886.x
  20. Erickson, S.J., Grauaug, A., Gurrin, L. and Swaminathan, M. (2002) Hypocarbia in the ventilated preterm infant and its effect on intraventricular haemorrhage and bronchopulmonary dysplasia. Journal of Paediatrics and Child Health, 38, 560-562. doi:10.1046/j.1440-1754.2002.00041.x
  21. Wallin, L.A., Rosenfeld, C.R., Laptook, A.R., Maravilla, A.M., Strand, C., Campbell, N., et al. (1990) Neonatal intracranial hemorrhage: II. Risk factor analysis in an inborn population. Early Human Development. 23, 129-137. doi:10.1016/0378-3782(90)90136-7
  22. Strauss, R.G. (1991) Transfusion therapy in neonates. American Journal of Disease of Children, 145, 904-911.
  23. Wiswell, T.E., Graziani, L.J., Kornhauser, M.S., Stanley, C., Merton, D.A., McKee, L., et al. (1996) Effects of hypocarbia on the development of cystic periventricular leukomalacia in premature infants treated with high-frequency jet ventilation. Pediatrics, 98, 918-924.
  24. Gothard, J.W., Busst, C.M., Branthwaite, M.A., Davies, N.J. and Denison, D.M. (1980) Applications of respiratory mass spectrometry to intensive care. Anaesthesia, 35, 890-895. doi:10.1111/j.1365-2044.1980.tb03950.x
  25. Ornato, J.P., Garnett, A.R. and Glauser, F.L. (1990) Relationship between cardiac output and the end-tidal carbon dioxide tension. Annals of Emergency Medicine, 19, 1104-1106. doi:10.1016/S0196-0644(05)81512-4
  26. Hatle, L. and Rokseth, R. (1974) The arterial to end-expiratory carbon dioxide tension gradient in acute pulmonary embolism and other cardiopulmonary diseases. Chest, 66, 352-357. doi:10.1378/chest.66.4.352
  27. Verschuren, F., Heinonen, E., Clause, D., Roeseler, J., Thys, F., Meert, P., et al. (2004) Volumetric capnography as a bedside monitoring of thrombolysis in major pulmonary embolism. Intensive Care Medicine, 30, 2129-2132. doi:10.1007/s00134-004-2444-9
  28. Capan, L.M., Ramanathan, S., Sinha, K. and Turndorf, H. (1985) Arterial to end-tidal CO2 gradients during spontaneous breathing, intermittent positive-pressure ventilation and jet ventilation. Critical Care Medicine, 13, 810-813. doi:10.1097/00003246-198510000-00007
  29. Evans, J.M., Hogg, M.I. and Rosen, M. (1977) Correlation of alveolar PCO2 estimated by infra-red analysis and arterial PCO2 in the human neonate and the rabbit. British Journal of Anaesthesia, 49, 761-764. doi:10.1093/bja/49.8.761
  30. Weinger, M.B. and Brimm, J.E. (1987) End-tidal carbon dioxide as a measure of arterial carbon dioxide during intermittent mandatory ventilation. Journal of Clinical Monitoring, 3, 73-79. doi:10.1007/BF00858353
  31. Takano, Y., Sakamoto, O., Kiyofuji, C. and Ito, K. (2003) A comparison of the end-tidal CO2 measured by portable capnometer and the arterial PCO2 in spontaneously breathing patients. Respiratory Medicine, 97, 476-481. doi:10.1053/rmed.2002.1468
  32. Fletcher, R. (1991) The relationship between the arterial to end-tidal PCO2 difference and hemoglobin saturation in patients with congenital heart disease. Anesthesiology, 75, 210-221. doi:10.1097/00000542-199108000-00007
  33. Short, J.A., Paris, S.T., Booker, P.D. and Fletcher, R. (2001) Arterial to end-tidal carbon dioxide tension difference in children with congenital heart disease. British Journal of Anaesthesia, 86, 349-353. doi:10.1093/bja/86.3.349