Comparative Thermometery in Paediatric Age Group: Is the Non-Touch Infrared Thermometer (NTIT) Reading Comparable to Regular Mercury-in-Glass Thermometer (MIGT) Reading? ()
1. Introduction
Evaluation of body temperature is one of the oldest and most important step in the diagnosis of disease [1] . Accurate temperature measurement is thus a critical step in evaluating health or disease especially in children and immmunocompromised subjects [1] [2] . Inaccurate measurement may lead to improper diagnosis, wrong treatment or inappropriate intervention [2] .
The temperature measured depends on the site and type of thermometer used. The most reliable means of temperature measurement is the core temperature, which is best taken at the pulmonary artery (site closest to the thermoregulatory center) [1] . Other sites for taking core temperature are the bladder, tympanic membrane, oesophagus, nasopharynx and trachea [1] . However these methods are invasive and not routinely feasible.
The Axilla, rectum and oral cavity are traditionally used to take temperatures in babies [3] Rectal thermometry is slightly invasive and poorly tolerated by children while the oral is painful, and uncomfortable, not recommended for use in patients with oral surgery, newborns or uncooperative and unconscious patients [4] Readings are also affected by consumption of hot or cold foods [4] .
Types of thermometers include electronic contact thermometers, chemical thermometers, infrared thermometers and mercury-in-glass thermometers. Mercury-in-glass thermometers (MIGT) are no longer used in the developed world because of the risk of mercury contamination and poisoning.
Non touch infra-red thermometers (NTIT) are newer and measure body temperature via infrared emission from the body [4] . They are non-invasive, convenient and easier to use [2] The risk of transmission of infections is minimized with NTIT especially in the paediatric age-group in whom minimal touch and infection control is very important [1] [2] .
Several studies have compared body temperatures using different thermometers [3] [4] [5] [6] . While Gasim et al. [5] and Edelu et al. [3] reported that infrared tympanic thermometers were as reliable and accurate as axillary mercury-in-glass thermometers in subjects studied. Yetman et al. [6] reported that temperatures obtained with infrared tympanic membrane thermometers did not accurately reflect an infant’s rectal or axillary temperatures. Similarly, Basak et al. [4] reported a weak correlation between temperatures taken with NTIT and chemical or tympanic membrane thermometers. On the other hand, Abdulkadir et al. [7] reported that infrared tympanic thermometry readings in febrile children better reflected rectal temperature than axillary temperature. None of these studies has actually compared thermometry readings from forehead NTIT and MIGT in the Paediatric age group. This study aims to compare this with the aim of suggesting a faster, more reliable and convenient means of taking temperature in children.
2. Method
Subject recruitment was carried out by the Researcher and two research assistants who had been duly trained by the Researcher after an Ethical clearance was obtained as part of a larger Poliomylitis Vaccine Trial from the University of Iloin Teaching Hospital Ethics and Research Committee. Inclusion criterial included: well baby with, aged 1 to 24 months, and consent given.All children that met the inclussion criteria were recruited over a six months period (between September 2016 and February 2017). After a clear explaination had been given to the mother/caregiver, a study proforma was administered in an interview form to obtain relevant socio-demographic features such as age of the baby, social class and immunisation history. A brief physical examination was also carried out on the babies. The elligible babies and their mothers were made to sit in the same room with a constant room temperature. Temperature of babies were taken thus:
Using the NTIT: The NTIT was held at a maximum distance of 5 cm from the midpoint of the forehead area until the signal tone was heard (average of 1 - 2 secs). Temparatures were taken 3 times and the average reading calculated and recorded.
Using the MIGT: The MIGT was cleaned with alcohol swab and shaken before each use, to drop the mercury column to below 35˚C and placed in the axilla for 3 - 5 minutes before the reading was taken. One reading was taken for each patient. The MIGT was then cleaned again with an alcohol swab.
3. Data Analysis
Data were entered into a computer and analysed with SPSS version 21.
Pearson correlation was used to determine the relationship between the two methods of temperature measurements, while Bland-Altmanmethod was used to test for level of agreement between them.Level of agreement was said to be good as if 95% of the data laid between 1.96 SD of the mean difference between the two methods [8] . According to previous studies, mean of difference was considered good if <0.5˚C, and satisfactory if <0.6˚C [9] [10] .
4. Results
A total of Four hundred and thirty-seven (437) children were recruited from the Immunisation clinic of UITH, Amilegbe Ilorin. They were aged 1 to 24 months. The mean age and SD was 5.81 ± 4.04 months. Almost Fifty-four percent (53.8%) of the children were males, while 46.2% were females. The mean birth weight was 3.23 kg. While 74.1% of the children were born in a hospital, 9.8% of them were born at home.
The other socio-demographic variables are as shown in Table 1 and Table 2.
Concerning the immunization status, 94.1% of the babies had immunization at birth, 92% had gone for the 6-week immunization visit (Table 3).
Not applicable were infants who were not old enough for the respective Vaccines.
The average temperatures obtained were 36.5˚C ± 0.30˚C for NTIT and 36.59˚C ± 0.44˚C for MGIT (Table 4).
Using the Pearson correlation technique, there was apositive relationship between the two methods of temperature measurements. (r = 0.281, p < 0.001) Further details are as shown in Figure 1.
The Bland-Altman plot showed good agreement between the two methods as 96.8% (423/437) of the paired differences in the readings fell within the limit of agreement as shown in Figure 2. Furthermore, the width of agreement was narrow (1.7791) with a lower limit of −0.8107 (95% CI, −0.8845 to −0.7369) and an upper limit of 0.9884 (95% CI, 0.9146 to 1.0623). Further details are shown in Figure 2.
Table 1. Socio-demographic variables of the study population.
Table 2. Schema of social classification [12] .
Table 3. Frequency of illness, use of antipyretics and immunizations of children.
Table 4. Mean temperature measurement.
Figure 1. Correlation of the temperature reading between Non-touch infrared thermometer (NTIT) and the Axillary reading with Mercury in Glass Thermometer (MGIT).
Figure 2. Bland Altman Graph showing the limit of agreement between the MIGT and NTIT readings.
5. Discussion
The mean axillary temperature of 36.59˚C ± 0.44˚C for MGIT and 36.5˚C ± 0.30˚C for NTIT in this study was similar to 36.82˚C ± 0.44˚C and 36.40˚C ± 0.42˚C reported by Sollai et al. [9] , but much lower than the 37.18˚C ± 0.96 ˚C and 37.30˚C ± 0.92 ˚C reported by Chiapinni et al. [11] . The reason for this disparity might be due to the fact that the current study recruited children from an immunization (well-baby) clinic while Chiapinni et al. [11] recruited patients from the Emergency unit. This is similar to the findings of Chiapinni et al. [11] .
The current study finding of a positive correlation between NTIT and MGIT is consistent with the findings of Gasim et al. [5] and Edelu et al. [3] . This demonstrates a linear relationship between the temperature readings by both methods and implies that one method can be used in place of the other. The finding of a good agreement between the two thermometers is consistent with the reports of Sollai et al. [9] and Chiapinni et al. [11] ; but contrary to Sethi et al. [10] and Basak et al. [4] who reported no agreement between infra-red thermometers and mercury thermometers. The difference could have been due to the fact that while the current research studied older children who were healthy, Sethi et al. [10] studied the sick neonates admitted in the intensive care unit.
Edelu et al. [3] , although reported a positive correlation between the two methods of measurements, found no agreement between them. The disparity in the two studies may however be due to the fact that Edelu et al. [3] used the infra-red tympanic thermometer in the oral mode; meaning the tympanic temperature reading was converted to an approximate oral temperature reading. This process of conversion may have altered the sensitivity. Also, rectal temperature was compared with the oral temperature as against forehead and axillary temperatures used in the current study.
6. Conclusion
MIGT and forehead NTIT had a good positive correlation and a good agreement.
Recommendation
For speed and convenience as well as hygiene and infection control, MGIT can be safely replaced by forehead NTIT in clinical Paediatric practice. Also use of MIGT can be discontinued to reduce the concerns about environmental mercury pollution.
Limitation of the Study
The neonatal age group was not included in the study. Also, Tympanic temperature, which is a better estimate of the core temperature was not done in this study and may be a line of future research in healthy children.
Funding
No specific Grants available for this work.