Theoretical modeling of airways pressure waveform for dual-controlled ventilation with physiological pattern and linear respiratory mechanics
Francesco Montecchia
.
DOI: 10.4236/jbise.2011.44042   PDF    HTML     5,929 Downloads   10,519 Views   Citations

Abstract

The present paper describes the theoretical treatment performed for the geometrical optimization of advanced and improved-shape waveforms as airways pressure excitation for controlled breathings in dual-controlled ventilation applied to anaesthetized or severe brain injured patients, the respiratory mechanics of which can be assumed linear. Advanced means insensitive to patient breathing activity as well as to ventilator settings while improved-shape intends in comparison to conventional square waveform for a progressive approaching towards physiological transpulmonary pressure and respiratory airflow waveforms. Such functional features along with the best ventilation control for the specific therapeutic requirements of each patient can be achieved through the implementation of both diagnostic and compensation procedures effectively carried out by the Advance Lung Ventilation System (ALVS) already successfully tested for square waveform as airways pressure excitation. Triangular and trapezoidal waveforms have been considered as airways pressure excitation. The results shows that the latter fits completely the requirements for a physiological pattern of endoalveolar pressure and respiratory airflow waveforms, while the former exhibits a lower physiological behaviour but it is anyhow periodically recommended for performing adequately the powerful diagnostic procedure.

Share and Cite:

Montecchia, F. (2011) Theoretical modeling of airways pressure waveform for dual-controlled ventilation with physiological pattern and linear respiratory mechanics. Journal of Biomedical Science and Engineering, 4, 320-340. doi: 10.4236/jbise.2011.44042.

Conflicts of Interest

The authors declare no conflicts of interest.

References

[1] Mushin, W.W., Rendell-Backer, L., Thompson, P.W. and Mapleson, W.W. (1980) Automatic ventilation of the lungs. Backwell, Oxford.
[2] Tobin, M.J. (2006) Principles and practice of mechanical ventilation. 2nd Edition, McGraw Hill, New York.
[3] MacIntyre, N.R. and Branson, R.D. (2001) Mechanical ventilation. WB Saunders, Philadelphia.
[4] Chatburn, R.L. (2003) Fundamentals of mechanical ventilation. Mandu Press Ltd, Cleveland Heights.
[5] Branson, R.D. and Chatburn, R.L. (1992) Technical description and classification of modes of ventilator operation. Respiratory Care, 37, 1026-1044.
[6] Chatburn, R.L. and Primiano Jr, F.P. (2001) A new system for understanding modes of mechanical ventilation. Respiratory Care, 46, 604-621.
[7] Chatburn, R.L. (2007) Classification of ventilator modes: update and proposal for implementation. Respiratory Care, 52, 301-323.
[8] Grianti, F., Montecchia, F., Di Bari, L. and Baldassarri, M. (1996) A versatile mechanical ventilator (DIGIT) with high flow stability and a programmable inspiratory phase flow pattern. IEEE Transactions on Biomedical Engineering, 43, 1062-1072. doi:10.1109/10.541248
[9] Montecchia, F., Guerrisi, M. and Canichella, A. (2007) Advanced lung ventilation system (ALVS) with linear respiratory mechanics assumption for waveform optimization of dual-controlled ventilation. Medical Engineering & Physics, 29, 259-276. doi:10.1016/j.medengphy.2006.03.006
[10] Campbell, R.S. and Davis, B.R. (2002) Pressure-controlled versus volume-controlled ventilation: does it matter. Res- piratory Care, 47, 416-424.
[11] Taylor, A.E., Render, K.R., Hyatt, R.E. and Parker, J.C. (1989) Clinical respiratory physiology. W. B. Saunders Company, USA.
[12] Hess, D.R. (2005) Ventilator waveform and the physiology of pressure support ventilation. Respiratory Care, 50, 166-186.
[13] MacIntyre, N.R., Gropper, C. and Westfall, T. (1994) Combining pressure-limiting and volume-cycling features in a patient-interactive mechanical breath. Critical Care Medicine, 22, 353-357.
[14] Branson, R.D. and MacIntyre, N.R. (1996) Dual-control modes of mechanical ventilation. Respiratory Care, 41, 294-305.
[15] Branson, R.D. and Davis Jr, K. (2001) Dual control modes: Combining volume and pressure breaths. Respiratory Care Clinics of North America, 7, 397-408. doi:10.1016/S1078-5337(05)70041-1
[16] Branson, R.D. and Johannigman, J.A. (2005) The role of ventilator graphics when setting dual-controlled modes. Respiratory Care, 50, 187-201.
[17] Laubscher, T.P., Heinrichs, W., Weiler, N., Hartmann, G. and Brunner, J.X. (1994) An adaptive lung ventilation controller. IEEE Transactions on Biomedical Engineering, 41, 51-59. doi:10.1109/10.277271
[18] W.G. Sanborn (1993) Microprocessor-based mechanical ven-tilation,” Respiratory Care, Vol. 38, pp. 72-109.
[19] Simon, F., Jenayeh, I. and Rake, H.H. (2000) Mechatronics in medical engineering: Advanced control of a ventilation device. Microprocessors and Microsystems, 24, 63-69. doi:10.1016/S0141-9331(99)00068-X
[20] Brunner, J.X. (2001) Principles and history of closed- loop controlled ventilation. Respiratory Care Clinics of North America, 7, 341-362.
[21] Branson, R.D., Johannigman, J.A., Campbell, R.S. and Davis Jr., K. (2002) Closed-loop mechanical ventilation. Respiratory Care, 47, 427-451.
[22] Anderson, J.R. and East, T.D. (2002) A closed-loop controller for mechanical ventilation of patients with ARDS. Biomedical Sciences Instrumentation, 38, 289-294.
[23] Tehrani, F.T., Rogers, M., Lo, T., Malinowski, T., Afuwape, S., Lum, M., Grundl, B. and Terry, M. (2004) A dual closed-loop control system for mechanical ventilation. Journal of Clinical Monitoring and Computing, 18, 111-129. doi:10.1023/B:JOCM.0000032744.99885.38
[24] Takeuchi, A., Abe, T., Hirose, M., Kamioka, K., Hamada, A. and Ikeda, N. (2004) Interactive simulation system for artificial ventilation on the internet: Virtual ventilator. Journal of Clinical Monitoring and Computing, 18, 353- 363. doi:10.1007/s10877-005-6268-0
[25] Chatburn, R.L. (2004) Computer control of mechanical ventilation. Respiratory Care, 49, 507-515.
[26] Tehrani, F.T. and Roum, J.H. (2008) Flex: A new computerized system for mechanical ventilation. Journal of Clinical Monitoring and Computing, 22, 121-130.
[27] Tehrani, F.T. (2008) Automatic control of mechanical ventilation. Part 1: Theory and history of the technology. Journal of Clinical Monitoring and Computing, 22, 409- 415. doi:10.1007/s10877-008-9150-z
[28] Tehrani, F.T. and Abbasi, S. (2009) Evaluation of a computerized system for mechanical ventilation of infants. Journal of Clinical Monitoring and Computing, 23, 93- 104. doi:10.1007/s10877-009-9170-3
[29] Barrera, R., Gabovich, N., Sogoloff, H., Groeger, J. and Miodownik, S. (2005) The effect of varying inspiratory flow waveforms on alveolar peak airway pressure and intrinsic positive end expiratory pressure in mechanically ventilated patients. The Internet Journal of Pulmonary Medicine, 5.
[30] Yang, S.C. and Yang, S.P. (2002) Effects of inspiratory flow waveforms on lung mechanics, gas exchange, and respiratory metabolism in COPD patients during mechanical ventilation. Chest, 122, 2096-2104.
[31] Niranjan, S.C., Bidani, A., Ghorbel, F., Zwischenberger, J. B. and Clark Jr, J.W. (1999) Theoretical study of inspiratory flow waveforms during mechanical ventilation on pulmonary blood flow and gas exchange. Computers and Biomedical Research, 32, 355-390.
[32] Lutchen, K.R., Hantos, Z. and Jackson, A.C. (1988) Importance of low-frequency impedance data for reliably quantifying parallel inhomogeneities of respiratory mechanics. IEEE Transactions on Biomedical Engineering, 35, 472-481. doi:10.1109/10.2118
[33] Burke, W.C., Crooke, P.S., Marcy, T.W., Adams, A.B. and Marini, J.J. (1993) Comparison of mathematical and mechanical models of pressure-controlled ventilation. Journal of Applied Physiology, 74, 922-933.
[34] Jonson, B., Beydon, L., Brauer, K., Mansson, C., Valind, S. and Grytzell, H. (1993) Mechanics of respiratory system in healthy anesthetized humans with emphasis on viscoelastic properties. Journal of Applied Physiology, 75, 132-140.
[35] Conti, G., De Blasi, R.A., Lappa, A., Ferretti, A., Antonelli, M., Bufi, M. and Gasparetto, A. (1994) Evaluation of respiratory system resistance in mechanically ventilated patients: The role of the endotracheal tube. Intensive Care Medicine, 20, 421-424.
[36] Crooke, P.S., Hota, S., Marini, J.J. and Hotchkiss, J.R. (2003) Mathematical models of passive, pressure-controlled ventilation with different resistance assumptions. Mathe- matical and Computer Modelling, 38, 495-502. doi:10.1016/S0895-7177(03)90021-5
[37] Milic-Emili, J., Gottfried, S.B. and Rossi, A. (1987) Non- invasive measurement of respiratory mechanics in ICU patients. International Journal of Clinical Monitoring and Computing, 4, 11-30.
[38] Baconnier, P.F., Carry, P.Y., Eberhard, A., Perdrix, J.P. and Fargnoli, J.M. (1995) A computer program for automatic measurement of respiratory mechanics in artificially ventilated patients. Computer Methods and Program in Biomedicine, 47, 205-220. doi:10.1016/0169-2607(95)01651-9
[39] Lucangelo, U., Bernabè, F. and Blanch, L. (2005) Respiratory mechanics derived from signals in the ventilator circuit. Respiratory Care, 50, 55-65.
[40] Bates, J.H., Hunter, I.W., Sly, P.D., Okubo, S., Filiatrault, S. and Milic-Emili, J. (1987) Effect of valve closure time on the determination of respiratory resistance by flow interruption. Medical and Biological Engineering Computing, 25, 136-140. doi:10.1007/BF02442841
[41] Bates, J.H., Baconnier, P. and Milic-Emili, J. (1988) A theoretical analysis of interrupter technique for measuring respiratory mechanics. Journal of Applied Physiology, 64, c f2204-2214.
[42] Lutchen, K.R., Yang, K., Kaczka, D.W. and Suki, B. (1993) Optimal ventilation waveforms for estimating low -frequency respiratory impedance. Journal of Applied Physiology, 75, 478-488.
[43] Barnas, G.M., Harinath, P., Green, M.D., Suki, B., Kaczka, D.W. and Lutchen, K.R. (1994) Influence of waveform and analysis technique on lung and chest wall pro- perties. Respiratory Physiology, 96, 331-344.
[44] Kaczka, D.W., Ingenito, E.P. and Lutchen, K.R. (1999) Technique to determine inspiratory impedance during mechanical ventilation: Implication for flow limited patients. Annals of Biomedical Engineering, 27, 340-355.
[45] Marini, J.J. and Crooke, P.S. (1993) A general mathematical model for respiratory dynamics relevant to the clinical setting. American Review of Respiratory Disease, 147, 14-24.
[46] Bates, J.H. (2007) A,” recruitment model of quasi-linear power-law stress adaptation in lung tissue. Annals of Biomedical Engineering, 35, 1165-1174. doi:10.1007/s10439-007-9291-0

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.